Ville Renvall1 and Jonathan R. Polimeni2
1Department of Neuroscience and Biomedical Engineering, Aalto University School of Science, Espoo, Finland, 2Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA, United States
Synopsis
T1 mapping using multiple inversion time IR-EPI can provide
a large number of different TI values in a short time, which can be utilized to
characterize the relaxation time distributions within individual voxels, as an
extension to multi-parametric fitting.Purpose
To improve the understanding of the constituents of voxels
beyond obtaining a single or a limited number of parameters per voxel.
Introduction
Quantitative relaxation parameters provide fundamental
information about tissues. However, especially T1 mapping using inversion
recovery (IR) is time consuming due to the requirements of waiting for T1
recovery and managing RF energy deposition (SAR) . Consequently, IR-based T1
mapping conventionally uses a limited set of samples (e.g., 1 or 2) during the
recovery. While these samples can characterize mono-exponential decay, voxels
with a spectrum of T1 values could be better described with increased sampling
of the signal change during the inversion recovery. In this study we used a Multiple
Inversion time Echo Planar Imaging sequence
1,2 (MI-EPI) to quickly obtain
many samples from the inversion recovery, and use these data to estimate T1
distributions to better describe the voxels at the microscopic level. The rapid
acquisition of EPI enables whole-brain coverage and dense sampling of the IR. Information
of the T1 distribution within a voxel can potentially provide the elucidation sub-voxel
composition and also to quantify dynamic partial volume effects due to tissue
motion that cause signal variation in functional MRI
3,4.
Methods
Four healthy volunteers participated in this study. Data
were acquired using a 7 Tesla whole-body scanner scanner (Siemens Healthcare)
with a 32-channel array head coil for receive and a birdcage transmit coil. We
acquired MI-EPI data where every inversion was followed by the readout of all
slices, such that each slice was acquired at a different inversion time (TI).
After each inversion recovery period, the slice acquisition order was permuted
by one slice. After repeating the acquisition a number of times equaling the
number of slices (Nslice), a set of data with Nslice equally
spaced TI values were available for every slice. A dictionary of signal evolution
was simulated using the Bloch equation with a range of values given to two
parameters, T1 and a remainder T2*-weighted term (called S0), and the least-squares
error (LSE) was computed between the data and the model.
Conventionally, the single T1/S0
assigned to a given voxel using this dictionary lookup procedure represents the
global minimum of the LSE across the dictionary entries. Depending on which subset
of the samples along the inversion recovery are used to model the measured
signal, different T1/S0 values can be obtained. E.g., using only the low TI
values could better represent the short T1 components whereas the higher TI
values could result in longer T1 estimates as the dynamics of the short T1
constituents would already have elapsed. Here we randomly selected the samples
along the IR curve (with replacement) into N sets and obtained N estimates of
T1 and S0; we then assembled a smoothed histogram of these N parameter
estimates for each voxel. Broad distributions could be regarded as representing
the partial volume contribution of tissues with different T1/S0 values. E.g., a
voxel containing both white matter (WM) and gray matter (GM) could have a “T1
spectrum” spanning the T1 values of those two tissue classes.
IR-EPI Protocol parameters: TR =
2.87 s, TE = 23 ms, excitation flip angle = 80°, field-of-view = 192 mm, matrix
= 184 × 184, slice thickness =
1.04 mm, Nslice = number of TIs = number of TRs = 50, GRAPPA R = 4, effective echo spacing = 0.205
ms, spacing of different TI values = 56 ms. The sequence employed FLEET ACS
reference acquisitions5 and trFOCI adiabatic inversions6.
Fitting: 2 or 5 TI values were randomly sampled 500 times for every parameter
histogram. A column of voxels spanning a cortical sulcus was selected
(arbitrarily) for showing the results.
Results
Figure 1 shows an axial T1 map generated from the whole data
set and a close-up where the voxels of interest have been marked. The figure
also includes the smoothed histograms resulting from the random sampling using
2 or 5 inversion times to obtain the T1 values.
Discussion
A wide and dense sampling at different TIs is required to
characterize the internal T1 composition of a voxel. The speed of EPI and SNR
provided by 7 T enabled investigation into individual voxels at a high spatial
resolution. Using a large amount of data, here 5 samples,
yielded robust T1 estimates with clear peak values at every WM and GM voxel
shown (T1 of cerebrospinal fluid (CSF) was less well defined). However, the
more wide-spread T1 distributions from the two-sample fitting provided
potentially useful information about the alternative possibilities for the T1
values within the voxels.
Acknowledgements
Supported by NIH NIBIB K01-EB011498, P41-EB015896, and R01-EB019437,the Academy of Finland (grant #265917), the Finnish Cultural Foundation Kalle and Dagmar Välimaa fund and
the Athinoula A. Martinos Center for Biomedical Imaging, and made possible
by NIH NCRR Shared Instrumentation Grants S10-RR023401 and S10-RR020948.References
1. Clare S, Jezzard P (2001). Rapid T(1) mapping using
multislice echo planar imaging. Magn. Reson. Med. 45:630–634.
2. Renvall V, Witzel T, Wald LL, Polimeni JR (2014).
Fast variable inversion-recovery time EPI for anatomical reference and
quantitative T1 mapping. In: Proc. Intl. Soc. Mag. Reson. Med., 22, 4282.
3. Polimeni
JR, Bianciard M, Keil B, Wald LL (2015). Cortical depth dependence of physiological
fluctuations and whole-brain resting-state functional connectivity at 7T. In: Proc.
Intl. Soc. Mag. Reson. Med., 23, 592.
4. Renvall
V, Witzel T, Bianciardi M, Polimeni JR (2014). Multi-contrast
inversion-recovery EPI (MI-EPI) functional MRI at 7 T. Proc. Intl. Soc. Mag.
Reson. Med., 22, 1488.
5. Polimeni JR, Bhat H, Witzel T, Benner T, Feiweier
T, Inati SJ, Renvall V, Heberlein K, Wald LL (2015). Reducing sensitivity losses
due to respiration and motion in accelerated echo planar imaging by reordering
the autocalibration data acquisition. Magn. Reson. Med. http://dx.doi.org/10.1002/mrm.25628.
6. Hurley AC, Al-Radaideh A, Bai L, Aickelin U, Coxon
R, Glover P, Gowland PA (2010). Tailored RF pulse for magnetization inversion at
ultrahigh field. Magn. Reson. Med. 63:51–58.