Ronald Mooiweer1, Shaihan J Malik2, Joseph V Hajnal2, Nico van den Berg1, Peter R Luijten1, and Hans Hoogduin1
1UMC Utrecht, Utrecht, Netherlands, 2Division of Imaging Sciences and Biomedical Engineering, King’s College London, London, United Kingdom
Synopsis
In
this work the design of SPINS excitation pulses has been expanded for use in
TSE sequences and was compared to dynamic RF shimming using DSC in a standard
T2w TSE sequence. We have demonstrated homogeneous 90 degree
excitation, but in itself this was not sufficient to make TSE images uniform.
Manipulating the refocusing pulses (using DSC) remains a necessity. Purpose
Turbo spin echo
(TSE) sequences with a long train of non-selective, variable flip angle,
refocussing pulses are very time-efficient1. However, they are also affected by
the transmit field (B1+) inhomogeneity that arises at
high field strength MRI (≥3T). Recently, signal homogenisation of such
sequences using Direct Signal Control (DSC) has been presented2. Here, RF shim settings are
dynamically updated between pulses to generate an optimal signal over the sequence.
Alternatively,
several methods are available that improve the homogeneity of excitation pulses,
mainly targeted at gradient echo sequences. In general, the calculation and
implementation of excitation pulses is more straightforward than that of a set
of RF shim settings, for which the entire sequence needs to be simulated. Custom
excitation pulses, such as Spiral Nonselective (SPINS) pulses3, are usually designed for small flip
angles and are ignorant to the phase of the excitation. TSE sequences, however,
require a 90˚excitation flip angle at a 90˚phase with
respect to the refocusing pulses, to realise the highest possible signal.
In
this work, SPINS excitation pulses have been adapted for use in TSE sequences
and were compared to dynamic RF shimming using DSC.
Methods
This study
has been conducted on a 3T Achieva system (Philips) with an 8-channel transmit body
coil and 8-channel receive head coil. Following local safety guidelines, the
system-calculated head SAR was kept below 12% of the 3.2W/kg limit to prevent
local SAR hotspots. B1+ and B0 were mapped as
described in (2).
SPINS
pulse design3, based on magnitude least squares optimisation,
was combined with reVERSE4 to stay within peak RF power limits(Fig.1). Bloch-simulations5 showed a non-zero phase across the excited volume, with only a small spread
around the mean value (Fig.2). The mean value was added as a global phase
offset to the RF pulses to steer
the effective phase towards zero: the system-defined phase setting for
excitation pulses in CPMG sequences. The fidelity of 90˚SPINS pulses was
measured in a 3D-AFI sequence6 with TR1/TR2=50/250ms,
to stay within SAR limits.
DSC
optimisation was targeted at the T1 and T2 of CSF. CSF is present in the centre
and periphery of the brain, so an improvement upon the periphery-shaded quadrature
mode is expected.
The
T2w-3D-TSE sequence has 106 echoes per shot (including 6 dummies), TR=2500ms,
ESP=4.0ms, asymptotic refocusing flip angle 35˚, resolution (1mm)^3, 2D-SENSE=1.7x1.7 and NSA=2.
Scan time is 7m35s for whole head coverage.
Results
The
B1+-maps in Figure 3 show a flattened excitation using
SPINS, approximately 7% below target. The phase relative to quadrature is close
to zero (±20˚).
Figure 4
shows that SPINS-excited TSE, compared to quadrature TSE, results in slightly
higher signal in grey and white matter, and a decrease in CSF signal. DSC has a
more pronounced effect, showing a more homogeneous intensity distribution of
the CSF over the brain volume.
Discussion
The phase distribution
of SPINS pulses is narrow, even though phase wasn’t constrained in pulse
optimisation. This is probably due to only low spatial frequencies being visited
in transmit k-space. Numerical optimisation specifically targeting a flat phase
might be beneficial in the future. Some degradation of the excitation magnitude
is seen in the frontal cortex in the simulation, but a matching signal gap
hasn’t been found any of the experiments.
The excitation
magnitude is quite homogeneous, but does not reach the desired FA. This is
partially inherent to the measuring method: 2-3˚ underestimation is expected at this FA6. The FA discrepancy could also be
due to the small tip angle approximation which was used to design the pulse.
Still, a slightly lower-than-ideal, but homogeneous signal, would be of more
use than the inhomogeneous signal generated by quadrature pulses.
Using SPINS
excitation, a decreased TSE signal ratio is observed in the CSF near the top of
the brain. This coincides with a measured phase offset, suggesting signal loss
due to violation of CPMG conditions. This wasn’t predicted in the Bloch simulation.
Even though
the SPINS excitation is flatter than quadrature excitation, when used in a TSE
sequence it improves the homogeneity only slightly. The >100 refocusing
pulses in quadrature mode following every SPINS excitation are probably the
cause of this. A more homogenous signal is found using DSC.
This work
is conducted at 3T, but has obvious extension to 7T where the illustrated
effects are more severe.
Conclusion
We have
demonstrated homogeneous 90˚ excitation, which by itself is not sufficient to make TSE images
uniform. Manipulating the refocussing pulses remains a necessity. Further
improvements might be made when dynamic RF shimming and excitation pulse design
are combined.
Acknowledgements
No acknowledgement found.References
1.
Busse RF, Hariharan H, Vu A, Brittain JH. Fast spin echo sequences with very
long echo trains: design of variable refocusing flip angle schedules and
generation of clinical T2 contrast. Magn. Reson. Med. 2006;55:1030–7. doi: 10.1002/mrm.20863.
2. Malik SJ, Beqiri A, Padormo F,
Hajnal J V. Direct signal control of the steady-state response of 3D-FSE
sequences. Magn. Reson. Med. [Internet] 2014;00. doi: 10.1002/mrm.25192.
3. Malik SJ, Keihaninejad S,
Hammers A, Hajnal J V. Tailored excitation in 3D with spiral nonselective
(SPINS) RF pulses. Magn. Reson. Med. 2012;67:1303–15. doi:
10.1002/mrm.23118.
4. Lee D, Grissom W a., Lustig M,
Kerr AB, Stang PP, Pauly JM. VERSE-guided numerical RF pulse design: A fast
method for peak RF power control. Magn. Reson. Med. 2012;67:353–362. doi:
10.1002/mrm.23010.
5. Bloch simulator: http://www-mrsrl.stanford.edu/~brian/blochsim/
6. Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn. Reson. Med. 2007;57:192–200. doi: 10.1002/mrm.21120.