Synopsis
A
new simultaneous multi-slice excitation method was proposed to reduce RF power
delivery by only concerning slice profiles within the imaging object. Compared to
MultiPINS, our approach used only 79% of the RF energy to the same result (slice
thickness = 3mm, MB factor = 5, 4 bandwidth time product, excitation duration
6380 $$$\mu s$$$). This
excitation method was experimentally demonstrated in spin-echo EPI with blipped
CAIPI acquisition.PURPOSE
Fast image acquisition can improve patient’s comfort, reduce
motion artifact, and allow delineating important physiological information with
high temporal resolution. Recently, blipped CAIPI for simultaneous multi-slice
(SMS) EPI
1 was applied successfully
to high temporal resolution fMRI and DTI acquisition time reduction.
Furthermore, WAVE-CAIPI
2 has been demonstrated that
13 slices can be simultaneously
acquired in Turbo Spin Echo imaging with small noise amplification penalty. However,
simultaneous-multi-slice excitation requires high RF power, because one straightforward
and typical approach is to achieve multi-slice excitation by applying linear
combination of RF pulses designed for individual slices. Using this method, the
transmitted energy and peak power is linearly proportional to the number of
slices. To overcome this problem,
blipped gradient and rectangular RF pulses were designed to excite aliasing slices
3. However, long time gap
between gradient blips is required to avoid excessive transmitted power., MultiPINS
can reduce transmitted power (roughly 50%) using the same excitation duration
by combing the typical multiband and PINS excitation method together to
distribute RF power evenly over time
4.
Since we only concern the excitation pattern inside an imaging
object, further reduction on the transmitted RF energy is possible if RF pulses
are designed to achieve the desired slices within the imaging object without
any restriction on the excitation pattern outside the imaging object. Accordingly,
we present simultaneous-multi-slice ROI-optimized excitation method.
Compared to MultiPINS, our approach used only 79% of the RF energy to the same
result (slice thickness = 3mm, MB factor = 5, 4 bandwidth time product,
excitation duration 6380 $$$\mu s$$$).
This excitation method was experimentally demonstrated in spin-echo SMS EPI
with blipped CAIPI acquisition.
METHODS
Without considering the excitation profile outside the
imaging object, the multi-slice excitation method can excite the desired slices
within imaging object with the minimal RF power. Given predetermined slice
gradient waveform G(t) and $$$k(t)=γ\int_0^tG(s)ds$$$, the RF waveform can be
derived from an optimization problem: $$RF_{ROI} (k(t))=argmin_{RF_{opt}} \| \int_0^TRF_{opt} (k(t)) e^{2πik(t)z} -m(z)\|_2^2+\alpha‖RF_{opt} ((k(t))\|_2^2$$
, where $$$m(z)$$$ is the desired excitation slice profile, $$$\gamma$$$ is the gyromagnetic ratio and $$$\alpha$$$ is a regularization parameter. Figure 1 shows RF waveforms, gradient waveforms, and
excited slice profile of typical MB excitation, PINS, MultiPINS, and ROI-optimized
methods. RF powers were calculated using the following excitation parameters: slice
thickness = 3 mm, slice separation = 30 mm, FOV = 150mm, MB factor = 5 , BWTP =
4, gap between slice gradient blips = 20, 40, 60, 80 us, and slice gradient
slew rate limit = 180 T/m/s. The duration of excitation using PINS and
MultiPINS methods can be calculated based on parameters listed above. The same
duration was used for typical MB excitation with constant slice gradient
strength. The same gradient waveform designed by PINS and MultiPINS methods were
used in ROI-optimized method. To compare RF power, the RF power in typical MB
excitation with 8720 $$$\mu s$$$ duration was set to 1.
Experiments on a 3T
scanner (Siemens) used spin-echo EPI with blipped CAIPI acquisition.to acquire images
covering the whole brain with the parameters listed above, except FOV (slice
direction) = 150 mm, FOV (imaging plane) = 192 mm, duration between two slice
gradient blips = 20 ms,, FOV shift between slices =1/3, TE = 60
ms, and TR = 800 ms. Both 90 degree pulse and 180 degree refocusing pulse were further
optimized by the optimal control method. Fat suppression was applied..
RESULT
Table 2 shows the RF power of the PINS method (2.2) was
larger than that of the typical MB method (1.4) when the time gap between two
gradient blips was short (20 $$$\mu s$$$).The ROI optimized method used 79%, 82%, 84%, and 86% of the RF power required
by MultiPINS method at 20, 40, 60, 80 $$$\mu s$$$ excitation gaps. Figure 2 shows simultaneously
excited and acquired 5 brain images of spin-echo EPI with ROI-optimized
excitation method and blipped CAIPI acquisition.
DISCUSSION
Both ROI-optimized method and MultiPINS method can
reduce RF power even the time gap between gradient blips is short. Unlike PINS only
delivering RF pulse during gaps between gradient blips, ROI-optimized and
MultiPINS methods distribute RF power over the whole excitation duration.
Furthermore, ROI-optimized method can further reduce RF power by designing RF
pulses with a higher degree of freedom, because only region within the imaging
object was considered in the pulse design. The RF power difference between
these methods gradually decreased as the excitation duration increases, because
blip gaps were wider to allow RF pulses between gaps.
Acknowledgements
This study was supported by Ministry of Science and Technology, Taiwan (MOST 104-2314-B-002-238, MOST 103-2628-B-002-002-MY3), National Health Research Institute, Taiwan (NHRI-EX104-10247EI), and Ministry of Economic Affairs, Taiwan (100-EC-17-A-19-S1-175).References
1. Setsompop K, Gagoski BA, Polimeni JR,
Witzel T, Wedeen VJ, Wald LL. Blipped-controlled aliasing in parallel imaging for simultaneous multislice
echo planar imaging with reduced g-factor penalty. Magnetic Resonance in Medicine 2012;67(5):1210-1224.
2. Gagoski
BA, Bilgic B, Eichner C, Bhat H, Grant PE, Wald LL, Setsompop K. RARE/turbo
spin echo imaging with simultaneous multislice Wave-CAIPI. Magnetic Resonance in Medicine
2015;73(3):929-938.
3. Norris
DG, Koopmans PJ, Boyacioglu R, Barth M. Power independent of number of slices
(PINS) radiofrequency pulses for low-power simultaneous multislice excitation. Magnetic resonance in
medicine 2011;66(5):1234-1240.
4. Eichner
C, Wald LL, Setsompop K. A low power radiofrequency pulse for simultaneous
multislice excitation and refocusing. Magnetic Resonance in Medicine
2014;72(4):949-958.