Hong Shang1, Hai Luo2, Xia Liu2, Gaojie Zhu2, and Leping Zha2
1Bioengineering, UC Berkeley - UCSF, Berkeley/San Francisco, CA, United States, 2AllTech Medical Systems, Chengdu, China, People's Republic of
Synopsis
Two classes of
previously proposed nonlinear phase RF pulses, the quadratic-phase pulse and
root-flip optimized pulse, are compared in terms of selectivity, peak B1 value,
pulse energy, and sensitivity to B1 variations, when applied for spatial outer
volume suppression. Root-flip pulses have lower peak B1 and energy given the
same transition width and pulse duration, or sharper transition given the same
peak B1, while quadratic-phase pulses have less sensitivity to B1 variations that
maintains profile shape with B1 deviations, and thus less prone to residual saturation
band magnetization. This work provides insights to pulse designers in regards
to nonlinear phase pulse design and application.Purpose
Nonlinear
phase RF pulses have been proposed and commonly used for broadband saturation
and inversion with reduced peak B
1 amplitude, including quadratic-phase pulse
1,2 and root-flip optimized pulse
3. However, practical comparison between them
is not well known. In this work, the two classes of pulses were studied in
terms of spatial selectivity, peak B
1 requirement, pulse energy, and B
1 sensitivity, when used for outer volume suppression (OVS) applications.
Methods
To achieve
high spatial selectivity for OVS a high time-bandwidth-product (TBW) is necessary,
typically by adjusting bandwidth and fixing pulse duration. A broadband pulse
can also reduce B0 sensitivity and displacement from chemical shifts. Nonlinear
phase pulses distribute RF energy more evenly over the pulse duration, which
helps keep peak B1 within system limits.
Both RF pulses
are designed based on the Shinnar-Le Roux (SLR) algorithm 4, which addresses
the nonlinearity of the Bloch equation at higher flip angles and achieves good
agreement with target pulse profile, by reversibly transforming the RF pulse
design to the well-established finite impulse response (FIR) filter design. The
FIR filters are designed and optimized within the convex optimization framework 5.
(A) Quadratic-phase pulse design. The
quadratic phase Beta polynomial is designed by specifying a quadratic phase in
the desired frequency response, and solved as a second order cone programming. Different
from the equal-ripple designs in 2, an equal-ripple-passband and
decaying-ripple-stopband pulse is designed to reduce perturbation into the imaging
volume. The amount of quadratic phase is set with a few design iterations.
(B) Root-flip pulse design. A minimum
phase Beta polynomial that has the minimal transition bandwidth by bisection
search 5 was first designed, then its passband roots are individually flipped
with respect to the unit circle, to find a configuration with minimum peak B1 amplitude
through all possible combinations. The monte-carlo flip pattern and advanced
algorithm for calculating roots introduced in 6 was used to reduce numerical
error.
Results
One example of design of
the quadratic-phase pulse and the root-flip pulse was shown in Fig. 1, with the
same pulse duration of 4ms, number of samples 400 (2x upsampling),
passband/stopband ripple
0.006/10-4, and transition width 6.25mm. Bloch simulation of selection profile
was shown in Fig. 2, notice the quadratic-phase pulse does have a quadratic Mxy
phase profile. Though optimized for peak B1, the root-flip pulse also shows an
approximate quadratic Mxy phase profile.
The pulse performance is
summarized in the table in Fig. 2.
Simulated
saturation profile with B1 deviation was shown in Fig. 3. Saturation profile measured on a whole body
1.5T clinical scanner (AllTech EchoStar, AllTech Medical Systems, Chengdu,
China) with a sphere water phantom is depicted in Fig. 4. Volunteer L-spine image examples with the quadratic-phase
OVS pulses are in Fig. 5 to demonstrate the practical effectiveness.
Discussion
Both root-flip
and quadratic-phase pulses can achieve high spatial selectivity while keeping
peak B1 within typical system limits in practical applications. Root-flip pulses
spread energy more evenly over time, thus has smaller peak B1 demands compared
to quadratic-phase pulse, especially at high TBW. Root-flip pulses also have
sharper transitions, because only the magnitude frequency response target is
considered during the FIR filter design. Thus, root-flip pulses tend to have
lower peak B1 and power given the same transition, or sharper transition given
the same peak B1.
With B1
deviations, the actual flip angles change for both pulses because they are not adiabatic.
Nevertheless, we found the quadratic-phase pulses maintain the target bandwidth
and profile shape much better than the root-flip pulses, as the selection
profiles of the latter are rapidly distorted, especially over passband
(saturation band), as demonstrated in simulation and imaging experiments. This
is because quadratic-phase pulse with quadratic phase modulation in the time
domain, as shown in Fig. 1, behaves similar as an adiabatic pulse with offset-independent
adiabaticity, but scaled down to the non-adiabatic regime 7. Adiabaticity is
not a design target in this work, as B1 insensitivity can also be achieved by a
train of OVS pulse, which contributes to T1 insensitivity as well 8. Retaining
similar flip angles over the saturation band with B1 variations makes quadratic-phase
pulse more robust when repeated multiple times in OVS. However, a non-uniform saturation
profile with B1 deviations may be less problematic for some other applications,
like nonlinear phase refocusing pulse, where signal is averaged over the slice 9.
This
comparison provides practical insights to pulse designers in choosing one of
the two classes of pulses in nonlinear phase OVS pulse design.
Acknowledgements
No acknowledgement found.References
[1] LeRoux et al. JMRI 8: 1022-1032 (1998), [2]
Schulte et al. JMR 166: 111-122 (2004), [3] Shinnar, MRM 32: 658–660(1994), [4]
Pauly et al. IEEE TMI 10: 53-65 (1991), [5] Shang et al. ISMRM (2015), [6] Lustig
et al. In: Proceedings of the ENC (2006), [7] Luo et al. MRM 45.6 1095-1102
(2001), [8] Tran et al. MRM 43 : 23-33 (2000). [9] Zhu et al. ISMRM (2014).