Spatiotemporally encoded anatomical shape in-plane excitation with reduced profile distortion from field inhomogeneity
Ying Chen1, Song Chen1, Zhong Chen2, and Jianhui Zhong1,3

1Center for Brain Imaging Science and Technology, Zhejiang University, Hangzhou, China, People's Republic of, 2Department of Electronic Science, Xiamen University, Xiamen, China, People's Republic of, 3Collaborative Innovation Center for Diagnosis and Treatment of Infectious Diseases, Zhejiang University, Hangzhou, China, People's Republic of

Synopsis

In-plane reduced field-of-view excitation based on two-dimensional radio-frequency pulse (2DRF) has been widely used in many applications. The EPI-style gradient waveform is commonly used in 2DRF implementation. However, at high field, the off-resonance effects during excitation would result in distortions of the profiles obtained. This work is to investigate the feasibility to achieve in-plane selective excitation of anatomically pre-defined regions using SPEN-2DRF pulse under different shim conditions. Experimental results show that the proposed method can produce profiles with significantly improved robustness to distortions at high field than the Fourier-based 2DRF pulse.

Purpose

In-plane reduced field-of-view (rFOV) excitation based on 2DRF is useful in many applications, such as to circumvent chest motion artifacts in cardiovascular MRI 1, to achieve localized B0 shimming with reduced SAR 2, to reduce the signal loss induced by intravoxel dephasing in T2*W fMRI 3, and to target the anatomically pre-defined ROI in single-voxel MR spectroscopy for minimized partial-volume effects 4. The EPI-style gradient waveform is commonly used in 2DRF implementation. However, at high field, the off-resonance effects during excitation cause severe distortions of the profiles obtained. Single-shot spatiotemporally encoded (SPEN) MRI is capable of maintaining the time efficiency of EPI but with significantly improved robustness to off-resonance induced distortions 5. Similar mechanism can be applied to 2DRF implementation to obtain excitation profiles of higher accuracy 6. In this work, we investigate the feasibility to achieve in-plane selective excitation of anatomically pre-defined regions using SPEN-2DRF pulse under different shim conditions.

Methods

In vivo rat brain imaging was performed on a 7T Varian animal scanner using two sequences depicted in Fig.1, in which the 2DRF pulses replace their 1D counterparts for exclusively exciting the brain regions. The 180° sinc pulses serve as slice selection and refocusing. The excitation and acquisition stages of both sequences are mirror symmetrical with identical gradient waveforms, magnitude and durations. The flowchart of the overall SPEN in-plane excitation scan is shown in Fig.2. Multi-shot GRE images were acquired for positioning and delineating the areas to be excited in subsequent rFOV single-shot scans. Compared with conventional method, the excited pattern of SPEN-2DRF is a profile with quadratic phase modulation. Partial Fourier reconstruction 5 was used to obtain the SPEN images. Scans were separately conducted after shimming and without shimming. The Fourier-based 2DRF and SPEN-2DRF pulses were both composed of 64 subpulses, and the latter had a time-bandwidth product of 256 along the SPEN dimension. Other parameters shared by the two sequences included FOV=60×60mm2 (axial and coronal), and 80×80mm2 (sagittal) with slice thickness=2mm, TR/TE=5000ms/34ms, durations for excitation and acquisition=26.1ms, acquisition matrix=64×64, and acquisition bandwidth=250kHz.

Results

Results of three orthogonal orientations were illustrated in Fig.3. For full FOV imaging, the images obtained with single-shot EPI were severely impaired by geometric distortions and local signal pileups due to shim inefficiency and susceptibility variations near tissue-air interfaces. In contrast, single-shot SPEN MRI provided images with significantly reduced deformation even without shimming. Correspondingly for rFOV imaging, the profiles obtained from spatiotemporally encoded 2DRF excitation and acquisition are of better geometric fidelity than those obtained from conventional method. Particularly, the shim conditions had a strong impact on the accuracy of the profiles obtained with conventional method, while the profiles obtained with SPEN method are largely immune to shimming conditions.

Discussion

In order to reduce the distortions caused by off-resonance effects during excitation and acquisition, the time-bandwidth product of the SPEN-2DRF pulse along the blipped (SPEN) axis should be several times larger than the number of subpulses, resulting in the periodic excitation of overlapped side-lobes. With small-tip-angle assumption, these excitation replicas share the same intensity and quadratic modulation, but differ by a linear phase. The undersampled acquisition along SPEN dimension also results in periodic aliases separated by the same length ∆L as the excitation replicas. With partial Fourier reconstruction, the value of a voxel is the sum of a series of overlapped side-lobes. Figure 4 shows the distribution of weights of all side-lobes, which can be divided into three levels. The second level is 10 times lower than the highest and the third is even smaller. Therefore, except for the four main lobes(5 for the center voxels only), the influence of other side-lobes can be neglected. Because the size of the rat head is small, only regions at the edge of the brain is overlapped by the nearest side-lobe in the results presented in Fig.3. For a larger object, such as the human brain, the signals of different lobes should be decomposed by further processing. Unlike conventional 2DRF excitation, these overlapped signals originate from different voxels separated by ∆L, providing a possibility to decompose them with parallel acquisition and reconstruction, and also the potential for applying this technology to clinical scanners not equipped with multi-channel transmission coils. Additionally, because the SPEN-2DRF pulse usually has a much larger duty cycle than Fourier-based 2DRF, the peak power needed is smaller.

Conclusion

A novel spatiotemporally encoded in-plane selective excitation and acquisition method is proposed in this work. Experimental results of in vivo rat shows that it can produce profiles less susceptible to off-resonance induced distortions at high field.

Acknowledgements

No acknowledgement found.

References

1. Abd-Elmoniem K Z, et al. MAGN RESON MED. 2012; 68:822-829.

2. Reynaud O, et al. MAGN RESON MED. 2015; 73:901-908.

3. Stenger V A, et al. MAGN RESON MED. 2000; 44:525–531.

4. Finsterbusch J, Busch M G. MAGN RESON MED. 2011; 66:333-340.

5. Chen Y, et al. MAGN RESON MED. 2013; 69:1326-1336.

6. Dumez J, Frydman L.J MAGN RESON. 2013; 226:22-34.

Figures

Fig.1. Diagram of the two sequences investigated in this study. (a) is the single-shot spin-echo EPI sequence with 2DRF excitation and (b) is the single-shot self-refocused SPEN sequence with hybrid 2DRF excitation. The excitation and acquisition stages of both sequences are mirror symmetrical with identical gradient waveforms, magnitude and durations.

Fig.2. Flowchart of the overall single-shot reduced field-of-view SPEN MRI scan with pre-defined excitation area.

Fig.3. In vivo rat imaging of 3 orthogonal orientations. The results respectively obtained under shimmed condition and without shimming are compared. 1st column: referential multi-shot GRE images; 2nd-6th columns: single-shot SE-EPI/SPEN images obtained with either 1D or 2DRF, Q represents time-bandwidth product of the SPEN-2DRF pulse.

Fig.4. Alias pattern of SPEN-2DRF excitation with pFT reconstruction. The reconstructed value of a voxel is the sum of a series of side-lobes shifted along SPEN dimension. The weights of different side-lobes are plotted, with j and l respectively representing the index of side-lobes due to undersampled excitation and acquisition.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1884