Ying Chen1, Song Chen1, Zhong Chen2, and Jianhui Zhong1,3
1Center for Brain Imaging Science and Technology, Zhejiang University, Hangzhou, China, People's Republic of, 2Department of Electronic Science, Xiamen University, Xiamen, China, People's Republic of, 3Collaborative Innovation Center for Diagnosis and Treatment of Infectious Diseases, Zhejiang University, Hangzhou, China, People's Republic of
Synopsis
In-plane
reduced field-of-view excitation based on two-dimensional radio-frequency pulse
(2DRF) has been widely used in many applications. The EPI-style gradient waveform is commonly used in 2DRF
implementation. However, at high field, the off-resonance effects during
excitation would result in distortions of the profiles obtained.
This work is to investigate the feasibility to achieve in-plane selective excitation
of anatomically pre-defined regions using SPEN-2DRF pulse under different
shim conditions.
Experimental results show that the proposed method can produce profiles with significantly improved robustness to distortions at high
field than the Fourier-based 2DRF pulse.Purpose
In-plane
reduced field-of-view (rFOV) excitation based on 2DRF is useful in many applications, such as to circumvent chest motion
artifacts in cardiovascular MRI
1, to achieve localized B
0 shimming with reduced SAR
2, to reduce the signal loss induced by intravoxel dephasing in T
2*W fMRI
3, and to target the anatomically
pre-defined ROI in single-voxel MR spectroscopy for minimized partial-volume
effects
4. The EPI-style gradient waveform is commonly used in 2DRF
implementation. However, at high field, the off-resonance effects during
excitation cause severe distortions of the profiles obtained. Single-shot
spatiotemporally encoded (SPEN) MRI is capable of maintaining
the time efficiency of EPI but with significantly improved robustness to off-resonance induced distortions
5. Similar mechanism can be applied to 2DRF
implementation to obtain excitation profiles of higher accuracy
6.
In this work, we investigate the feasibility to achieve in-plane selective excitation
of anatomically pre-defined regions using SPEN-2DRF pulse under different
shim conditions.
Methods
In vivo rat brain imaging was performed on a 7T Varian animal scanner using two sequences
depicted in Fig.1, in which the 2DRF
pulses replace their 1D counterparts for exclusively exciting the
brain regions. The 180° sinc pulses serve as slice selection and
refocusing. The excitation and acquisition stages of both sequences are mirror
symmetrical with identical gradient waveforms, magnitude and durations. The
flowchart of the overall SPEN in-plane excitation scan is shown in Fig.2. Multi-shot
GRE images were acquired for positioning and delineating the areas to be
excited in subsequent rFOV single-shot scans. Compared with conventional method,
the excited pattern of SPEN-2DRF is a profile with quadratic phase modulation.
Partial Fourier reconstruction
5 was used to obtain the SPEN images.
Scans were separately conducted after shimming and without shimming. The
Fourier-based 2DRF and SPEN-2DRF pulses were both composed of 64 subpulses, and
the latter had a time-bandwidth product of 256 along the SPEN dimension. Other
parameters shared by the two sequences included FOV=60×60mm
2 (axial
and coronal), and 80×80mm
2
(sagittal) with slice thickness=2mm, TR/TE=5000ms/34ms, durations for
excitation and acquisition=26.1ms, acquisition matrix=64×64, and acquisition
bandwidth=250kHz.
Results
Results of three orthogonal orientations were illustrated in Fig.3. For full
FOV imaging, the images obtained with single-shot EPI were severely impaired
by geometric distortions and local signal pileups due to shim inefficiency and susceptibility
variations near tissue-air interfaces. In contrast, single-shot SPEN MRI
provided images with significantly reduced deformation even without shimming. Correspondingly
for rFOV imaging, the profiles obtained from spatiotemporally encoded 2DRF
excitation and acquisition are of better geometric fidelity than those obtained
from conventional method. Particularly, the shim conditions had a strong impact
on the accuracy of the profiles obtained with conventional method, while the
profiles obtained with SPEN method are largely immune to shimming conditions.
Discussion
In
order to reduce the distortions caused by off-resonance effects during excitation and
acquisition, the time-bandwidth product of the SPEN-2DRF pulse along the
blipped (SPEN) axis should be several times larger than the number of subpulses,
resulting in the periodic excitation of overlapped side-lobes. With
small-tip-angle assumption, these excitation replicas share the same intensity
and quadratic modulation, but differ by a linear phase. The undersampled
acquisition along SPEN dimension also results in periodic aliases separated by
the same length ∆L as the excitation replicas. With partial Fourier reconstruction,
the value of a voxel is the sum of a series of overlapped
side-lobes. Figure 4 shows the distribution of weights of all side-lobes, which
can be divided into three levels. The second level is 10 times lower than the
highest and the third is even smaller. Therefore, except for the four main
lobes(5 for the center voxels only), the influence of
other side-lobes can be neglected. Because the size of the rat head is small,
only regions at the edge of the brain is overlapped by the nearest side-lobe in
the results presented in Fig.3. For a larger object,
such as the human brain, the signals of different lobes should be decomposed
by further processing. Unlike conventional 2DRF excitation, these overlapped signals
originate from different voxels separated by ∆L, providing a possibility to decompose
them with parallel acquisition and reconstruction, and also the
potential for applying this technology to clinical scanners not equipped with multi-channel
transmission coils. Additionally, because the SPEN-2DRF pulse usually has a much
larger duty cycle than Fourier-based 2DRF, the peak power needed is smaller.
Conclusion
A novel
spatiotemporally encoded in-plane selective excitation and acquisition method
is proposed in this work. Experimental results of in vivo rat shows that it can
produce profiles less susceptible to off-resonance induced distortions at high
field.
Acknowledgements
No acknowledgement found.References
1. Abd-Elmoniem K Z, et al. MAGN RESON MED. 2012; 68:822-829.
2. Reynaud O, et al. MAGN RESON MED. 2015; 73:901-908.
3. Stenger V A, et al. MAGN RESON MED. 2000; 44:525–531.
4. Finsterbusch J, Busch M G. MAGN RESON MED. 2011; 66:333-340.
5. Chen Y, et al. MAGN RESON MED. 2013; 69:1326-1336.
6. Dumez J, Frydman L.J MAGN RESON. 2013; 226:22-34.