Multi-Band Slice-Selective and 2D-Selective RF Excitations with Band-Specific Dephasing Moments for Tailored z-Shimming
Jürgen Finsterbusch1

1Systems Neuroscience, University Medical Center Hamburg-Eppendorf, Hamburg, Germany

Synopsis

T2*-weighted acquisitions used for functional neuroimaging can suffer from through-slice dephasing. Additional, so-called z-shim gradient pulses can be applied in the slice direction to minimize related signal losses. Thereby, a single, slice-specific gradient moment may be sufficient for small target regions like the spinal cord. To combine this z-shimming approach with multi-band acceleration, the individual bands must provide different dephasing moments in the slice direction. This can be realized with appropriate temporal shifts of the individual envelopes as is demonstrated for conventional slice-selective RF excitations and 2D-selective RF excitations for inner-field-of-view imaging.

Introduction

Multi-band acceleration (e.g. [1]) has become an important tool to speed up acquisitions, in particular in functional neuroimaging. However, the T2*-weighted acquisitions usually used suffer from through-slice dephasing and related signal dropouts. Additional, so-called z-shim gradient pulses can be applied in the slice direction to compensate the dephasing effect and minimize signal losses (e.g., [2]). For small target regions like the spinal cord, a single, slice-specific moment may already reduce signal drop outs significantly without affecting the temporal resolution [2]. Here, it is shown that with an appropriate temporal shift of the different bands individual dephasing moments can be realized for the different slices of a multi-band RF pulse as is demonstrated for slice-selective and 2D-selective RF excitations [3, 4].

Methods

Figure 1 sketches the basic principles of the current RF pulse design for slice-specific z-shim moments. For conventional single-band RF excitations, the individual z-shim moments can be realized by manipulating the slice rephasing gradient pulse accordingly (Fig. 1a). For multi-band RF pulses the moments can be realized by a temporal shift of the corresponding RF envelope which requires to prolong the slice-selection gradient pulse to avoid truncation artifacts (Fig. 1b). The application of this approach to 2D-selective RF excitations with a fly-back blipped-planar trajectory, e.g. for inner-field-of-view imaging (e.g. [5]), is straightforward (Fig. 1c).

Multi-band RF pulses were calculated for an acceleration factor of 3 as a superposition of phase-modulated RF envelopes with an appropriate temporal shift if required. The pulses were implemented in an FID echo-planar imaging pulse sequence. To avoid excessive peak power, multi-band pulses were applied with longer pulse duration, i.e. a lower slice-selection gradient pulse. For single-band RF pulses, a manipulation of the slice rephasing gradient was used to realize the desired z-shim gradient moment as shown in Fig.1a.

Measurements were performed using a 3 T whole-body MR system (Magnetom TIM Trio) with a 32-channel head coil (image resolution 2.0×2.0 mm2). A slice thickness of 2.0 mm was used for slice-selective RF excitations. For the 2D-selective RF excitations the excitation profile was 5.0×40 mm2 (slice thickness×field-of-view) which is in line with typical inner-field-of-view applications in the spinal cord [5] A water phantom and a cucumber cut into two pieces with an air-filled gap in-between to obtain susceptibility differences and related through-slice dephasing, were investigated.

Results

Figures 2 and 3 show images acquired without and with arbitrary, slice–specific z-shim gradient moments with single- and multi-band RF excitations that are slice-selective and 2D-selective, respectively. The signal intensities for the slices acquired with multi-band RF excitations are very similar to those obtained for the corresponding single-band acquisitions which demonstrates that (i) the multi-band RF pulse envelopes were calculated correctly and (ii) the desired, slice-specific z-shim gradient moments can be realized as for the conventional, single-band acquisitions.

The results obtained for the cucumber phantom exhibiting different degrees of through-slice dephasing are presented in Fig. 4 and 5 for slice-selective and 2D-selective RF excitations, respectively. Without z-shim gradient pulses the slices close to the gap between the two cucumber pieces suffer from significant signal loss. Applying a fixed, averaged z-shim moment to all slices of a multi-band excitation, increases the signal intensity in some of these slices but is at the expense of a signal loss in slices far away from the gap that do not require a z-shim moment but share the same multi-band excitation. The best signal intensities are obtained when an individual moment is applied to all slices in a multi-band excitation yielding significantly increased signal intensities close to the gap without affecting the other slices. Compared to the measurements without z-shim or with an averaged z-shim moment, intensities are increased by up to 70% and 40%, respectively, for the slice-selective RF excitations, and by up to 130% and 80%, respectively, for the 2D-selective RF excitations.

Discussion and Conclusion

For significant z-shim moments and related temporal shifts the multi-band RF excitation may be prolonged by several milliseconds which usually should not be crucial. However, in 2D-selective RF excitations multiple slice-selection gradient lobes are applied (see Fig. 1c). i.e. the prolongation may be come significant. To avoid excessive prolongations, the gradient amplitude of the outer k-space lines could be increased like in VERSE.

In conclusion, multi-band RF excitations can be designed to provide a slice specific z-shim gradient moment by applying a temporal shift of the corresponding RF envelope. Thus, application that benefit from slice-specific z-shimming like functional neuroimaging of the spinal cord, can be accelerated with multi-band imaging without additional signal loss.

Acknowledgements

No acknowledgement found.

References

[1] Setsompop K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn. Reson. Med. 2012; 67: 1210-1224.

[2] Finsterbusch J, Sprenger C, Büchel C. Single, slice-specific z-shim gradient pulses improve T2*-weighted imaging of the spinal cord. Neuroimage 2012; 59: 2307-2315.

[3] Bottomley PA, Hardy CJ. Two-dimensional spatially selective spipn inversion and spin-echo refocusing with a single nuclear magnetic resonance pulse. J. Appl. Phys. 1987; 62: 4284-4290.

[4] Pauly J, Nishimura D, Macovski A. A k-space analysis of small-tip-angle excitations. J. Magn. Reson. 1989; 81: 43-56.

[5] Finsterbusch J, Improving the performance of diffusion-weighted inner field-of-view echo-planar imaging based on 2D-selective radiofrequency excitations by tilting the excitation plane. J. Magn. Reson. Imaging 2012; 35: 984-992.

Figures

Fig. 1: Basic principles of realizing a slice-specific z-shim moment by (a) manipulating the slice rephasing of a single-band RF excitation and (b, c) temporally shifting the individual RF envelopes accordingly in a (b) slice-selective and (c) 2D-selective multi-band RF excitation.

Fig. 2: Comparison of images obtained using single- (left) and multi-band (three bands; right) slice-selective RF excitations without (upper) and with an arbitrary, slice-selective z-shim, i.e. dephasing, gradient moment (lower).

Fig. 3: Comparison of images obtained using single- (left) and multi-band (three bands; right) 2D-selective RF excitations without (upper) and with an arbitrary, slice-selective z-shim, i.e. dephasing, gradient moment (lower).

Fig. 4: Images of a cucumber with through-slice dephasing due to field inhomogeneities acquired with slice-selective, three-band RF excitations without z-shim (upper), with a fixed, averaged z-shim moment applied to all bands of an RF excitation (middle), and with individual, slice-specific z-shim moments for all bands (lower).

Fig. 5: Images of a cucumber with through-slice dephasing due to field inhomogeneities acquired with 2D-selective, three-band RF excitations without z-shim (upper), with a fixed, averaged z-shim moment applied to all bands of an RF excitation (middle), and with individual, slice-specific z-shim moments for all bands (lower).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1881