Gary George McGinley1,2, Atle Bjørnerud3,4, and Øystein Bech Gadmar3
1Institute for Experimental Medical Research, Oslo University Hospital, Oslo, Norway, 2KG Jebsen Cardiac Research Center and Center for Heart Failure Research, University of Oslo, Oslo, Norway, 3The Intervention Centre, Oslo University Hospital, Oslo, Norway, 4Department of Physics, University of Oslo, Oslo, Norway
Synopsis
This study aims to compare the effectiveness of three reverse-gradient method susceptibility
artefact correction tools (EPIC, TOPUP, and HySCO) in the correction of spin-echo (SE) and gradient-echo
(GE) EPI images of the brain, and to measure the effect of pixel bandwidth, SENSE factor
and slice thickness on artefact correction.
This was achieved by co-registering the artefacted and corrected images to an anatomical scan and measuring the normalised mutual information (NMI). It was found that EPIC correction resulted in the largest gains in NMI and that more mutual information was recovered at lower
pixel bandwidths after EPIC correction.Purpose
To
quantitatively assess and compare the effectiveness of three reverse-gradient
method susceptibility artefact correction tools in the correction of spin-echo (SE)
and gradient-echo (GE) EPI images of the brain acquired using a dual echo EPI
sequence. A secondary objective was to measure the effect of pixel bandwidth,
SENSE factor and slice thickness on artefact correction.
Methods
Five
healthy volunteers were scanned using a 3T Philips Achieva MRI system with an
8-channel SENSE head coil. A T1-weighted sagittal whole brain scan was
acquired, which was used as an anatomical reference for co-registration.
Multiple GE-EPI and SE-EPI images were acquired for each patient using a single-shot
dual-echo EPI sequence with the following parameters:
TR = 1500ms, TE(GE) = 35ms, TE(SE)=120ms, FOV = 192mm x
192mm, Echo Train length (ETL) = 62.
The
following parameters were varied:
Pixel
bandwidth: 636Hz, 893Hz, 1861Hz
Slice
thickness: 2mm, 4mm, 6mm
SENSE
factor: 1, 2 3
For
each set of parameters, the EPI sequence was run twice, with opposite polarity
of the phase-encoding gradient.
The
acquired images were then corrected using three correction tools; EPIC1, FSL TOPUP2, and SPM HySCO3.
These tools employ the reverse gradient method of susceptibility correction,
whereby two images acquired with opposite polarity of the phase-encoding
gradient (therefore having pixel displacements in opposite directions) are used
to calculate a pixel displacement map (Figure 1). This map can then be used to
correct image series acquired in the same way. The correction tools were used
with their default settings, as these were considered representative of typical
usage.
The
corrected images were then independently co-registered with the high resolution
T1 weighted scan and the maximal value of normalized mutual information (NMI)
was recorded4. NMI is a measure of geometric similarity between two
images, and was therefore used to determine the effectiveness of the correction
of the geometric distortions5. Mann-Whitney and Kruskal-Wallis
non-parametric tests were used to determine significant differences in NMI with
a confidence value of p<0.05.
Results
The
change in mutual information after correction was found to be significant only
for the EPIC correction method, according to the Mann-Whitney u-test (Figure 2).
We found that this was the case for both SE-EPI and GE-EPI images. The
difference in NMI gain using the EPIC correction tool on SE-EPI images with
varying bandwidth was found to be significant (Figure 3). It was generally
observed that susceptibility artefact correction reduced the variation in NMI
between images of different pixel bandwidth, slice thickness, and SENSE factor.
Discussion
The
EPIC correction tool achieved the most significant gains in mutual information
for both GE-EPI and SE-EPI images. Upon inspection, EPIC correction resulted in
images with fewer artefacts and noticeable improvements in geometric
distortions (Figure 4). The HySCO correction performed poorly in this study,
and did not satisfactorily reverse the geometric distortions. TOPUP corrected
the geometric distortions, but was found to introduce banding artefacts. It
should be noted that parameter optimization might improve the performance of
these tools; our experience with these settings suggests that this may be the
case, however due to the large number of free parameters, an investigation of
the effects of each of them was outwith the scope of this study. The effect of
pixel bandwidth on the image correction using EPIC was found to be significant
(Figure 5); however the effects of SENSE factor and slice thickness on the correction
quality were less apparent. Generally, these tools were equally effective in
the correction of geometric distortions in both GE-EPI and SE-EPI images.
Conclusion
EPIC
was found to be the most effective of the three correction tools in the present
study, both in terms of the gain in NMI and with respect to the reduction of
visible artefacts in the correction of both GE-EPI and SE-EPI images. Furthermore,
more mutual information was recovered at lower pixel bandwidths, reducing the
impact of acquisition bandwidth on the geometric accuracy of SE-EPI images post
correction.
Acknowledgements
This project was carried out in collaboration with the Intervention Centre at Oslo University Hospital and the University of Oslo.I would like to thank my supervisors, Atle Bjørnerud and Øystein Bech Gadmar for their help and guidance. In addition, I would like to thank Oliver Marcel Geier, Kyrre Eeg Emblem and Ingrid Digernes for providing me with additional guidance and imaging data.References
1.
Holland D, Kuperman JM, Dale AM. Efficient Correction of
Inhomogeneous Static Magnetic Field-Induced Distortion in Echo Planar Imaging. NeuroImage.
2010;50(1):175. doi:10.1016/j.neuroimage.2009.11.044
2.
Smith, S. M. et al. Advances in
functional and structural MR image analysis and implementation as FSL. Neuroimage
23 Suppl 1, S208–19 (2004).
3.
Ruthotto et al. (2013)
HySCO - Hyperelastic Susceptibility Artifact Correction of DTI in SPM
4.
SPM, By members &
collaborators of the Wellcome Trust Centre for Neuroimaging, http://www.fil.ion.ucl.ac.uk/spm/
5.
Fritz, L. et al. (2014)
Comparison of EPI distortion correction methods at 3T and 7T