Synopsis
One of the motivations for 4D MRI is the need to
characterize patient respiratory motion in the context of radiation therapy treatment
planning (RTP). Here, we compare two
approaches for generating 4D MRI data using dynamic, navigator-based
acquisitions with retrospective respiratory compensation, including single-shot
balanced-SSFP and single-shot fast spin echo with variable refocusing flip angle (vrfSSFSE). The results presented suggest that both
sequences offer a straightforward approach to generating 4D MRI data for
MR-guided RTP.Purpose
MRI offers a number of well-documented
advantages over CT, including superior soft-tissue contrast and the lack of
ionizing radiation. The need to
characterize patient respiratory motion is an important aspect of radiation
therapy treatment planning (RTP), where MRI has the potential to exceed the
performance of 4D CT, the current standard of care. Possible advantages include target
identification, depiction of organs-at-risk, the ability to image for longer
periods of time and/or more frequently, and ultimately, better outcomes for
patients.
1 One approach to 4D
MRI (vs. 4D CT), is to generate a time-resolved 3D volume via a multi-slice,
multi-phase MR acquisition, where the images are retrospectively sorted by
respiratory phase.
2 Here, we
compare two dynamic, multi-slice pulse sequences, each with navigator-based
retrospective respiratory compensation.
Methods
A cylindrical navigator pulse was added to both
single-shot balanced-SSFP and single-shot fast spin echo (aka SSFSE, or HASTE, SSH-TSE, etc.) sequences capable of multi-phase imaging. Additionally, the SSFSE pulse sequence
incorporated variable refocusing flip angle (vrf), useful in this context for
reduced repetition times (due to lower SAR), and therefore higher frame
rates. Phantom and volunteer scanning
was performed on a wide-bore 3.0T MRI scanner (MR750w, GE Healthcare, Waukesha,
WI). Following informed consent, volunteer
images were acquired for both single-shot sequences. Imaging was performed in the coronal plane
for 20, 5-mm slices, 20 phases per slice, FOV = 40 x 36 cm, 2x parallel imaging
acceleration (ARC, GE Healthcare, Waukesha, WI). Additional imaging parameters for the
single-shot bSSFP acquisition were TR/TE = 3.7/1.65 ms, frames per second = 1.57,
35°
flip angle, 224 x 256 matrix, ±125 kHz bandwidth, and 1.0 NEX. Additional imaging parameters for the vrfSSFSE
acquisition were TR/TE = 642/80 ms, frames per second = 1.56, 256 x 224 matrix,
±125
kHz bandwidth, 0.64 effective NEX, and refocusing flip angle targets
3
of 130, 90, 100 and 45 degrees, respectively.
Total (arbitrary) acquisition time for both sequences was approximately 4
minutes and 15 seconds for 400 total images.
The navigator profiles were used retrospectively to measure the position
of the liver dome between imaging shots, which was then used to sort the images
retrospectively, based on slice location and respiratory phase.
Results
Figures 1 and 2 are composite images, showing
all acquired phases for a single slice location, retrospectively sorted by
respiratory phase, for both the single-shot bSSFP and vrfSSFSE in vivo
acquisitions. The bSSFP images (Figure
1) demonstrate high signal-to-noise, bright-fluid T2/T1-weighted contrast,
robustness to motion, and expected off-resonance effects, including banding and
fat-water cancelation. The vrfSSFSE images (Figure 2) demonstrate T2-weighted
contrast, robustness to off-resonance, a modest (intrinsic) black-blood effect
due to the flip angle modulation, and lower relative SNR. As intended, in both cases, the respiratory
phase is successfully represented by a monotonic displacement of the diaphragm from end
inspiration to end expiration. Pitfalls
of both sequences are illustrated in Figure 3, showing typical off-resonance
effects encountered with bSSFP and shading due to cardiac motion encountered
with vrfSSFSE.
4Discussion
In spite of their differences, both single-shot
acquisitions demonstrate comparable effectiveness in their ability to
thoroughly sample the respiratory cycle and to satisfy the basic requirements for
respiratory-guided 4D MR without the need for respiratory bellows or associated
reliability concerns. Balanced SSFP is
already commonly used in multi-phase imaging, generally prized because of its
steady state properties. In this case,
the reduced SAR of the vrfSSFSE sequence, however, allows it to match the
relatively high frame rate of the bSSFP acquisition (of approximately 1.5 fps).
5 Additional strengths of the vrfSSFSE
acquisition are its familiar T2-weighted contrast, and robustness to off-resonance
effects, which may translate to a relative advantage in terms of target and
at-risk organ delineation over large fields of view. Relative to CT, both sequences exhibit superior
soft-tissue contrast, and the total (arbitrary) acquisition time of
approximately 4 minutes, represents an unacceptable exposure time for CT. This long scan time may translate into a
valuable advantage for MR, via a more thorough characterization of patient
respiratory motion over time than CT can safely achieve. Finally, image artifacts common to both
sequences remain a concern, though they may be mitigated to some extent with
well-established techniques such as phase cycling, cardiac gating, and/or retrospective
rejection of corrupted images.
Conclusion
We believe these single-shot sequences offer a
straightforward approach to generating 4D MRI data for MR-guided RTP. In spite of their differences, we believe
both sequences offer sufficient soft-tissue contrast, with each exhibiting its
own set of unique characteristics. Future
work will include the evaluation of these techniques in a clinical radiation
therapy treatment planning setting.
Acknowledgements
No acknowledgement found.References
1. McGee KP, Hu Y, Tryggestad
E, Brinkmann D, Witte B, Welker K, Panda A, Haddock M, Bernstein MA. MRI in
radiation oncology: Underserved needs. Magn Reson Med. 2015 Jul 14.
2. Tryggestad E, Flammang A,
Han-Oh S, Hales R, Herman J, McNutt T, Roland T, Shea SM, Wong J.
Respiration-based sorting of dynamic MRI to derive representative 4D-MRI for
radiotherapy planning. Med Phys. 2013 May;40(5).
3. Busse RF, Hariharan H, Vu
A, Brittain JH. Fast spin echo sequences with very long echo trains: design of
variable refocusing flip angle schedules and generation of clinical T2
contrast. Magn Reson Med 2006 May;55(5):1030-7.
4. Litwiller DV, Holmes JH,
Saranathan M, Loening AM, Glockner JF, Vasanawala SS, Bayram E. Sensitivity of
modulated refocusing flip angle single-shot fast spin echo to impulsive
cardiac-like motion. Proceedings of the ISMRM 2014.
5. Loening AM, Saranathan M, Ruangwattanapaisarn
N, Litwiller DV, Shimakawa A, Vasanawala SS. Increased speed and image quality
in single-shot fast spin echo imaging via variable refocusing flip angles. J
Magn Reson Imaging. 2015 Jun 19.