Claus Benjaminsen1, Rasmus Ramsbøl Jensen1, Paul Wighton2, M. Dylan Tisdall2, Helle Hjorth Johannesen3, Ian Law3, Andre J. W. van der Kouwe2, and Oline Vinter Olesen1
1DTU Compute, Technical University of Denmark, Lyngby, Denmark, 2Athinoula. A. Martinos Center for Biomedical Imaging, Dept. of Radiology, Massachusetts General Hospital, Boston, MA, United States, 3Department of Clinical Physiology, Nuclear Medicine & PET, Rigshospitalet, University of Copenhagen, Copenhagen, Denmark
Synopsis
Prospective
motion correction for MRI neuroimaging has been demonstrated using MR
navigators and external tracking systems using markers. The drawbacks of these
two motion estimation methods include prolonged scan time plus lack of compatibility
with all image acquisitions, and difficulties validating marker attachment resulting
in uncertain estimation of the brain motion respectively. We have developed a
markerless tracking system, and in this work we demonstrate the use of our
system for prospective motion correction, and show that despite being
computationally demanding, markerless tracking can be implemented for real time
motion correction.Purpose
Prospective
motion correction for MRI neuroimaging has become increasingly prevalent in
practice, with a variety of techniques and hardware being actively developed1.
Reliable motion estimation is crucial for prospective correction, since incorrect
estimation of the motion will degrade the image.
Tracking
can be accomplished using MR navigators or external devices. MR navigators
often prolong the scan time, lack simultaneous 3D spatial and temporal
resolution, and are not compatible with all image acquisitions. External tracking systems using markers have
difficulties validating marker attachment and as a result the estimated brain
motion is uncertain. Furthermore, these methods require extra patient
preparation for attaching the markers, which also makes them inconvenient for
clinical use.
In the
search for a clinically feasible method, we have developed a markerless
tracking system, and previously demonstrated it for retrospective correction2.
In this work we demonstrate prospective motion-correction in a 3D FLASH sequence
using our system. Markerless tracking is non-trivial and can be computationally
demanding, but here we demonstrate that it can be successfully implemented for real-time
prospective correction in MRI.
Methods
Our
tracking system was set up on a hybrid 3T mMR Biograph, Siemens3. The
control unit of our system was connected to the internal scanner Ethernet
network. A healthy volunteer was scanned using a custom 3D FLASH sequence, modified
to query our control unit for the current patient position4. Changes
in orientation and shifts in position were corrected for in real time by dynamically
adjusting the gradients, RF pulse frequencies, and k-space sample phase every
TR, thus generating a corrected image on the scanner upon completion of the
scan.
One
volunteer, having given informed consent, was instructed to perform a
repeatable motion pattern or remain motionless while being scanned with and
without motion correction enabled. The parameters of the 3D FLASH sequence were:
matrix 128x128, pixel spacing 1.72 mm, 56 slices, slices thickness 3 mm, TR 30 ms,
TE 2.2 ms, and flip angle 35 degrees. A queue lag of pose estimates of 4x TR was
maintained by the scanner to prevent correction queue underflow. This lag prolonged
the latency between performed motion and motion update in addition to the pose
estimation and communication latency.
A standard
MPRAGE image volume (512x512x192, pixel spacing 0.49 mm, and slice thickness 1
mm) was used to determine the geometric alignment between the tracking system
and the scanner coordinate system. A rigid transformation was fit iteratively between
a surface of the volunteer’s face obtained from our tracking system, and the
surface of the volunteer’s head extracted from the MPRAGE volume. The alignment
transformation was subsequently used to transform the volunteer’s position into
the scanner coordinate system. No further calibration of the system was
required.
Results
Figure 1
shows 3D point clouds of the volunteer inside the MR head coil. These constitute
some of the outer positions, corresponding to the maximal motion shown in Fig. 2.
Figure 2
shows the motion recorded during the image acquisitions with/without motion
correction (moco) and the reference scan with no motion. The curves represent
the motion of the centroid of the 3D point cloud.
It is noted that the motion pattern is similar between the scans without and
with motion correction. All of the plots have ripples corresponding to
respiratory motion with amplitude of 0.2 to 0.5 mm (Fig. 3). This indicates a
tracking accuracy substantially better than the general image resolution. Figure
4 shows an image slice from each of the three FLASH volumes. A much better
result is obtained with motion correction than without. The slice with motion
correction is comparable in quality to the slice with no motion.
Discussion
The only
correction applied was adaptively updating the imaging coordinates according to
the tracked motion. Further steps such as B0 and B1 correction were not
included, and these can be expected to improve image quality further,
especially for large motions. The tested motions are relatively large
demonstrating that this system is capable of correcting larger-than-usual
ranges of motion.
Determining
a position quickly and accurately are key challenges in tracking for
prospective motion correction. Generally it is possible to trade speed for
accuracy and vice versa. The presented results show that it is possible to find
a good tradeoff between these two objectives with further optimization still
possible.
Conclusion
We have
shown that prospective motion correction can be implemented using markerless
tracking. Markerless tracking has great potential for clinical routine as
failures from maker attachment are eliminated, it does not introduce any extra
patient discomfort or preparation time, and it also does not prolong the scan
time.
Acknowledgements
This work
was supported by Novo Nordisk Foundation and Arvid Nilssons Foundation.References
1. Maclaren
J, Herbst M, Speck O. "Prospective motion correction in brain imaging: a
review" Magnetic Resonance in Medicine 69.3 (2013): p. 621-636.
2. Jensen RR, Benjaminsen C, Hansen AE, Larsen R, Olesen
OV. Markerless
Motion Correction in MRI. Joint Annual
Meeting ISMRM-ESMRMB and SMRT 23rd Annual Meeting 2015. p. 1205.
3. Olesen OV, Wilm J,
van der Kouwe AJ, Jensen RR, Larsen R, Wald LL. An MRI Compatible Surface
Scanner. Joint Annual Meeting
ISMRM-ESMRMB and SMRT 22nd Annual Meeting 2014. p. 1303.
4. Wighton
P, Tisdall MD, Bhat H, Nevo E, van der Kouwe AJ. Slice-by-slice prospective
motion correction in EPI sequences. International Society for Magnetic
Resonance in Medicine's (ISMRM) Workshop on Motion Correction in MRI 2014,
Tromsø, Norway.