Auto-Calibrating Wave-CS for Motion-Robust Accelerated MRI
Feiyu Chen1, Tao Zhang1,2, Joseph Y. Cheng1,2, John M. Pauly1, and Shreyas S. Vasanawala2

1Electrical Engineering, Stanford University, Stanford, CA, United States, 2Radiology, Stanford University, Stanford, CA, United States

Synopsis

In this work, we propose a motion-robust auto-calibrating Wave-CS technique. This technique uses the wave-encoded center k-space and the known point-spread-function (PSF) of wave-encoding to reconstruct a Cartesian central k-space for calibration. The coil sensitivity maps are subsequently estimated with ESPIRiT for CS-SENSE reconstruction of under-sampled k-spaces. Results show this approach can reduce the motion artifacts and the aliasing artifacts due to sensitivity variations between the calibration and accelerated wave-encoded acquisitions.

Purpose

Wave-encoding1 has been demonstrated to improve the performance of accelerated MRI reconstruction by distributing the under-sampling in both phase-encode and readout directions. Wave-CS2 combines wave-encoding with compressed sensing and has been shown to significantly reduce aliasing artifacts in brain imaging. However, a separate calibration scan is required. Subject motion between the calibration scan and the wave-encoded acquisition may introduce sensitivity variation and therefore degrade image quality. In this work, we propose a motion-robust auto-calibrating Wave-CS technique for abdominal and pelvic imaging.

Methods

The concept is to use the wave-encoded center k-space and the known point-spread-function (PSF) of wave-encoding to reconstruct a Cartesian central k-space for calibration. In detail, the wave-modulated signal $$$S_{wave}$$$ and the Cartesian signal $$$S_{Cartesian}$$$ have the following relationship in the kx-y-z domain1:$$S_{wave}[k_x,y,z]=PSF[k_x,y,z]\cdot S_{Cartesian}[k_x,y,z]$$The PSF can be expressed as: $$\begin{aligned} PSF[k_x,y,z] &=\exp(-i\gamma\int g_y(\tau)y+g_z(\tau)z d\tau) \\ &=\exp(C_1\cdot y+C_2\cdot z) \\ &=\exp(C_1\cdot \Delta y \cdot i_y+C_2\cdot \Delta z \cdot i_z)\end{aligned} $$ where $$$C_1$$$ and $$$C_2$$$ are constants associated with gradients $$$g_y$$$ and $$$g_z$$$, $$$y$$$ and $$$z$$$ are positions in image space, $$$\Delta y$$$ and $$$\Delta z$$$ are the corresponding spatial resolutions, and $$$i_y$$$ and $$$i_z$$$ are spatial indices. The wave-encoded central k-space can be treated as an independent low-resolution k-space modulated by the same PSF with lower resolution. Thus, by replacing $$$\Delta y$$$ and $$$\Delta z$$$ with $$$\Delta y_{calib} = \frac{N_y}{N_{y, calib}}\cdot \Delta y$$$ and $$$\Delta z_{calib} = \frac{N_z}{N_{z, calib}}\cdot \Delta z$$$, where $$$N_y$$$ and $$$N_z$$$ are the acquisition matrix size, $$$N_{y, calib}$$$ and $$$N_{z, calib}$$$ are the calibration matrix size, and $$$\Delta y_{calib}$$$ and $$$\Delta z_{calib}$$$ are the spatial resolution of the calibration k-space, we can obtain a set of low-resolution PSF (Fig.1b), and subsequently reconstruct the Cartesian central k-space using inverse Fourier transform of the PSF. The resulted Cartesian center can then be used for calibration.

To test the accuracy of this approach, five phantom experiments with both fully sampled Cartesian acquisition and wave-encoded acquisition were conducted. The normalized root-mean-square-error (RMSE) between low-resolution images from the calibration data in Cartesian acquisition and those reconstructed by the proposed method were calculated as an indicator of calibration accuracy (Fig. 2). Two volunteer scans (Figs. 3 and 4) were acquired on a 3T GE MR750 scanner (GE Healthcare, Waukesha, WI) with wave encoding (3 cycles of sinusoids, 4mT/m amplitude) using a 32-channel cardiac coil (Invivo Corp., Gainesville, FL) and a 32-channel torso array coil (NeoCoil, Pewaukee, WI), respectively. In the second scan, spatial selective excitations in frequency-encoding directions were used to maintain a reasonable over-sampling factor of ~1.5. Down sampling using a VDRad trajectory3 was simulated (Fig. 3) and implemented (Fig. 4) in a 3D SPGR sequence with a 16×16 calibration region at a reduction factor of 5.1 and 6.2, respectively. Sensitivity maps were estimated using ESPIRiT4 from Cartesian acquisition directly and from wave-encoded acquisition with the proposed method. CS-SENSE reconstruction using the same sampling pattern was implemented for the auto-calibrating method and compared with the conventional Wave-CS method. Acquisition parameters for the 3D SPGR sequence were: TR/TE 12/2.2ms, FA 15°, and BW $$$\pm$$$125kHz. Acquisition matrices were 308(kx)×128(ky)×128(kz) (with partial readout factor 0.6) for FOV 420mm (R/L)×210mm (A/P)×210mm (S/I), and 256(kx)×256(ky)×64(kz) for FOV 400mm (S/I)×400mm (R/L)×256mm (A/P).

Results

Normalized RMSE for low-resolution calibrating images decreases with increased size of calibration for the auto-calibrating method (Fig. 2). Artifacts due to motion and under-sampling occur in both reconstructions (Fig. 3). Auto-calibrating Wave-CS reduces the artifacts in both free breathing and breath-holding cases, in comparison with conventional Wave-CS approach (Figs. 3 and 4).

Discussion

To maintain a low RMSE of less than 2%, a calibration size of ~20×20 is necessary. Using larger calibration size would be helpful until the calibration size reaches 40×40. When the calibration size is larger than 40×40, the non-zero RMSE is mainly caused by inaccurate PSF estimation of wave encoding. In comparison with conventional Wave-CS, the proposed auto-calibration approach can reduce the motion artifacts and the aliasing artifacts due to sensitivity variations between the calibration and the wave-encoded acquisitions, as pointed by yellow arrows in Figs. 3 and 4. As the acquisition parameters were designed for post-contrast T1 weighted acquisitions, the image quality may improve in clinical contrast-enhanced scans.

Conclusion

In this work, we evaluated the feasibility of auto-calibrating Wave-CS for potential applications with subject motion. Applying this method in dynamic contrast enhanced abdominal imaging will be our future work.

Acknowledgements

We gratefully acknowledge the support of the Tashia and John Morgridge Faculty Scholars Fund, NIH/NIBIB R01 EB009690, R01 EB019241, and GE Healthcare.

References

1. Bilgic B, et al. Wave-CAIPI for highly accelerated 3D imaging. Magnetic Resonance in Medicine. 2015, 73(6): 2152-2162.

2. Curtis A, et al. Wave-CS: Combining wave encoding and compressed sensing. Proc. Intl. Soc. Mag. Reson. Med. 2015, 23: 0082.

3. Cheng JY, et al. Free-breathing pediatric MRI with nonrigid motion correction and acceleration, Journal of Magnetic Resonance Imaging. 2014, 42(2): 407-420.

4. Uecker M, et al. ESPIRiT-an eigenvalue approach to autocalibrating parallel MRI: where SENSE meets GRAPPA. Magnetic Resonance in Medicine. 2014, 71(3): 990-1001.

Figures

Fig. 1 (a) Phase of the full-size PSF displayed in the y-z plane in [-π, π] for kx = 2, 78, 154, 230, and 306; (b) Phase of the corresponding low-resolution PSF for the 16(y)×16(z) calibration region.

Fig. 2 Normalized RMSE for low-resolution calibrating data decreases while size of calibration kspace increases, as shown in: (a) simulated wave-encoded kspace with an accurate PSF, and (b) five phantom experiments with estimated PSFs.

Fig. 3 Comparison of reconstructed images using full acquisition (a), Wave-CS (b), and the auto-calibrating Wave-CS (c) during free breathing scans at a reduction factor of 5.1. 1.5x zoomed-in images are shown in the second row. Yellow arrows point the major difference between these reconstructions. The VDRad sampling mask with a 16×16 calibration region is shown in (d).

Fig. 4 Comparison of reconstructed images using conventional Wave-CS (a) and auto-calibrating Wave-CS (b) during breath-held scans at a reduction factor of 6.2. Conventional Wave-CS (a) uses a separated breath-held calibration scan to achieve the coil sensitivity maps. 1.5x zoomed-in images are shown at the bottom right corners. Yellow arrows point the major difference between these reconstructions. The VDRad sampling mask with a 16×16 calibration region is shown in (c).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1857