Hirofumi Hata1, Yusuke Inoue2, Ai Nakajima1, Shotaro Komi1, Yutaka Abe1, Keiji Matsunaga2, and Hiroki Miyatake1
1Department of Radiology, Kitasato University Hospital, Sagamihara, Japan, 2Department of Diagnositic Radiology, Kitasato University School of Medicine, Sagamihara, Japan
Synopsis
We compared navigator-gating techniques
for free-breathing 2D SPGR images of the liver using pencil-beam excitation and
self-navigation techniques in 3 T MRI. In
pencil-beam navigator, single-check (PB-SC) and double-check (PB-DC) modes were
examined. In self-navigator scans, self-navigator signals were acquired in two
fashions; before (SN-Pre) or after (SN-Post) the imaging read-out. Visual
analysis shows that respiratory waveforms fluctuated in SN-Post. Quantitative
and qualitative image evaluations show that PB-DC and SN-Post had better image
qualities than the others. Considering scan time was about doubled in SN-Post, PB-DC
should be the best for respiratory navigation in 2D SPGR imaging at this stage.Purpose
A T1-weighted 2D spoiled gradient-recalled acquisition in the steady state (2D SPGR)
image is one of the basic contrast images, and commonly acquired in abdominal
MRI. In abdominal MRI, a respiratory navigation technique is widely used to
reduce respiratory motion artifacts when a patient has difficulty in breath-holding
for an adequate duration. The purpose of this study was to compare
navigator-gating techniques for free-breathing 2D SPGR images of the liver
using pencil-beam (PB) excitation1 and self-navigation2 (SN)
techniques in 3 T MRI.
Methods
MR imaging: Twelve healthy volunteers underwent axial 2D
SPGR MRI of the liver using PB navigator and SN techniques. Figure 1 outlines
the four navigator techniques used in this study. In PB navigator, single-check
(PB-SC) and double-check (PB-DC) modes were examined; the former accepts the
imaging data when the following navigator value falls within the acceptance
window,3 whereas the latter accepts only when both the precedent and
following values are in the window. Self-navigator signals were acquired at the
k-space origin and signals from the superior half of the imaging slices were
combined for reducing the spin saturation effects,4 and
self-navigator echo was acquired before or after the imaging read-out. We refer
to these two self-navigation acquisition scheme as SN-Pre and SN-Post,
respectively. Free-breathing (FB) and breath-held (BH) 2D SPGR scans were also
performed without any respiratory navigation.
All
scans were performed on a 3 T clinical scanner (Discovery
750w; GE Healthcare, Waukesha, WI) with floating anterior and fixed
posterior coil arrays. Sequence
parameters included: parallel imaging using ASSET with an acceleration factor
of 2.5, TR = 245 ms, TEs (SN-Pre) = 2.3/5.8 ms, TEs (others)
= 1.1/2.3 ms, rBW = ±83.3 (SN-Pre)/±166.7
(others) kHz, FA = 60°, FOV = 36 cm, slice thickness/gap = 6.0/1.5 mm, matrix = 288 ×192, number of slices = 22, NEX = 1.
Waveform/Image
analysis: Visual waveform quality was graded using a three-point scoring
system: 1 (poor) = respiratory movement not apparent, 2 (fair) = apparent
respiratory movement with substantial fluctuation, 3 (good) = smooth and well
representative of respiratory movement. Image analysis was performed quantitatively
and qualitatively. The ghost signals in the SPGR images were measured using a rectangular
ROI (20 mm × 50 mm) placed anteriorly outside
the body. The final ghost levels were calculated with the averaged ghost
signals from the upper, middle, lower slices divided by the liver signals in
the middle slices. Visual motion artifacts were graded using a four-point
scoring system: 1 (poor) = unacceptable for clinical use, 2 (fair) = distorted
diagnostic capability, 3 (good) = evident artifacts without distorting
diagnostic capability, 4 (excellent) = negligible artifacts. Statistical
analysis was conducted by the Friedman test, followed by the Wilcoxon signed-rank test with
corrections for multiple comparisons.
Results
Table 1 shows the scan time and the waveform analysis results.
SN scans took longer than PB; especially the scan time of SN-Post was almost
doubled compared to those of PB methods. The PB techniques generated better
waveform quality than the self-navigator, and SN-Pre was better than SN-Post. Ghost
levels of all the navigator scores were higher than BH and lower than FB (Table
2). Among navigator techniques, PB-DC and SN-Post scores were lower than PB-SC
and SN-Pre in several conditions. Visual analysis showed that all the navigator
scores were significantly lower than BH and higher than FB (Table 3). Among
navigator techniques, PB-DC and SN-Post scores were significantly better than
PB-SC and SN-Pre in several comparisons. Representative images were shown in
Fig. 2.
Discussion and Conclusion
PB generated better waveform quality than SN, showing
that the PB navigator reflected pure respiratory motion whereas the SN signals are
influenced by cardiac motion and spin saturation effects.2 In
comparison between the two self-navigator techniques, SN-Post was worse than
the SN-Pre regarding waveform quality, though ghost level and image quality
scores were better. One possible reason may be larger cardiac motion effects are
involved in SN-Post signals, causing fluctuations in waveforms and consequently
allowing both monitoring of respiratory and cardiac motions. Decomposition of
respiratory and cardiac motions can improve scan efficiency and/or image
quality. Among navigator techniques, PB-DC and SN-Post had better image quality
than the others, and PB-DC should be the best considering scan time among the
techniques tested in this study. Further investigation should be conducted in the
future to reduce the difference between navigator and BH techniques.
Acknowledgements
We
thank Yuji Iwadate for his technical support.References
1. Pauly
J, Nishimura D, et al. A k-space analysis of small-tip-angle excitation. J Magn Reson. 1989;81:43–56.
2.
Brau AC, Brittain JH. Generalized self-navigated motion
detection technique: Preliminary investigation in abdominal imaging.
Magn Reson Med. 2006;55(2):263-270.
3.
Yoon JH, Lee
JM, et al. Navigated three-dimensional T1-weighted
gradient-echo sequence for gadoxetic acid liver magnetic resonance imaging in
patients with limited breath-holding capacity. Abdom Imaging. 2015; 40(2):278–288.
4.
Iwadate Y, Brau AC, et al. In:
Proceedings of the 22th Annual Meeting of ISMRM, Milan, Italy, 2014. (abstract
4352).