Synopsis
A
turbo spin echo based sequence for robust and fast diffusion imaging is
proposed. It overcomes the non-CPMG problem by split-echo acquisition of turbo
spin echo signals. EPI-like readout is used to sample the separated echoes and generate
multiple blades for a single k-space. Each blade was corrected for both the
inherent phase of separated echoes and off-resonance phase, to avoid the
destructive inference. With this technique, the non-CPMG problem can be
effectively mitigated at low flip angle refocusing pulses to reduce SAR.
Moreover, the off-resonance artifacts can also be reduced especially when high
acceleration factor is applied.Purpose
Diffusion imaging has been widely used in clinical applications. By far
single-shot EPI is the most commonly used sequence due to its high speed and
insensitivity to bulk motion and physiologic movement. However, this technique
suffers from image blurring caused by signal decay during the long acquisition
window and static field related problems, e.g., geometric distortion and
susceptibility artifact caused by B0 inhomogeneity. Comparing to EPI based
methods, diffusion weighted (DW)-PROPELLER sequence, which is based on Turbo
Spin Echo (TSE) sequence and PROPELLER(BLADE) trajectory, has the
advantage of reduced sensitivity to B0 inhomogeneity and T2 decay induced image
blurring [1]. However the application of this sequence is limited by its
relatively long scan time, high SAR and violation of CPMG condition. X-PROP [2] with multi-blade k-space filling
strategy, was developed to speed up the acquisition and reduce SAR. In this
technique the non-CPMG problem is mitigated by phase cycling the refocusing RF
pulses, which requires the flip angle of refocusing pulses as close to 180° as
possible. Therefore, SAR is still high especially at high fields and B1 field
variation may result in unstable image quality. This technique also suffers
from off-resonance artifacts when large acceleration factor is applied, due to
the contribution from more gradient echoes. The SPLICE (split acquisition of
fast spin-echo signals) technique can address the non-CPMG problem reliably by
separating two signal components to avoid destructive phase interference[3]. Therefore,
we propose a novel technique by sampling two separated echoes with multi-blade
(X-PROP) strategy and placing the blades to a single k-space dataset, and
demonstrate its application for fast and robust DWI. With this technique, the
non-CPMG problem can be effectively mitigated with low-flip-angle refocusing
pulses to reduce SAR. Moreover, the off-resonance artifacts can also be reduced
with high acceleration factor.
Methods
The diagram of the proposed sequence is shown in figure 1. Note that
time (t) starts at the end of diffusion preparation (not shown). At this point the
spins lie in the transversal plane with spatially variant and unknown phase,
which violates the CPMG condition. The signal components with different phases
are split into two groups (E1 and E2), as the first spin echo is refocused
asymmetrically with regard to the interval between consecutive refocusing
pulses. Multi-blade(X-PROP) acquisition strategy is applied to sampling these
echoes. EPI-like readout gradients and blips are generated on both readout and
phase encoding directions, producing radial-like k-space lines which are placed
into separate blades. The readout is repeated throughout the whole echo train,
producing all k-space lines for each blade. These blades spans in one single
k-space, as shown in figure 2. The echoes from E1 and E2 group have different
phases, which are removed by the blade-by-blade phase correction in the
following image reconstruction step.
A phase insensitive preparation is also
applied to push E1 and E2 to reach steady state as soon as possible by applying
a dephasing gradient right after the end of diffusion preparation [4]. Additional rephrasing and dephasing gradients
must be added before and after the EPI readout, respectively. Note that the
signal acquisition starts from the second interval during which E2 is formed.
This
prototype sequence was implemented on a Siemens MAGNETOM Spectra 3T scanner and
validated by in vivo experiment. For comparison, conversional DW single-shot
EPI and TSE-based T2W images were also acquired.
Results
3 out of 25 slices of the images acquired by single-shot EPI and the
proposed method with b = 1000 s/mm2 are shown in figure 3.
The geometric distortion, susceptibility artifact and blurring were obvious in
EPI images while alleviated in the images acquired with proposed method.
Figure
4 shows images acquired at basal brain area where B0 inhomogeneity is severe. Images
acquired by the proposed sequence (b = 0 s/mm2) (figure 4b) corresponded well
to the TSE images (figure 4a), exhibiting no obvious off-resonance artifacts. This
is benefit from more contributions from spin echo signals even high
acceleration factor is used.
Discussion and Conclusion
SPLICE technique is robust for overcoming the non-CPMG problems,
while multi-blade technique is highly effective for data acquisition. Our study
shows that the proposed sequence can reliably produce diffusion weighted images
with fast speed, minimized artifact and moderate SAR. It offers the potential
for robust high quality diffusion application.
Acknowledgements
The authors thank Alto Stemmer and Dr. Shi Cheng for help with the preparation of the abstract.References
1.Pipe JG, et al. Multishot diffusion-weighted FSE using PROPELLER MRI. Magn Reson Med 2002; 47: 42–52.
2. Li Z,et al. X-PROP: a fast and robust diffusion-weighted propeller technique. Magn Reson Med 2011;66:341–347.
3.Schick F. SPLICE: Sub-Second Diffusion-Sensitive MR Imaging Using a Modified Fast Spin-Echo Acquisition Mode. Magn Reson Med 1997; 38: 638-644.
4. Alsop DC. Phase insensitive preparation of single-shot rare: application to diffusion imaging in humans. Magn Reson Med 1997; 38: 527–533.