Synopsis
We extended previous work using 3D-FatNavs to enable motion-correction of ultra-high resolution structural acquisitions, including T1-, T2- and T2*-weighted images. Images are of exceptional quality and detail for in-vivo acquisitions.Introduction
We recently successfully demonstrated the
use of a highly-accelerated 3D-fat-excitation GRE at 2 mm isotropic resolution
(3D-FatNav) to allow tracking and retrospective correction of T1-weighted
MP2RAGE and T2-weighted 3D-TSE with sufficient precision to follow the
microscopic movements of the head which are inevitable even for compliant
subjects during extended scan durations [1]. Previous work involved incorporating
the 3DFatNav into sequences where sufficient dead-time already exists for their
insertion without increasing scan time or perturbing the steady-state of the
host sequence. In the current work we
extend their use to the GRE sequence, where the steady-state of the host
sequence must also be considered. We also apply the technique to each of these
sequences with ultra-high resolution protocols (<400 μm isotropic) covering
the whole brain with very long scan times (up to 42 mins per scan, some scans
repeated for averaging) which would be unachievable without motion correction
at high precision.
Methods
All experiments were performed on a Siemens
7T head-only system fitted with a single-channel transmit, 32-channel receive
RF coil (Nova Medical Inc.).
We have recently demonstrated that the
spatial resolution and acceleration factor of the 3D-FatNav can be selected to
reach the desired compromise between the precision of the motion estimates and
the duration of the 3D-FatNav [2]. For this work we decided to use a
conservatively long FatNav duration for very high quality motion estimates –
using the same parameters as described in ref. [1]: 2 mm isotropic resolution,
4×4 GRAPPA acceleration, ¾ partial Fourier undersampling in both PE directions
– resulting in 1152 ms per 3D-FatNav. For the MP2RAGE and TSE protocols this
still fits in existing dead-time without extending the scan-time, but for GRE will
result in a longer scan.
Previous work using motion-navigators with
GRE was careful to create segments for the navigator with a matched-TR to the
host sequence in order to preserve the steady-state condition [3]. We
hypothesized that insertion of a longer navigator would also be possible after
the acquisition of each k-space partition – as the transition into the steady
state would then be the same during each partition (except for the very first).
Figure 1 shows the relative timing of the 3 sequences.
Estimated rigid-body motion-parameters were
obtained using ‘spm_realign’ (http://www.fil.ion.ucl.ac.uk/spm/)
and the full 3D k-space of the host sequence was corrected for this motion,
using a NUFFT (http://web.eecs.umich.edu/~fessler/code/)
to account for the resulting non-Cartesian k-space sampling. For MP2RAGE and
TSE each reconstructed motion-corrected image was co-registered using
rigid-body FLIRT (http://fsl.fmrib.ox.ac.uk/)
before averaging. As the MP2RAGE data were acquired over 2 scanning sessions
each image was also unwarped for gradient nonlinearities (https://github.com/ksubramz/gradunwarp)
prior to co-registration. TSE and GRE images were bias-field corrected using
‘smoothn’ (http://www.biomecardio.com/matlab/smoothn.html).
For the reconstructed TSE data the hippocampus was segmented manually using
itk-SNAP (http://www.itksnap.org/).
Results and Discussion
Figure 2 shows the MP2RAGE ‘uniform’ image with
350 μm isotropic voxels, acquired with 4×31-minute scans over 2 sessions. Figure
3 shows the TSE image at 380 μm resolution (2×29-minute scans) and Figure 4
shows the GRE image at 380 μm resolution (1×42-minute scan).
There is no evidence of image artifact
arising from the disturbance to the steady-state caused by the addition of the
3D-FatNav.
All the images
show excellent contrast and detail.
Figure 5 shows a 3D rendering of the
hippocampus manually segmented from the TSE image. The exceptionally high
isotropic resolution of the scan and the strong contrast allow both the
digitations of the surface, as well as the intricately convoluted ‘tail’ to be
clearly visualized.
Conclusion
Motion-correction enables high-quality
images to be acquired at 7T at ultra-high spatial resolution for a range of
image contrasts. We expect these kinds of protocols to be useful for probing
the in-vivo brain in exceptional detail.
Acknowledgements
This work was in part supported by
the Centre d’Imagerie BioMédicale (CIBM) of
the EPFL, UNIL, UNIGE, HUG, CHUV and the Leenards and Jeantet Foundations, as
well as SNSF project number 205321_153564.References
1. Gallichan et al, MRM (Early View online)
2. Gallichan and Marques, MRM (under review)
3. Tisdall et al, proc ISMRM 2014 p882