Synopsis
In this study, an optimized slice acquisition ordering method
was proposed to improve the signal attenuation and contrast alteration caused
by the crosstalk and MT effect in multi-slice HASTE imaging using a short TR. It demonstrated that a shorter acquisition time is
possible with an optimized slice acquisition order. It allows a shorter TR,
whilst maintaining the SNR and contrast similar to the conventional one, which
is particularly useful in the abdominal imaging.PURPOSE
Half-Fourier single shot spin echo
sequence (HASTE) is a turbo spin echo (TSE) technique and uses a single-shot
method to acquire an entire 2D image in a single TR. Multi-slice HASTE is
widely used in the abdominal imaging with breath hold due to its short scan
time. However the imperfect slice profile of the selective RF pulse will lead
to the undesired excitation of the neighboring regions. The adjacent slice will
be partially saturated, leading to signal attenuation along the slice
direction. Unless a very long TR time is used, crosstalk between slices could
cause artifacts or SNR loss in 2D acquisitions. One conventional solution for
this problem is to modify the slice acquisition order and increase the
effective gap between sequentially excited slices
1. In addition, the
RF pulse train can be considered as off-resonance magnetization transfer (MT)
pulses for other slices, resulting in signal attenuation and contrast
alteration. The MT effect could also be reduced by increasing the slice gap and
altering the slice order
2-3. However the slice gap should not be
increased unlimitedly because small lesions can be missed. The crosstalk and MT
induced artifacts could not be eliminated by simply acquiring slices with
odd/even ordering, in particularly for the imaging with thinner slices. In this
study, an optimized slice acquisition ordering method was proposed to improve
the signal attenuation and contrast alteration in multi-slice HASTE imaging
using a short TR.
METHODS
Theoretically,
the longer interval between excitation pulses is preferred for the neighboring
slices. In order to maximize the slice gap between successively excited slices
and interval time between excitation pulses of neighboring slices, the acquisition
of neighboring slices could be interleaved iteratively. Take 20 slices with 2
concatenations as an example, illustrated in Fig.1. The slices represented by
squares are numbered based on its location in the imaging volume. Different acquisition orders are depicted
using dot lines with different colors. The conventional odd/even slice acquisition
order is indicated with light gray squares and yellow dot line. In the proposed
method, the slices are divided into two groups at first, in which odd slices
are acquired followed by even slices. Then slices within each group will be
interleaved iteratively, shown with gray squares and green line. In Fig.1, the
slices can be interleaved three times at most if there is no criterion,
represented by dark squares and red line. However this acquisition order is not
the optimized one, e.g. the interval is too short between #13 and #11 slices.
Therefore delicate criteria are needed to avoid excessive iterations. In this
study, we assume that the signal attenuation is negligible when the slice gap of
successive excited slices is larger than 3 times of slice thickness or the excitation
interval neighboring slices is longer than 2 second. Although the diagram only illustrates
the situation for the slice number of 20 with 2 concatenations, the
same principle is also applicable to the other slice numbers with different
concatenations.
The
proposed method was implemented in HASTE sequence and validated via in-vivo
study. Abdominal images were acquired from a healthy volunteer using a
commercial 1.5T scanner (MAGNETOM Aera, Siemens Healthcare). Conventional and non-product
HASTE were acquired with same parameters respectively. The imaging parameters are
as follows: TE/TR = 83/(1300, 800, 600 and 400) ms, slice thickness = 6 mm, 20%
slice gap, matrix size = 198×256, slice number = 30, concatenation = 2, in-plane
GRAPPA factor = 2.
RESULTS
Fig.2 shows abdominal images acquired with conventional and
non-product HASTE sequence respectively. Fig.2a is acquired with original HASTE sequence, TR = 1300ms. Fig.2b-d
are acquired with modified HASTE
sequence, TR = 800ms, 600ms and 400ms respectively. For these
images, it notes that the SNR and contrast in Fig.3b-d are very similar to that
of Fig.2a, even with one third of TR time (Fig.2d). Fig.3 shows the corresponding 3D reconstructed abdominal images. All
images are showed with same windowing values. In comparison to images
acquired with conventional odd/even slice acquisition ordering, cerebrospinal
fluid (CSF) signal is more homogenous along the slice direction in Fig.3b,d-f,
even with a shorter TR.
DISCUSSION AND
CONCLUSION
It demonstrated that a
shorter acquisition time is possible with an optimized slice acquisition order.
It allows a shorter TR, whilst maintaining the SNR and contrast
similar to the conventional one, which is particularly useful in the abdominal imaging.
The proposed slice acquisition ordering method could be widely used in HASTE
sequence. In addition, it could also be easily adapted to the other multi-slice
imaging sequences.
Acknowledgements
No acknowledgement found.References
1. M. A.
Bernstein, K. F. King and X. J. Zhou, Handbook of MRI pulse sequences,.
Elsevier Academic Press, 2004, ISBN: 0-12-092861-2
2. R.T.
Constable, A.W. Anderson, J. Zhong, J. C. Gore, Factors influencing contrast in
fast spin-echo MR imaging,. Magn. Reson. Imaging, 10,497, 1992
3. R. Pawluk, A.
Cooney, B. Chomay, Reduction of Magnetization Transfer Effects in Multi-slice
HASTE
Imaging by varying Slice
Excitation Ordering,. ISMRM 10, 2383, 2002