Holger Eggers1, Christian Stehning1, Mariya Doneva1, Elwin de Weerdt2, and Peter Börnert1,3
1Philips Research, Hamburg, Germany, 2Philips Healthcare, Best, Netherlands, 3Department of Radiology, Leiden University Medical Center, Leiden, Netherlands
Synopsis
3D Dixon TSE scans
essentially provide the same information as several conventional 2D TSE scans
in different orientation, without and with fat suppression. However, their
scan time is usually still too long for clinical practice. In this work, the
basic feasibility of accelerating a 3D Dixon TSE scan with PD weighting by a
combination of compressed sensing and parallel imaging was investigated in knee
imaging. Results obtained in half the scan time compared to the use of parallel
imaging alone are presented, which indicate that 3D Dixon TSE scans may become
as fast as current, conventional 3D TSE scans with fat suppression.Introduction
Today, routine
knee examinations consist of several 2D turbo spin echo (TSE)
scans, each covering multiple slices in different
orientation. These scans differ in
contrast, usually including proton density (PD) and
T2 weighting, both without and with fat suppression. A 3D TSE scan with
high, isotropic resolution promises to replace typically two of these scans with similar contrast but different orientation, because images
can be reformatted retrospectively to any orientation without loss of resolution. At the same
time, a chemical shift encoding-based water-fat, or Dixon, TSE scan allows
substituting two of these scans with similar basic contrast and same orientation, but without
and with fat suppression, respectively. Consequently, two 3D Dixon TSE scans may suffice for routine knee examinations, one with PD and one with T2 weighting. However,
the scan time of such 3D Dixon TSE scans is usually still too long for clinical practice, not to mention the increased risk of motion artifacts. Therefore,
the purpose of this work was to explore the basic feasibility of exploiting the
additional acceleration provided by the combination of compressed sensing (CS) and
parallel imaging compared with parallel imaging alone for reducing the scan time of
3D Dixon TSE scans to that of conventional 3D TSE scans with fat suppression.
Methods
We limited this preliminary
investigation to the representative case of PD weighting. For reference, we
took a conventional 3D PD TSE scan, once without and once with fat suppression.
The 3D PD TSE scan without fat suppression had the following parameters: TSE
factor: 60, echo spacing: 4.5 ms, shot duration: 300 ms, profile order: low-high radial, apparent TE: 27 ms, TR: 1000 ms, pixel bandwidth: 595 Hz. Using
a SENSE acceleration of 5.0 (2.0 x 2.5), this scan took 3:18 min. The 3D PD TSE
scan with fat suppression employed SPAIR and the following parameters: TSE factor: 63, echo spacing: 6.4
ms, shot duration: 438 ms, profile order: linear, apparent TE: 59 ms, TR: 1200 ms, pixel bandwidth: 345 Hz. Using a
SENSE acceleration of 4.4 (2.0 x 2.2), this scan took 4:51 min. To introduce chemical shift encoding, the 3D PD TSE scan
without fat suppression served as starting point. Using a multi-repetition strategy,
the number of shots was doubled, and the readout gradient and acquisition
window were shifted in every second shot, requiring in total at least twice the scan time. This was
compensated by complementing SENSE with variable density Poisson
disk sampling and an L
1 norm-based CS reconstruction using pre-calibrated coil sensitivities
1,2. In this way, two single-echo images were
reconstructed separately, before performing the water-fat separation
3. Compared
to the 3D PD TSE scan without fat suppression, the 3D PD Dixon TSE scan had the
following parameters: TSE echo spacing: 6.7 ms, shot duration: 451
ms, apparent TE: 41 ms, pixel bandwidth: 633 Hz. Using an effective CS-SENSE acceleration of 6.8 (2.6 x 2.6),
this scan took 4:49 min, about as long as the 3D PD TSE scan with fat
suppression.
Healthy subjects were examined on a 3T Ingenia scanner (Philips Healthcare, Best, Netherlands) using an
8-element knee receive coil (InVivo, Gainesville, USA). All scans covered
a single knee with a FOV of 145 x 160 x 160 mm
3 and an isotropic resolution of 0.7 mm.
Results
Representative
results obtained in one of the subjects are summarized in Figs. 1-3. Images acquired for reference with the conventional 3D PD TSE scans
are provided in Fig. 1. Corresponding images
reconstructed from the described 3D PD Dixon TSE scan are shown in Fig. 2. They were produced by first generating the water and fat images displayed in Fig. 3 and then calculating linear combinations of them to obtain contrasts similar to those in Fig. 1.
Discussion
The results obtained
so far suggest that the addition of compressed sensing may indeed allow the
acquisition of corresponding images without and with fat suppression in similar scan
times as the acquisition of one of these images today. The examples shown in Figs. 1-3 indicate that both
SENSE, at least in the scan without fat suppression, as well as
CS-SENSE were operated close to the limits of
the employed coil array and the sparsity of the underlying data. Further
acceleration, or artifact reduction, may be expected from a multi-echo
instead of a multi-repetition approach to chemical shift encoding
4.
Acknowledgements
No acknowledgement found.References
1. Lustig M, et al. Magn Reson Med 2007; 58:1182-1195. 2. Liang D, et al. Magn Reson Med 2009; 62:1574-1584. 3. Eggers H, et al. Magn Reson Med 2011; 65:96-107. 4. Madhuranthakam AJ, et al. J Magn Reson Imaging 2010;
32:745-751.