Yong Liu1 and Paul R. Harvey2
1Philips Research China, Shanghai, China, People's Republic of, 2Philips MR, Best, Netherlands
Synopsis
Extraneous in-band Radio Frequency Interference (RFI) signals can mix with the MR signal of a subject creating artefacts in the image data, usually zipper like artefacts. In order to eliminate/reduce the impact of RFI, this abstract describes a software de-noising method which can be implemented both online and off-line through k-space manipulation.Target audience
MR Architects, MR System Engineers and MR Software Engineers
Introduction
Extraneous in-band Radio Frequency Interference (RFI) signals can mix
with the MR signal of a subject creating artefacts in the image data, usually zipper
like artefacts. This can happen if the RF shield/Faraday cage develops a leak
or is under-specified in some way. In order to eliminate/reduce the impact of
RFI, a variety of de-noising methods [1-4] have been proposed that use
dedicated noise pickup coils to sense the ambient noise. The noise signal is
usually multiplied by a scale and delay factor [1-2] before it is subtracted
from the MR signal. In some cases direct subtraction [3-4] is used. The
correction factors are obtained during a pre-scan session for reception [1-2]
or transmission [4]. This abstract describes a software de-noising method which
eliminates the need of a dedicated noise pickup coil, or even a physical noise sampling
coil. The proposed method can be implemented both online and off-line through k-space
manipulation without a pre-scan calibration.
Method
Extraneously received in-band RFI co-exists with the subject MR signal in
every k-space/channel of a multi-channel receiving (Rx) coil. If the RFI signal
would be sampled independently and simultaneously with the reception of MR
signal, such sampled RFI reference could be processed together with the
“contaminated” k-space. In this way, the RFI component in each channel would be
identified and removed. The de-noised k-space data can then be used to
reconstruct a MR image without/with reduced zipper artefacts.
The RFI signals may be obtained in two ways. One
way is to reuse a standard MR Rx coil as a Sniffer Coil (SC). The sniffer coil is
typically placed outside the viable imaging volume in order not to pick up any
MR signal. The sniffer coil can be connected to the MR receiving subsystem and
works as an additional Rx coil. Both sniffer coil and Rx coil acquire their independent
k-space samples. The k-space data of the Rx coil contains the information of
both MR signal and RFI signal while the k-space data of the sniffer coil contains
RFI that was sampled at the location of the sniffer coil. The alternative way,
as demonstrated here, is to use statistical means to extract the RFI information
by utilizing the multiple channel outputs of a multi-channel Rx coil. We refer
to this as the Virtual Sniffer Coil (VSC) method. Each contaminated channel output
contains both MR signal and RFI signal. The correlation of MR information among
these channels is different from that of the RFI component. A statistical
method, e.g. Principle Component Analysis (PCA), is used to separate the RFI
signal cluster from the MR signal cluster.
Due to the spatial distribution of Rx coil
channels, and related various path losses/time-of-arrival from the remote RFI
source, the RFI signal component is not identical in each channel and different
from the RFI reference of either the SC or VSC method. However, all channels are
sampled simultaneously. The mapping between RFI reference and the RFI component
is modeled by a complex ratio R. The magnitude and phase parts of R compensate the
differences in path loss and time-of-arrival at baseband respectively.
In the outer regions of contaminated k-space,
the RFI component dominates. As a first order approximation, equation (1)
holds.
[Snif]·R=[RFIcomponent]≈[Rx]outer_regions (1)
The
compensation ratio R can be calculated, using the least squares method,
line-by-line in k-space, i.e. one R per TR, in equation (2).
min ||[Rx]outer_regions - [Snif]·R||22 (2)
Results
De-noising experiments were performed using a Philips Multiva 1.5T MRI with
a controlled external RFI source. Fig. 1 displays a single channel image in the
VSC de-noising experiment from a Philips 8-channel SENSE head coil. An AM RFI
with an in-band carrier frequency was introduced during the experiment. The k-space
data and the estimated RFI component in this experiment are compared in Fig. 2.
It is found that their triangular AM profile match well in the outer regions of
k-space. Fig. 3 shows the setup of a de-noising experiment using the SC method.
One coil loop is placed on the top of a phantom, which travels into the bore.
Another coil loop is left outside the bore on the table and performs the role
of a SC. A CW RFI was used in the SC de-noising experiment. Fig. 4 presents the
result of SC de-noising method.
Discussion and Conclusion
Two
experimental setups comparing the SC and VSC methods demonstrate their de-noising
effect. The de-noised images appear very close to those without a RFI signal. The
de-noising performance could be improved when a more sophisticated approximation
is adopted in RFI component estimation.
The
SC and VSC method can be used at a clinical site, in the event that RFI occurs
and a technologist is unable to resolve it in a short time. In particular, the
VSC method is suitable for retrospective correction. Both methods may help technologists
continue their tight MR scan schedule with satisfactory image quality.
Acknowledgements
No acknowledgement found.References
[1]. Stephen Hushek, et.al US20080048658A1; [2]. Samuel Patz, et.al,
WO2013016639A1; [3]. Shiro Oikawa, et.al JPS63272336A; [4]. Uri Rapoport, et.al
WO2014167561A2.