Ina Nora Kompan1 and Matthias Guenther2
1mediri GmbH, Heidelberg, Germany, 2Fraunhofer Mevis, Bremen, Germany
Synopsis
3D radial
acquisitions are prone to gradient delay and phase artifacts, which need to be
corrected for. Ideally, the phase can be retrospectively corrected for by
assuming the same phase in k-space
center where all projections meet. However, if data are affected by gradient
delays, profiles are shifted at varying angles and might not intersect at all. Here,
a simple to apply, novel phase correction method is presented in which the
changed trajectory due to gradient delays is incorporated into 3D phase
correction. It was shown that artifacts are reduced by 12% compared to only
gradient delay corrections.Introduction
Advantages of
radial
k-space sampling schemes over
Cartesian sampling are a regular acquisition of the
k-space center, giving increased information on changes to contrast
and insensitivity to motion artifacts. Furthermore, under-sampling artifacts
appear as incoherent streaking artifacts, which are in many applications more
tolerable than Cartesian artifacts [1] and are more suited to a compressed
sensing reconstruction [2]. However, since projections are acquired at varying
angles, radial acquisitions are more sensitive to gradient delays and eddy
currents, leading to image distortions. Additionally, phase errors introduced
by B
0 inhomogeneities may cause artifacts due to signal
cancellations during gridding. Gradient delays can be measured and corrected
for by adapting the gradient timing or by retrospectively shifting the
trajectory [3], [4]. Phase corrections have also been shown to improve image
distortions [5]. Under ideal conditions, the phase of intersection points of
two projections is identical. Without gradient delays, all profiles meet in
k-space center and can be corrected for
by measuring the center phase φ and multiplying the projection by e
-iφ. However,
if data are affected by gradient delays, profiles will not intersect in
k-space center anymore. In the 3D case,
they might not intersect at all. In this work, a novel method is presented, in
which gradient delays are retrospectively corrected by shifting the trajectory
during gridding and these shifts are incorporated into a 3D phase correction.
Methods
Phantom data
were acquired using a 3D golden angle radial sequence on a 3T MRI system (Magnetom
Skyra, Siemens Healthcare, Erlangen, Germany). Gradient delays were measured
and corrected for according to [3]. For additional phase correction, the
following steps were performed:
1) A random reference profile p0 is chosen and its
corresponding complex data are stored in the vector pref.
2) All projections {pj} are found for which the
closest absolute distance d to p0 is smaller than a given
threshold t.
3) The projections {pj} are phase corrected as
described below with respect to p0
and also recorded in pref.
4) This procedure is repeated for
all projections in pref on an individual basis, where each corrected
profile serves as new additional reference projection.
5) The algorithm stops when all
projections are corrected for.
The phase correction
of projection p1 with
respect to projection p2
is performed as follows (fig. 1). Both profiles are linearly interpolated and
the points r1 on p1 and r2 on p2 with
the closest distance d between both
projections are determined. The phases φ1 and φ2 at the
points r1 and r2 and their difference
∆φ = φ1-φ2 are calculated. For a small threshold t (here t=0.001 pixel), the phase in both points should be approximately
the same, when no phase errors are present. Under this assumption, the complex
signal of p2 is multiplied
by a factor e-iΔφ making both
phases consistent.
Results
For
a spherical phantom, the gradient delays were measured to be ∆t
x=1.638
μs, ∆t
y=0.792 μs and ∆t
z=0.335 μs. The uncorrected, only
gradient delay corrected, and additionally phase corrected images along with
their difference images are shown in figure 2. The red arrows indicate regions
of artifact reduction. The ratio
s of
the mean background and the phantom signal is calculated for the 3D volume. A
reduction of 62 % for gradient delay correction and 12 % for the applied phase
correction is found.
Discussion
A
novel method for phase correction incorporating gradient delays has been shown
to reduce artifacts. In comparison with other works such as [6], no additional
measurements are needed. The algorithm is simple and straightforward to apply.
Phase corrections have been performed in a similar study [5], however, there it
was not taken into account that profiles do not intersect in the center in the
presence of gradient delays. Since the gradient delays are assumed to be
exactly known, the accuracy of gradient delay measurements affects the quality
of the phase correction. The presented algorithm is furthermore based on the
assumption that two points in close proximity have approximately the same
phase, which may become invalid in the case of fast changing phases. In the
experiment shown, phase errors are relatively small compared to gradient delay
errors, however still clearly visible. Especially in the case of field strengths greater than 3T, B
0
effects increase and phase errors are likely to become more relevant.
Acknowledgements
No acknowledgement found.References
[1] Peters, D., et al (2000). Undersampled projection reconstruction
applied to MR
angiography. Magn Reson Med,
43(1):91–101.
[2] Block KT
et al (2007). Undersampled radial MRI with multiple coils. Iterative image
reconstruction using a total variation constraint. Magn Reson Med 57:1086-1098.
[3] Block, KT et al (2011). Simple method for adaptive gradient-delay
compensation
in radial MRI. In Proc Intl Soc Mag Reson
Med 19.
[4] Peters, D. et al (2003). Centering the projection reconstruction
trajectory: Reducing gradient delay errors. Magn
Reson Med, 50(1):1–6.
[5] Moussavi, et al (2013). Correction of gradient-induced phase
errors in radial MRI. Magn Reson Med,
71(1):308–12.
[6] Brodsky, E. K., et al (2013).
Rapid measurement and correction of phase errors from B0 eddy currents: impact
on image quality for non-cartesian imaging. Magn
Reson Med, 69(2):509–15.