David S Smith1, Saikat Sengupta1, Aliya Gifford1, and E Brian Welch1
1Institute of Imaging Science, Vanderbilt University, Nashville, TN, United States
Synopsis
We achieve significantly improved image quality in whole-body continuous moving table MRI
without time penalty by using a multichannel, golden angle radial acquisition coupled to a
calibrationless reconstruction.Purpose
Whole-body MRI is of interest in the assessment of many
diseases such as obesity, diabetes, cancer and circulatory malfunctions.
1-4 Continuously Moving Table (CMT) MRI is a high-throughput imaging technique that
offers a powerful alternative for rapid whole-body MR examination. In CMT MRI,
data are acquired with continuous motion of the patient table, allowing scanning of multiple locations in the body in a single sweep. Together
with a continuous golden angle (GA) radial acquisition, CMT MRI image the whole-body in 1-2 min.
5 One downside of the
high speed is lower signal-to-noise ratio. Using an RF receive coil with
multiple channels can improve SNR if the channels are combined appropriately. Here
we describe the development of a single-pass, calibrationless, multi-channel
whole-body CMT MRI technique with significantly improved image quality and no
additional scan time requirement.
Methods
Experiments were performed on a Philips Achieva 3T (Philips
Healthcare, Best, Netherlands) scanner equipped with two-channel parallel
transmit capability. A 16-channel Torso-XL surface coil (Invivo, Gainesville,
Florida) was used with an X-tend tabletop (X-tend ApS, Hornslet, Denmark) to
enable the acquisition of multichannel CMT data. Figure 1 shows the general
setup for the X-tend tabletop. The anterior 8-channel section of the Torso-XL
hung from the top of the scanner bore in a fabric sling. The sling was allowed
to hang low enough to slide against the subject’s body in order to maximize
SNR. The posterior portion of the coil was placed in a rolling “coil wagon”
sandwiched between two layers of the X-tend tabletop. As the X-tend table moved
through the scanner bore, the coil wagon was held stationary at isocenter by
straps attached to the scanner bore covers. The straps limited the range of z
direction table movement to 1.6 m.
An adult male volunteer was scanned at 3 T under an
IRB-approved protocol. The volunteer entered the magnet feet first, supine with
arms at his side. Imaging was performed with full zFOV = 1.6 m, table speed = 2
cm/s, in-plane FOV = 40 cm x 40 cm transverse, in-plane voxel size: 1.56 mm x
1.56 mm, TR/TE: 3.7/1.35 ms, flip angle: 15°,
radial samples: 256, excited slice thickness: 12 mm, readout bandwidth: 854
Hz/pixel, 21,622 radial projections, total scan time: 80 s. Pre-scan
optimization consisting of center frequency and RF drive scale determination
was performed in the abdomen at the position of the umbilicus. After the
preparation phase, the table moved such that the head was slightly inferior to
the z position at isocenter, followed by the CMT scan with the 16-channel array
covering the 1.6 m FOV.
Slice-wise complex image reconstructions were performed using
the Trajectory Optimized Non-Uniform FFT (TRON) algorithm,6 a GPU-accelerated reconstruction customized for GA radial acquisitions. Reconstructed slice centers were separated by 1.56
mm (21 radial projections), matching the in-plane resolution, and were 18.9 mm
(256 radial projections) thick. Reconstruction of the 953 slices of 512 x 512
images required 11.3 s for the 16-channel data and 9.7 s for the body coil data
on a Xeon workstation with dual Nvidia GeForce GTX TITAN X cards. A Gaussian
gridding kernel with width of 3.0 was used with a grid oversampling factor of
1.25. As part of the reconstruction, coil combination was performed in image
space using the adaptive method of Walsh et al.7 because it preserves the relative
complex phase and slightly improves the signal-to-noise level.
Results
Figure 2 shows three axial slices from the volume from both
the built-in quadrature body coil and Torso-XL data. The multichannel image has significantly
reduced artifacts, and SNR improved from 12.9 with the body coil to 19.6 with the 16-channel coil. Uniformity is similar in both images, with notably better sensitivity with the 16-channel coil to the spinal column.
Discussion
The improvement over using the built-in body coil is clear
in even this simple adaptive combination of the channels, suggesting that
simply having the receive coils closer to the subject and a larger number of
receive channels is a great benefit. A notable downside is that the additional
hardware inside the bore reduces the useful bore radius and limits the
potential for applications with large, e.g. obese, subjects. Hardware
improvements could include a coil designed specifically for CMT MRI, with the
majority of elements concentrated at the isocenter and with thin profile or
even built into the scanner. Software improvements could include moving to an
iterative reconstruction with phase constraints.
Conclusions
Significantly improved image quality and SNR in whole-body CMT MRI
was achieved without time penalty using a multichannel acquisition coupled to a
calibrationless reconstruction.
Acknowledgements
NCI K25 CA176219, NCATS UL1 TR000445References
1. Börnert & Aldefeld, 2008. JMRI, 28, 1-12.
2. Barkhausen et al. 2001. Radiology, 220, 252-6.
3. Johnson et al. 1997. Radiology, 202, 262-7.
4. Schaefer et al. 2010. Rofo, 182, 954-64.
5. Sengupta et al. 2015. MRM, Early View (doi:
10.1002/mrm.25848).
6. http://github.com/davidssmith/TRON
7. Walsh et al. 2000, MRM, 43, 682.