Lin Chen1, Shuhui Cai1, and Congbo Cai1
1Department of Electronic Science, Xiamen Unviersity, Xiamen, China, People's Republic of
Synopsis
As an emerging ultrafast imaging method,
spatiotemporally encoded single-shot MRI is advantageous because it can resist
off-resonance effects while retaining spatial and temporal resolutions
comparable to the classical EPI. In this work, a variable density spiral
sampling (VDSS) scheme is proposed for SPEN MRI. An optimization algorithm is
used to design the sinusoidal readout gradient waveform. A specific gridding
algorithm and non-Cartesian super-resovled reconstruction are proposed to
retrieve image. Compared to the Cartesian sampling, VDSS can
provide images with less artifacts and better spatial resolution.Target audience
The
target audience is basic scientists and clinical scientists who are interested
in ultrafast imaging and non-Cartesian sampling.
Purpose
Spatiotemporally
encoded (SPEN) single-shot MRI is an ultrafast MRI technique proposed recently,
which has a great advantage in resisting off-resonance compared to the echo
planar imaging (EPI).
1 However, due to insufficient sampling rate,
the SPEN images are vulnerable to aliasing artifacts.
2 Besides, for
the Cartesian trajectory, the transition between negative and positive
gradient lobes will reduce the acquisition efficiency of SPEN MRI.
In this work, a variable density spiral sampling (VDSS) scheme is proposed
based on bi-SPEN MRI
3 to overcome the above limitations. An
optimization algorithm is used to generate sinusoidal gradient waveforms for
the SPEN spiral MRI. A non-Cartesian super-resolved (SR) reconstruction method
is proposed in combination with a specific gridding algorithm and compressive
sensing (CS) to reconstruct the spiral data.
Methods
The single-shot SPEN MRI
spiral sampling pulse sequence is shown in Fig. 1. The 90° chirp pulse and 180°
chirp pulse are imposed in y and x direction, respectively. An
optimization algorithm is used to design the sinusoidal gradient waveform within limits of gradient magnitude and slew-rate.
4 The flowchart of SPEN spiral imaging is
shown in Fig. 2. Due to the quadratic phase modulation, the phase of acquired
signal oscillates severely, as shown in Fig. 2(d). Based on the characteristic of SPEN approach, the oscillating
phase can be smoothed by removing the quadratic phase modulation without loss
of useful information.
2,3 Since the signal is piecewise smooth after
the quadratic phase offset is removed, as shown in Fig. 2(e), gridding can be
realized by interpolating the spiral data. CS-based SR reconstruction is
employed to retrieve SR image by enforcing the sparsity of the image in
transform domain.
Results
Experiments were performed on a Varian 7.0 T MRI system using
a quadrature-coil probe. The experimental sample was an in vivo rat. The
FOV was 45 ×
45 mm
2 and the slice thickness was 2 mm. For Cartesian bi-SPEN MRI,
the bandwidths and the durations of the phase-encoding chirp pulse and frequency-encoding
chirp pulse were 64kHz/3ms and 8kHz/4ms respectively. The imaging matrix size was
64 ×
64, and the spectral width (sw) was 250 kHz, the gradient slew-rate was
156 G/(cm×ms), the
acquisition time was 27.14 ms. For SPEN spiral MRI, the bandwidths and the
durations of the chirp pulse along the y
direction and x direction were 64kHz/3ms
and 32kHz/3ms respectively. The number of sampling points was 6991, the circle
number was 31, and the acquisition time was 27.96 ms. The results are shown in
Fig. 3. The zoom-in region in Fig. 3 proves that the SPEN spiral
scheme can provide more details and higher image resolution than Cartesian
scheme. As indicated by the yellow arrows in Fig. 3b, there are artifacts
in the Cartesian result, whereas the artifacts are obviously reduced in VDSS
result.
Discussion
Different
from the Cartesian sampling, VDSS is a non-Cartesian sampling scheme. The
aliasing artifacts are less coherent and spread out efficiently across the field
of view in VDSS, which is favorable for the application of CS to further
eliminate the aliasing artifacts. Besides, the profile of readout gradient is
smooth overall in VDSS. The quick switching of gradients can be avoided in
VDSS. Compared to the Cartesian sampling, VDSS can acquire more points within
the same time. Furthermore, VDSS can aggregate the sampling points in signal
regions, which is favorable for the spatial resolution of resulting image.
Conclusion
A VDSS scheme together with an image reconstruction
method is proposed for SPEN MRI. Compared to the Cartesian scheme, VDSS can
provide images with less aliasing artifacts and more detail information.
Acknowledgements
This work was supported by the NNSF of China under Grants
11474236, 81171331 and U1232212.References
1. Tal A, Frydman L.
Single-scan multidimensional magnetic resonance. Prog Nucl Magn Reson Spectrosc
2010;57(3):241-292.
2. Cai CB, Dong JY,
Cai SH, et al. An efficient de-convolution reconstruction method for
spatiotemporal-encoding single-scan 2D MRI. J Magn Reson 2013;228:136-147.
3. Li J, Chen L, Cai
SH, et al. Imaging with referenceless distortion correction and flexible
regions of interest using single-shot biaxial spatiotemporally encoded MRI.
NeuroImage 2015;105(1):93-111.
4. Lustig M, Kim
S-J, Pauly JM. A fast method for designing time-optimal gradient waveforms for
arbitrary k-space trajectories. IEEE Trans Med Imaging 2008;27(6):866-873.