Dinghui Wang1, Zhiqiang Li1, Ryan K. Robison1, and James G. Pipe1
1Imaging research, Barrow Neurological Institute, Phoenix, AZ, United States
Synopsis
Spiral in-out readout is an efficient
sampling scheme for T2-weighted (T2w) spin-echo (SE) sequences. Two sets of spiral in-out data at different TEs are typically acquired so that deblurred water and fat images can be extracted from the spiral-in and the spiral-out data separately, which are then combined together.
A method has been recently proposed
to reconstruct water and fat images from a single set of spiral in-out data. This work demonstrates the feasibility of using this method as a
fast and scan efficient means of fat suppression to reduce the scan time of the
spiral T2w SE by a factor of 2.Introduction
Spiral in-out readout is an
efficient sampling scheme for T2-weighted (T2w) spin-echo (SE) sequences
1.
Two sets of spiral in-out trajectories are typically acquired with a shift of ΔTE between the two
acquisitions. Water and fat images can be successfully extracted
and deblurred from two spiral-in images and two spiral-out images separately
2
and then combined. This method is referred to as the separate deblurring hereafter for simplicity. A new method (referred to as the joint deblurring hereafter) has been recently proposed to reconstruct water and fat
images from a single set of spiral in-out data
3, with a time
delay ΔTE
io inserted between spiral-in and the spiral-out parts.
The purpose of this work is to investigate the feasibility of reducing the scan
time of the spiral T2w SE (by a factor of 2) with the joint deblurring by both experiments and numerical simulation.
Methods
Two volunteer data sets were collected
on a 3T Philips Ingenia scanner, with one and two sets of spiral in-out acquisitions respectively. ΔTE was 1.15 ms for the latter. The readout time was
20.7 ms, including a ΔTEio of 1.0 ms. Other scan parameters include:
FOV = 230×230 mm2, resolution =1×1 mm2, slice thickness =
4 mm, slice gap = 1mm, slice number = 20, TR = 2350 ms and TE = 80 ms. A map of B0 inhomogeneity
(Δf0) was obtained by a Cartesian mDixon scan. The spiral-in and spiral-out
images were reconstructed after corrections for trajectories4 and B0
eddy currents5-6. Water and fat images were reconstructed from the
two data sets by the joint deblurring and the separate deblurring,
respectively. In the joint deblurring, a
slowly varying phase map (PM) was evaluated from the first-pass results3 by low-pass filtering (applying 20% Hann window in the k-space),
which was removed from water and fat images in the subsequent process.
The water and fat images from the separate deblurring in the
experiment were utilized to simulate the spiral in-out k-space data by the
discrete Fourier transform. Noise was added to the simulated k-space data to characterize
the dependence of the signal-to-noise
ratio (SNR) on ΔTEio. Furthermore, the influence of T2 decay was also evaluated in the simulation. The water images were approximately segmented to white matter, gray matter and cerebrospinal fluid (CSF) by thresholding. T2 was assigned 70 ms for white matter, 100 ms for gray matter, 2000 ms for CSF as shown in Fig. 1. T2 was assumed 70 ms for fat. A constant T2' (the relaxation rate due to field inhomogeneity across a voxel) of 50 ms was also included in the simulation.
Results and Discussion
The water images reconstructed
by the joint deblurring demonstrate similar overall sharpness and contrast in
comparison with that of the separate deblurring (Fig. 2). The SNR reaches the highest value when ΔTEio is between 1.1 and 1.3 ms (Fig. 3). The optimal SNR of
the joint deblurring is about $$$\sqrt{2}/2$$$ of the optimal
SNR of the separate deblurring that uses twice the scan time. Therefore, the scan efficiency (SNR/scan time½) is comparable between the two methods.
The difference between the reconstructed images from the simulated data with and without T2 decay is negligible, which implies that the error due to
T2 decay is insignificant. However, there is subtle but observable deviation in the reconstructed fat images when the evaluated PM (PMeval) is used instead of the known PM (PMtrue) as shown in Fig. 4. Therefore, inaccurate PM estimation may in part
contribute to the artifacts in the fat images of the joint deblurring in the
experiment.
Conclusion
This work has demonstrated the joint deblurring as a
promising means of fast, scan efficient fat suppression for spiral T2w SE
imaging.
Acknowledgements
This work was
funded by Philips Healthcare. References
1. Li Z, et al. ISMRM
2015; 2444. 2. Wang D, et al. MRM
2015; doi:10.1002/mrm.25620. 3. Wang D, et al.
ISMRM 2014; 1661. 4. Alley MT, et al. MRM 1998; 39:581-587. 5. Brodsky EK, et al. MRM 2013; 69:509-515. 6. Robison RK, et al. ISMRM 2014; 1620.