Michael Germuska1, Alberto Merola1, and Richard G Wise1
1CUBRIC, Cardiff University, Cardiff, United Kingdom
Synopsis
An
emerging method for quantitative mapping of OEF is by dual calibration of the
BOLD signal. 1,2 However, this method is highly sensitive to
measurement noise, resulting in unstable estimates of OEF. An
alternative approach is to use flow-diffusion equations to calculate the
biophysically supported OEF. 3 However, this approach is limited by
the need to assume a tissue oxygen tension (PtO2). We propose a
method for combining these two approaches, producing calibrated BOLD estimates
of OEF that are constrained by a modelled flow-diffusion relationship of oxygen
extraction. The proposed method is shown to produce stable estimates of OEF and
PtO2.Purpose
To
improve the robustness of cerebral oxygen extraction fraction estimation by
dual calibrated fMRI, and to produce in-vivo estimates of tissue oxygen tension
without additional data acquisition.
Methods
We
have recently proposed a forward modelling method for the estimation of OEF
using the dual calibrated fMRI methodology4. This method allows for
the simultaneous estimation of resting OEF, resting blood flow (CBF), BOLD-weighted
blood volume (CBV) and the flow-related cerebral vascular reactivity (CVR). In
this work we re-parameterise the minimisation problem in terms of PtO2,
CBF, CBV and CVR. In this parameterisation OEF is an internal parameter that is
derived as a function of PtO2, CBV and CBF. This follows from the
flow-diffusion modelling undertaken by 3, where the biophysically
supported OEF is modelled as a function of the mean transit time (MTT), PtO2,
and an equal forward and reverse diffusion rate constant k. The MTT can be
estimated from CBV/CBF and the rate constant is fixed to achieve a typical OEF
(0.3) for normal resting physiological values (MTT = 1.4 s, PtO2 =
25 mmHg), as per 3. The steady state value of OEF for a wide range
of MTT and PtO2 values can be calculated using numerical integration
techniques to provide a lookup table for OEF (see figure 1). Thus, estimates
OEF via the dual calibrated BOLD method can be constrained to obey the
flow-diffusion model. In the work presented by 3 a model was
presented that allows capillary transit time heterogeneity (CTTH) to be
included in estimates of OEF. However, for consistency with standard BOLD
models, CTTH effects have not been included in our current implementation.
The performance of the proposed technique was assessed against the standard dual calibrated fMRI technique in a cohort of 10 healthy volunteers. Image data were acquired on a 3 T whole body MRI system. Functional data were acquired using a pulsed arterial spin labelling (ASL) using a QUIPSS II acquisition scheme. This sequence used a dual-echo gradient echo (GRE) readout (TE1 = 2.7 ms TE2 = 29 ms, TR = 2.2 s, flip angle 90°, FOV 22 cm, matrix 64 × 64, 12 slices of 7 mm thickness with an inter-slice gap of 1 mm, TI1 = 700 ms, TI2 = 1500 ms). Respiratory challenges consisted of 3 periods of hypercapnia (5% CO2) and 2 periods of hyperoxia (50% O2) interleaved with room air, for a total acquisition time of 18 minutes. Data were analysed using our proposed forward modelling method, 4 using a regularised non-linear least squares minimisation to solve for the parameter estimates.
Results
Estimates
of OEF are more closely correlated to the mean transit time (MTT) when they are
constrained by the biophysical flow-diffusion relationship (see figure 1).
Using the non-constrained fitting method a positive relationship is found
between MTT and OEF with an R
2 of 0.51. This correlation is expected
as a longer MTT allows for greater extraction of oxygen. When the OEF estimates
are constrained by the biophysical flow-diffusion relationship the R
2
increases to 0.88. This increase in R
2 represents better agreement
with the biophysical model of oxygen extraction and so is expected in the
proposed method. The coefficient of variance (COV) of OEF estimates is reduced
from 35% to 27% when using the constrained fitting method. This reduction in
COV results in a significant reduction in implausible estimates of resting OEF,
as can be seen from the example parameter maps in figure 2, and the associated
histograms in figure 3. Estimates of PtO
2 have a group mean value of
24.5 mmHg, which are found to have a negative correlation with resting CBF (R
2
= 0.23).
Conclusions
The
proposed method of OEF estimation by calibrated fMRI demonstrates reduced
variance in parameter estimates and better agreement with expected biophysiological
processes when compared to standard analysis methods. We suggest that the incorporation of flow-diffusion modelling into quantitative
calibrated fMRI measurements could significantly improve the robustness and
accuracy of in-vivo estimates, while simultaneously providing additional
information on tissue oxygen tension that may prove useful in the assessment of disease and brain function.
Acknowledgements
We
would like to thank Sune Jespersen for sharing his well-commented Matlab modelling
code with us, and would like to acknowledge the support of the UK Engineering
and Physical Sciences Research Council (EP/K020404/1) for this work.References
1. Bulte D, Kelly M, Germuska M, Xie J, Chappell M, Okell T, Bright M, Jezzard P.
Quantitative measurement of cerebral physiology using respiratory-calibrated MRI.
NeuroImage 60 (2012) 582–591
2. Gauthier C and Hoge R.
Magnetic resonance imaging of resting OEF and CMRO2 using a generalized
calibration model for hypercapnia and hyperoxia. NeuroImage 60 (2012) 1212–1225.
3. Jespersen S and Østergaard L. The roles of
cerebral blood flow, capillary transit time heterogeneity, and oxygen tension
in brain oxygenation and metabolism. Journal of Cerebral Blood Flow &
Metabolism (2012) 32, 264–277.
4. Germuska M, Merola A, Stone A, Murphy K, Wise
RG. A Bayesian framework for the estimation of OEF by calibrated MRI. 23rd
ISMRM Annual Meeting (2015).