Aiming Lu1, Xiangzhi Zhou1, Mitsue Miyazaki1, Masao Yui2, Masaaki Umeda2, and Yoshiharu Ohno3,4
1Toshiba Medical Research Inst., Vernon Hills, IL, United States, 2Toshiba Medical System Corp, Otawara, Japan, 3Advanced Biomedical Imaging Research Center, Kobe University Graduate School of Medicine, Kobe, Japan, 4Division of Functional and Diagnostic Imaging Research, Department of Radiology, Kobe University Graduate School of Medicine, Kobe, Japan
Synopsis
T2* mapping with a
single-exponential model have been demonstrated to be useful in accessing pulmonary
functional loss. However, the model does
not fully explain the signal evolution at longer TEs. We
propose to improve the T2* characterization in the
lung parenchyma with a bi-exponential
model. Using a 3D multi-echo
radial sequence, our results demonstrated that short T2* values and the
volume fractions of the two compartments could be obtained on a clinical 3T
scanner. In addition to the improved accuracy
of the short T2* measurement, the added fraction values could also potentially
be used as biomarkers.Introduction
Pulmonary imaging is of
clinical significance. Currently pulmonary imaging is mainly performed with CT,
which raises concern of ionization radiation especially in the young. Pulmonary
imaging with MRI is desirable but challenging due to the lower proton density,
short T2* in lung parenchyma, and the respiratory motion. Recently, Ultra-short
TE (UTE) MRI has been shown to perform similarly as standard and low
dose CT in diagnosing pulmonary diseases owing to its short TE capability and resilience to motion (1). Lung parenchyma T2*
values measured with a single exponential model have also been demonstrated to
have a role in pulmonary functional loss assessment (2). However, the single compartment model does
not fully describe the signal evolution especially at longer TEs. In this work, we propose to improve the characterization of T2*s
in the lung parenchyma with a bi-exponential model. Using a 3D multi-echo radial sequence, our results demonstrated that short T2* (T2*S)
values and the volume fractions of the two compartments could be readily
obtained on a clinical 3T scanner.
Materials and methods
Pulmonary imaging was performed on two volunteers on a
3T Toshiba Vantage Titan 3T scanner (Otawara, Japan) with a four-half-echo 3D radial sequence (Figure 1). Images
were acquired during the expiration phase with respiratory-gating. To minimize
the impact of eddy currents, a fly-back readout was used. The echo space was set
to 1.1ms so that the short term eddy currents (time constant tc ~100us) induced by
the gradients before the next readout decayed to a negligible level (~5tc). Three
acquisitions were carried out on each volunteer with the TEs of the first echoes
set to 0.1ms, 0.5ms and 0.9ms, respectively. Other acquisition parameters
include TR/flip angle/FOV/slab thickness/resolution/# of radial lines/# of
segments: 5.8 ms/6°/FOV 50cm/25.6cm/3.9 mm
isotropic/18432/64.
Image reconstruction was performed using
gridding followed by complex Fourier transform. Eddy currents during the
readout was accounted for with the measured actual trajectories (3). All 12
echo images were used for nonlinear bi-exponential fitting in Matlab (Mathworks
Inc.) on a voxel by voxel basis. T2*S range was set to be 0.1ms-10ms, and long
T2* (T2*L) was set to be 5ms -1000ms for the fitting.
Results and discussion
Typical signal
evolutions in several typical voxels are shown in Figure 2. Voxel #1 was
selected in liver tissue and the signal was nearly flat (not shown). At least
two compartments are visible in plots of most lung parenchyma voxels (#2-#5),
especially voxels #4 and #5. The bi-exponential model fits the data fairly
well, although a single exponential
model may also work for other tissues such as the vessels and the lung
parenchyma voxels with high water content (e.g., #6). A small peak is observed
in some plots, which may be due to the chemical shift of pulmonary surfactant.
The
calculated T2*S maps and the volume fraction maps of the short and long T2* components
are shown in Figure 3. Lung parenchyma tissue show higher T2*S values and
volume fractions than other tissues. The calculated T2*S values of lung
parenchyma range from 0.4-1.1 ms, consistent
with those reported in literature. The fractions of T2*S values are different
at the anterior and posterior sides, which increased from ~50% to ~70% in the
ROIs placed in lung parenchyma from anterior to posterior. T2*S fraction values were low (below 20%) in other tissue such as liver,
kidney, as expected.
The origin of multiple compartments in lung
parenchyma voxels
could be due to the existence of multiple tissue types in lung parenchyma. However,
the partial volume effect of lung parenchyma with
blood could also contribute to the observed phenomenon. In the latter case, the
bi-exponential model may help to differentiate the contributions from lung
parenchyma and blood and consequently increase the accuracy of T2*S measurement.
There are several limitations in our studies. The accuracy of the T2*L needs further
investigation due to the relatively short TEs used. As can be seen in the plots, chemical shift could
also compromised the accuracy of T2* characterization. Inclusion of chemical shift in modelling could help. In
addition, all examinations were performed in the supine position in this study. Positional difference needs to be further
investigated.
Conclusion
We have demonstrated the application of a multi-echo
3D radial sequence with UTE capability to image lung parenchyma on healthy
volunteers. Our data indicate the
existence of multiple compartments in lung parenchyma voxels. In addition to
the improved accuracy of T2* measurement in lung parenchyma with the
two-compartment model, the added fraction values of the two compartment could
also potentially be used as biomarker for lung diseases.
Acknowledgements
No acknowledgement found.References
1. *Ohno Y, et al., JMRI,
doi:
10.1002/jmri.25008. ( Epub ahead of print). 2. Ohno et al., Eur J Radiol 2013. 82:1359-1365. 3.
Lu. et al., Magn Reson Med.
2011 66:1582-9