dcQSM: Quantitative Susceptibility Mapping by Directly Fitting Complex Images
Zhe Liu1, Pascal Spincemaille2, and Yi Wang1,2

1Biomedical Engineering, Cornell University, Ithaca, NY, United States, 2Radiology, Weill Cornell Medical College, New York, NY, United States

Synopsis

The quality of Quantitative Susceptibility Mapping (QSM) depends critically on a correct estimation of total magnetic field, which may sometimes be degraded by phase unwrapping failure. We propose to bypass the traditional field estimation and phase unwrapping steps and estimate both background field and local susceptibility distribution directly from complex GRE images, which is referred to as dcQSM. Since no field is explicitly existent in our method, dcQSM eliminates phase unwrapping errors in tradition methods.

Purpose

Quantitative susceptibility mapping (QSM) allows quantification of magnetic biomarkers such as iron, calcium and gadolinium in the brain, blood or liver [1]. Traditional QSM is generated from gradient echo (GRE) data using steps of field estimation, field/phase unwrapping, background removal and local dipole inversion [1] as shown in the left of Fig.1. Traditional QSM has a major drawback that an error in field unwrapping, often occurring at brain boundary due to low SNR, can cause severe artifacts in the final susceptibility map. Here we propose a novel QSM method directly from the complex GRE data (dcQSM) that requires neither field mapping nor phase unwrapping, by fitting both the background field and the local susceptibility distribution directly to original complex GRE data.

Methods

The optimization model for nonlinear background removal is:

$$\chi_B^*=arg\min_{\chi_B}\sum_j^{N_B}\parallel I_j - \mid I_j \mid e^{-i2\pi TE_j(d\star\chi_B)} \parallel_2^2 (1)$$

with $$$I_{j}$$$ the complex GRE image for the j-th echo, $$$TE_{j}$$$ the j-th echo time, $$$d$$$ the dipole kernel and $$$\chi_{B}$$$ the background susceptibility distribution responsible for background field inside the region of interest $$$M$$$. This is a nonlinear version of projection onto dipole field (PDF) [2] model but, additionally, fields are no longer explicitly obtained by instead fitting the susceptibility directly to complex images $$$I_{j}$$$, hence bypassing the field estimation as well as phase unwrapping. From the estimated $$$\chi_B^*$$$, the effect of the background inhomogeneity can be removed from the original complex GRE data:

$$ I_{Lj}=I_je^{i2\pi TE_j(d\star\chi_B^*)} (2) $$

Here the modified GRE data $$$I_{Lj}$$$ are complex images whose phase is induced by local field only. The local susceptibility $$$\chi_L^*$$$ in $$$M$$$ can now be directly estimated from this complex signal:

$$ \chi_L^{*}=arg\min_{\chi_L}\sum_j^{N_B}\parallel I_{Lj} - \mid I_{Lj} \mid e^{-i2\pi TE_j(d\star\chi_L)} \parallel_2^2+\lambda\parallel M_G\triangledown\chi_L \parallel_1 (3) $$

Here the binary mask $$$M_{G}$$$ is derived from anatomic magnitude image as in traditional QSM [1]. This dcQSM differs from traditional QSM [1] in that we are fitting to multi-echo complex signal rather than the local field. Both minimization problems (1) and (3) are solved using the iterative Gauss-Newton linearization and Conjugate Gradient search [1].

We applied the proposed dcQSM on an agarose phantom in which 4 balloons filled with Gadolinium solutions at prepared susceptibilities of 0.4, 0.8, 1.6 and 3.2ppm. The phantom was scanned at 3T (GE Healthcare, Waukesha, WI) with FA=15, FOV=18cm, 0.8$$$\times$$$0.8$$$\times$$$0.8 mm3 and echo spacing 3.5ms. We estimated mean susceptibility inside each balloon from the QSM. We also applied our proposed dcQSM on 5 human brain data to test its robustness against phase unwrapping problem. Subjects were scanned at 3T (Siemens Skyra), FA=15, FOV=24cm, 0.9$$$\times$$$0.9$$$\times$$$2 mm3 and echo spacing 5ms. For both phantom and in vivo experiment, dcQSM was compared with traditional QSM.

Results

In the phantom experiment, as shown in Fig. 1, both traditional QSM and proposed dcQSM achieved comparable estimation accuracy for balloon susceptibilities (traditional QSM: 0.48, 0.84, 1.61, 3.13 ppm; Proposed dcQSM: 0.45, 0.84, 1.62, 3.25 ppm). In an example from in vivo experiments, the traditional method didn’t generate a correct local field due to phase unwrapping failure, resulting in a shadow artifact in QSM in Fig. 2. In contrast, the proposed dcQSM method reconstructed a susceptibility map without the artifact.

Discussion

Our proposed dcQSM removes the need for phase unwrapping by directly fitting the background field to the complex raw GRE data using nonlinear optimization (1). This dcQSM method achieved the same accuracy as the traditional method in phantom susceptibility estimation. In clinical MRI data, dcQSM is robust against the phase unwrapping errors sometimes seen with the traditional method. In the future work, we will focus on developing faster algorithm in solving our nonlinear background removal (1) and local susceptibility inversion problem (3).

Conclusion

Our proposed dcQSM algorithm eliminates the need for phase unwrapping in traditional QSM.

Acknowledgements

We acknowledge support from NIH grants RO1 EB013443 and RO1 NS090464

References

1. Liu, Tian, et al. "Nonlinear formulation of the magnetic field to source relationship for robust quantitative susceptibility mapping." Magnetic Resonance in Medicine 69.2 (2013): 467-476.

2. Liu, Tian, et al. "A novel background field removal method for MRI using projection onto dipole fields (PDF)." NMR in Biomedicine 24.9 (2011): 1129-1136.

Figures

Figure 1. Flowchart for traditional QSM (Left) and proposed dcQSM (Right) on a balloon phantom. Both methods produced accurate balloon susceptibilities.

Figure 2. Brain QSM with traditional method and proposed dcQSM. Notice the failure in phase unwrapping in cortical brain, and the induced shadow artifact with traditional QSM. Proposed dcQSM removed that artifact.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1570