Synopsis
A
criterion of A-optimality was used to optimize SSFP sequence
acquisition parameters in order to perform multi-parametric mapping
of the physical parameters proton density, relaxation rates and
apparent diffusion coefficient. A fast calculation of the
steady-state was used to estimate the Fisher information matrix from
which fitted parameter error was determined from its inverse.
Considering a range of possible T1 and T2 values, and relevant ADC,
the acquisition parameters were optimized over the four dimensions of
TR, prescribed flip angle, RF phase increment and spoiling gradient
to achieve the minimum error on the effectively fitted physical
parameters. It is demonstrated that choosing targeted T1 and T2
values over a wide range of expected values enables defining
acquisition protocols that minimize the error over this range.PURPOSE
To
analyze contrast in fast steady-state sequences, a given set of
acquisition parameters can be considered: prescribed flip angle
(alpha),
Repetition Time (TR),
phase cycling increment (phase)
and spoiling gradient (characterized by a,
the distance along which a dephasing of
2π occurs
between 2 TRs). The measured signal then depends on: proton density (M0),
background phase (Phi0),
effective flip angle alphaeff, longitudinal and transversal
relaxation time (T1,
T2)
and the apparent diffusion (ADC).
Our objective is to
define the optimal set of acquisition parameters that will give the
smallest estimation error over a predefined range of physical
parameters.
The
Fisher information matrix (FIM)
and the Cramer-Rao lower bounds (CRLB)
are standard tool for experiment-design optimization [2,3] or for
fitting error evaluation [4]. Here, it is shown that the acquisition
parameters can be optimized by minimizing the CRLB
of
each physical parameters, thus reducing scan time/enhancing precision
for multi-parametric mapping.
METHODS
For any
unbiased joint estimation method of the above-mentioned Npar=6
physical parameters, a lower bound of the error
σθk
in the parameter θk,
is given by the CRLB: $$\sigma_{\theta_{k}}\geq CRLB = \sqrt{FIM^{-1}_{kk}}$$
When Nmeas
observed data Yi
are normally distributed about Si=S(xi,θk)
with variance σ
the FIM is :
$$FIM_{jk}=\sum_{i=1}^{N_{meas}}\frac{1}{\sigma^{2}}\frac{\partial
S_{i}}{\partial \theta_{j}} \frac{\partial S_{i}}{\partial
\theta_{k}}$$
Note that
the MR signal is kept as a complex value, thus the assumption of
Gaussian noise is reasonable. For all simulations, a SNR of 500
(σ=1/500)
was used. The FIM was calculated numerically with a fast algorithm
[1]. In order to optimize the acquisition protocol, the objective
function J was defined as the sum of the normalized CRLB:
$$J=\sum_{k=1}^{N_{par}}\frac{\sigma_{\theta_{k}}^{2}}{\theta_k^2}$$
To
normalize the CRLB for $$$alpha_{eff}$$$ and
Phi0, nominal references of 90° and π
were used, respectively. To obtain optimized acquisition parameters,
the minimization of J was perform with a stochastic optimization
algorithm, SOMA [5].
Four sets
of acquisition parameters were considered and compared. The first one
was defined empirically and the others were obtained by considering
12 acquisitions and the following constraints: a constant TR between 6 and 60 ms. alpha between 2° and 180°, the phase increment between 0
and 359° by 1° steps and a
between 125µm and 1250µm.
Protocol1:
empirically defined set with 28 acquisitions: TR=12.5ms, alpha=24°,
a=312µm,
and 28 phase increments
[0°/1°/-1°/2°/-2°/4°/-4°/8°/-8°/16°/-16°/180°/90°/-90°/45°/-45°/0°/117°/-117°/150°/-150°/32°/-32°/64°/-64°/128°/-128°/0°]
Protocol2:
optimization performed considering T1=0.9s, T2=0.08s and
D=1.10-9m2s-1
Protocol3:
optimization performed considering T1=5s, T2=0.05s and D=1.10-9m2s-1
Protocol4:
optimization performed by summing the normalized CRLB over 4 values : T1=[0.5/5/0.5/5]s, T2=[0.05/0.05/2/2]s, and
D=1.10-9m2s-1
RESULTS
optimized parameters:
Protocol2: TR=12.5ms, alpha=[2/8/12/20/20/21/21/27/32/38/48/49], phase increments =[356/4/1/178/283/3/3/357/240/356/357/182] a=[280/742/1222/125/126/125/125/147/131/1076/943/1022].
Protocol3: TR=12ms, alpha=[4/4/5/9/11/12/13/13/14/16/16/18], phase increments =[359/147/140/360/1/359/359/1/359/359/358/359] a=[845/547/928/965/168/157/146/154/143/816/1234/1060].
Protocol4: TR=10.3ms, alpha=[4/8/12/14/14/15/16/19/19/23/23/61], phase increments =[2/0/359/1/359/178/2/182/357/0/2/180] a=[915/1006/146/142/138/1204/131/163/1117/872/940/1219].
It covers very different flip angles, phase increments and
spoiling gradients.
Table1 provides the expected error for each protocol estimated for typical
brain tissue relaxation and diffusion parameters. As protocol2 was
specially designed for these expected relaxation and diffusion
parameters, it provides the minimal error. It can be seen
that it provides more precise estimations than protocol1 with much
less Nmeas
acquisitions demonstrating the gain in total scan time when using
optimization as compared to empirically defined protocols.
To
estimate if the protocols provide adequate precision over a wide
range of relaxation rates, the relative error was computed for a
large range of T1 and T2 (from 0.5 to 5s and 50ms to 2000ms
respectively) (figure 1-4). Results are shown for D=1.10-9mm2s-1,
but were similar for values between 0.5 to 3 demonstrating a limited influence of diffusion. For all
parameters, excepted for T2, the relative error was increasing with
increasing T1 and decreasing T2. Protocol3, which was optimized for
targeted T1/T2=5s/50ms provides the smallest maximum relative error
over the range of T1 and T2 values (reduced by a factor 2 as compare
to protocol2 for M0, T1 and ADC). When using an objective function
designed for extreme T1 and T2 values, such as in protocol4, more precise estimates with more uniform errors are obtained in a
wider range.
DISCUSSION/CONCLUSION
Optimization
of acquisition parameters through the CRLB may provide a valuable
tool for objective protocol definition in MRI, especially with SSFP
sequence in which the contrast mechanisms are leading to complex
signal models. In the context of
multi-parametric mapping, it was shown that protocols can be
optimized for proton density, background phase, effective flip angle,
relaxation times and apparent diffusion coefficient mapping with
targeted relative errors on the fitted physical parameters. Based on
an adequate choice of the objective function, it was shown that error
on fitted parameters can be minimized for a given range of T1 and T2.
Acknowledgements
Grants Institut des
neurosciences translationnelles - ANR-10-IAIHU-06, and Infrastructure d’avenir
en Biologie Santé - ANR-11-INBS-0006References
[1] de Rochefort, ISMRM 2015
[2] Alexander DC, MRM 2008 60:439-448
[3] Anastasiou A., MRI 2004 22:67-80
[4] Lankford L., MRM 2013 69: 127-136
[5] Zelinka I. SOMA Self organizing migrating algorithm. In Babu BV and Onwubolu G, editors, New optimization techniques in engineering. Springer, 2004.