Meher Juttukonda1, Manus Donahue1, Melissa Gindville2, and Lori Jordan2
1Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States, 2Pediatrics, Division of Pediatric Neurology, Vanderbilt University Medical Center, Nashville, TN, United States
Synopsis
Quantitative
CBF maps derived from pseudo-continuous ASL (pCASL) may be useful in assessing
stroke risk in sickle cell anemia (SCA) patients, but T1 relaxation
of SCA blood must first be characterized. Venous blood samples were collected
from SCA patients as well as normal subjects, and an inversion recovery
approach was used to quantify the T1 relaxation times ex vivo. For similar hematocrit,
oxygenation, and temperature, T1 relaxation times of SCA blood
appear similar to those of normal blood. Therefore, computation of CBF in SCA
patients may not be affected by the assumption of normal blood T1
relaxation.Purpose
Quantitative CBF maps derived from pseudo-continuous
ASL (pCASL) may be useful in assessing stroke risk in patients afflicted with
sickle cell anemia (SCA). The computation of CBF
1 requires knowledge
of the T1 relaxation time of blood which has been elegantly
characterized for normal subjects as a function of oxygenation level,
hematocrit, and field strength
2,3. However, longitudinal relaxation times
have not been characterized in SCA patients whose hemoglobin consists of the
variant hemoglobin S (HbS) in addition to normal hemoglobin A (HbA). In this
study, we measure the T1 relaxation times of SCA blood over a normal
range of hematocrit and percent HbS.
Methods
Experiment. Six
samples (n=3 SCA; n=3 non-SCA) of human venous blood were collected into
heparinized tubes from volunteers who provided informed, written consent. For
SCA blood, laboratory tests were performed to obtain the hematocrit and percent
HbS values. For normal blood, hematocrit values were obtained using an i-STAT
handheld blood analyzer (Abbott, Ottawa, Canada). Blood oxygenation
measurements were obtained for all six samples using the blood analyzer. MR
measurements were performed on the blood samples at 3T (Philips), and care was
taken to ensure
in vivo temperature
conditions. Prior to scanning, the
temperature of the blood samples was stabilized to approximately 37 °C using a heated water bath. During scanning, the
tubes were positioned inside a jar of peanut butter which was also previously heated
to 37 °C. The container was then wrapped in a warm water
circulation system and placed in the scanner. T
1-weighted MR images
were acquired using an inversion recovery sequence: TE=90ms; TI=30ms, 60ms,100ms, 250ms, 500ms, 1000ms, 2000ms, 3000ms, 5000ms with constant recovery time=8000ms; in-plane
spatial resolution=1.25x1.25mm
2.
Analysis. The T
1 of each sample was computed by fitting
the mean magnitude MR signal from a region inside the blood sample at each TI
using a two-parameter model and a non-linear least squares algorithm
4.
Results
For the normal blood samples, the mean (±standard deviation) values were 39% (±0.6) for the hematocrit and 1.55s (±0.11) for the T
1. For the SCA blood
samples, the mean (±standard deviation) values were 32% (±7) for the hematocrit, 78.8% (±4.8) for the percent HbS, and 1.47s (±0.13) for the T
1.
Representative T
1 maps (Figure 1A)
and the two-parameter fitting for extracting T
1 (Figure 1B) are
shown for representative normal and SCA blood samples.
Discussion
CBF values computed from quantitative pCASL
measurements may be useful in assessing stroke risk in SCA patients. However, T
1
relaxation of blood from SCA patients, which contains HbS in addition to HbA,
must be characterized before CBF can be reliably quantified. For similar hematocrit,
oxygenation, and temperature, T
1 relaxation times of SCA blood appear
similar to those of normal blood, and assumption of normal blood T
1
for SCA patients would result in a relatively low error in CBF of
approximately 8.6%. These results should
be considered in the context of the following limitations. First, the
temperature of the samples during scanning was lower than 37 °C but was identical for both normal and SCA blood. Next,
only venous blood oxygenation was considered. Finally, the experiments were
conducted on
ex vivo samples;
in vivo SCA blood may have slightly
different T1 relaxation characteristics due to flow velocity,
turbulence, and other physiological factors. Ongoing work involves examining
the dependence of the T1 and T2 times of SCA blood on
hematocrit and oxygenation.
Conclusion
T
1 relaxation times in SCA blood
appear to be similar to those observed in normal blood. Computation of CBF using pCASLfor
stroke risk assessment in SCA patients may not be affected by the assumption of
normal blood T
1 relaxation.
Acknowledgements
No acknowledgement found.References
1. Alsop
DC, Detre JA, Golay X. Recommended implementation of arterial spin-labeled
perfusion MRI for clinical applications. Magn Reson Med. 2015; 73:102-116.
2. Lu H, Clingman C, Golay X, van Zijl PC. Determining the
longitudinal relaxation time (T1) of blood at 3.0 Tesla. Magnetic resonance in
medicine. 2004; 52(3): 679-682.
3. Kalavagunta C, Michaeli S, Metzger GJ. In vitro Gd-DTPA
relaxometry studies in oxygenated venous human blood and aqueous solution at 3
and 7T. Contrast Media Mol Imaging. 2014; 9(2):169-176.
4. Coleman TF, Li Y. An Interior, Trust
Region Approach for Nonlinear Minimization Subject to Bounds. SIAM Journal on Optimization. 1996; 6:418–445.