Temperature dependence of R1, R2* and magnetic susceptibility of ferritin at 7T
Mobeen Ali1, Penny Gowland1, and Richard Bowtell1

1SPMIC, School of Physics and Astronomy, University of Nottingham, Nottingham, United Kingdom

Synopsis

Comparison of post mortem and in vivo MR images requires an understanding of the temperature dependence of the NMR parameters that generate relevant image contrast. Here, we therefore evaluated the temperature dependence of the susceptibility and relaxivity of ferritin-doped agar. A phantom containing cylinders doped with different ferritin concentrations was scanned at 7T at temperatures ranging from 5–35⁰C. R1, R2* and field maps were generated and the variation of each parameter with ferritin concentration was evaluated. The variations of susceptibility, R2* and R1 with ferritin concentration all decreased with increasing temperature with R2* showing the strongest temperature dependence.

Background

Magnetic resonance imaging (MRI) of post-mortem tissue is typically carried out with samples chilled or at room temperature. However, the magnetic susceptibility and relaxivities of different tissue components can have different temperature dependencies. This needs to be taken into account when comparing post mortem results to in vivo data.

Ferritin is the main storage molecule for non-haeme iron in many organisms, which is required for respiration and various other metabolic processes [1]. It is one of a few sources of iron known to occur in sufficient concentrations to affect MRI [2,3]. The ability to map iron distributions is clinically useful in brain, liver and heart iron overload diseases, but to validate in vivo methods we need to be able to compare in vivo and post mortem data.

The aim of this study was to characterise the effect of temperature on the NMR behaviour of ferritin at varying concentrations.

Method

Four cylinders of agar were doped with ferritin at concentrations ranging from 0.5 to 3mg/ml. These cylinders were embedded in a spherical agar phantom of 12cm diameter, in a parallel two by two arrangement. The phantom was then placed inside a tank (Figure 1), around which cold water was pumped, using a Tamson Instruments water bath, to achieve a temperature of 5⁰C. The water round the phantom was then warmed in 5⁰C increments to 35⁰C, allowing the temperature to stabilize after each temperature increment, which took approximately 30 minutes each time. The phantom and surrounding tank were scanned at each temperature in a Philips 7T MRI scanner with the cylinders oriented parallel to the field, using a varying inversion time, inversion recovery fast field echo sequence (1.25x1.125x1.25mm3 voxel, FOV=200x180 x72.5mm3, 7 TI values in the range 192-2652ms) and a multi-echo gradient echo sequence (0.94x0.94x1.7mm3 voxel, FOV=210x210x42.5mm3, TE1= 4ms, ΔTE=4ms, 10 echoes).

R1 and R2* values were calculated from an ROI in the centre of each tube (Figure 2). The signal magnitude from the multi GE sequence was linearly fitted to measure R2* and the polarity restored magnitude data from the inversion recovery was fitted to obtain R1. The phase evolution with TE from an ROI inside each tube was compared to that in an annulus around the tube to calculate the field offset inside the tube which is given by $$\triangle B_{int}=\frac{B_0 \triangle \chi}{2}(\cos^2\theta-\frac{1}{3})$$ where B0 is the scanner field and θ is the angle of the tube to B0. This was used to calculate the susceptibility difference, Δχ.

Results

The susceptibility of each cylinder increased linearly with increasing ferritin concentration with a constant of proportionality of 0.88ppm per mg Fe per gram agar at 293K and decreased with increasing temperature. The gradient of the susceptibility variation with concentration, increased with inverse of the temperature (Figure 3) as expected.

Figure 4 shows that R2* decreased for higher concentrations and increased for lower concentrations with increasing temperature so the R2* values showed less variation with concentration. The R2* of the undoped agar around the tubes was found to increase with temperature, with a similar increase found over multiple ROIs. The gradient of R2* with ferritin concentration was used to calculate the R2* relaxivity which increased significantly with the inverse of the temperature (changing by nearly a factor of 2 over the temperature range 5-35⁰C).

R1 for the ferritin samples decreased with increasing temperature, as did R1 for the undoped agar as shown in Figure 5. Ferritin-induced R1 relaxivity also increased with the inverse of the temperature.

Discussion

Phantoms and support platforms for post-mortem samples are usually made from agar and this study confirms previous work that showed that the T2 of agar decreases with temperature [4], and showed that agar T1 increases with temperature.

In ferritin-doped agar gels R1 and R2* are both sensitive to the magnetic susceptibility effects of iron. Ferritin contains a superparamagnetic core whose average magnetization has a linear dependence on inverse temperature over the temperature range studied here [1] as confirmed by the results in Figure 3. The more significant increase in R2* with inverse temperature may be explained by the additional temperature sensitivity of water diffusion rates.

Conclusion

The magnetic susceptibility variation with ferritin concentration is 10% higher at 5⁰C compared to 35⁰C. R1 and R2* change much more significantly over this temperature range potentially leading to significant discrepancies between post mortem and in vivo measurements.

Acknowledgements

Financial support from the MRC and The University of Nottingham.

References

1. Brooks et al., Relaxometry and Magnetometry of Ferritin, MRM 40227-235 (1998)

2. Schenck and Zimmerman, High-field magnetic resonance imaging of brain iron: birth of a biomarker? NMR in Biomedicine, 17(7):433–445, 2004.

3. Haacke, et al. Imaging iron stores in the brain using magnetic resonance imaging. Magnetic resonance imaging, 23(1):1–25, 2005.

4. Vre et al., The Use of Agar Gel as a Basic Reference Material for Calibrating Relaxation Times and Imaging Parameters, MRM 2, 176-179 (1985)

Figures

Figure1: Sample holder built to fit in the RF head coil. This was attached to a circulating water bath, which circulated water in pipes (indicated by arrows) around the phantom. It included a port for a fibre optic thermometer.

Figure 2: T1 map at inversion time 1621ms. Red squares and blue annuli centred on cylinders with concentration increasing from 0.5mg/ml (top right), 1mg/ml, 2mg/ml clockwise to 3mg/ml (top left). Foam blocks to hold phantom in centre of scanning volume appear as dark regions, above, below, left and right.

Figure 3: Susceptibility variation with temperature for different ferritin concentrations (left). Gradient of susceptibility with ferritin concentration plotted against inverse of temperature (right).

Figure 4: R2* variation with temperature in cylinders doped with different ferritin concentrations and in surrounding agar(left). Ferritin R2* relaxivity plotted against inverse of temperature (right).

Figure 5: R1 variation with temperature in cylinders doped with different ferritin concentrations and in surrounding agar (left). Ferritin R1 relaxivity plotted against inverse of temperature (right).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1528