Improved Image Quality when estimating Perfusion Parameters using Bayesian Fitting Algorithm
Irene Klærke Mikkelsen1, Anna Tietze1,2, Lars Ribe1, Anne Obel3, Mikkel Bo Hansen1, and Kim Mouridsen1

1CFIN, Aarhus University, Aarhus, Denmark, 2Dept. of Neuroradiology, Aarhus University Hospital, Aarhus, Denmark, 3Neuroradiology, Aarhus University Hospital, Aarhus, Denmark

Synopsis

Dynamic Contrast Enhanced Perfusion Imaging (DCE) allows for quantification of the blood-brain barrier integrity in tumor patients. A key post-processing step is to fit a hemodynamic model to DCE data. The fitting procedure can, however, cause spurious voxels and image degradation. We compared the widely used Levenberg-Marquardt fitting algorithm to a Bayesian algorithm. Image quality was assessed in 42 tumor patients. The Bayesian approach provided the highest image quality scores. This was confirmed in simulated data with fewer outliers (spurious voxels) when using the Bayesian approach. The hemodynamic two-compartment model that separates cerebral blood flow and leakage, provides reliable Ve images, when the robust Bayesian fitting algorithm is used.

Background

The most widely used curve-fitting algorithm is the Levenberg-Marquardt algorithm (LM)1, but Bayesian modeling (Bayes) is an alternative approach2 not previously used for DCE. We fit patient- and simulated data by two different hemodynamic models using each of the two fitting algorithms, with the aim to improve parameter accuracy and image quality.

Among the hemodynamic models, the Tofts Extended Model (TOFTs) is the most commonly used. Its leakage parameter, Ktrans, is however hampered by the cerebral blood flow, CBF, especially in non-leaky healthy tissue. On the contrary, the two-compartment exchange model (2CX) is potentially able to separate leakage, K1, and CBF, but K1 is not well-determined in non-leaky tissue. We suggest that a more robust Bayesian fitting algorithm will provide more accurate values in both healthy and diseased tissue.

Methods

Hemodynamic parameters were estimated in 42 glioma patients using both the TOFTs and the 2CX models with each of the two fitting-algorithms, LM and Bayes. Patlak estimates3 were used as the starting guess for the fitting routines. The same approach was pursued in 10.000 simulated data sets for each of 4 disease stages and healthy tissue. Scanner parameters and noise levels were used for the simulations. The hemodynamic parameter maps from the patients were visually rated by two neuro-radiologists. The parameter estimates of simulations were compared to the underlying true values (see fig. 3).

Results

The TOFTs Ktrans estimates have confounding high values from the vessels (see fig. 1 A and B). These are vanishing in K1 of 2CX (fig. 1 C and D). The vessel contribution is instead visible in the CBF estimates (fig 1 E and F). Using Bayes fitting algorithm (right column) leads to clear separation of healthy and leaky tissue in the K1 image (fig 1 D). And the number of spurious voxels in the maps of extra-vascular volume, Ve, (fig. 1H) is reduced compared to LM (fig. 1G). The average rater scores for K1 and Ve were higher for Bayes compared to LM, see fig. 2.

In simulations, TOFTs Ktrans estimates reproduced neither K1 nor CBF in no- or mildly diseased tissue, regardless of fitting algorithm (fig. 3A). Opposing this, 2CX largely reproduced the underlying K1, CBF and Ve estimates for all disease stages (fig. 3B-D). In simulated healthy tissue, outliers were seen for K1 for both fitting algorithms, whereas outliers in Ve only were seen when using LM.

Conclusion

Only the 2CX model separates blood flow and leakage for all disease stages. When optimizing the 2CX model parameters, the Bayesian algorithm offers similar accuracy as the Levenberg-Marquardt algorithm with fewer outliers resulting in less image scatter and improved image rating. The Bayesian algorithm may be what is needed for the 2CX model to become clinical applicable.

Acknowledgements

No acknowledgement found.

References

[1] Marquardt DW. An algorithm for least-squares estimation of non-linear parameters J Soc Ind Appl Math. 1963;11(2):431-41.

[2] Mouridsen K, Friston K, Hjort N, Gyldensted L, Østergaard L, Kiebel S. Bayesian estimation of cerebral perfusion using a physiological model of microvasculature. Neuroimage. 2006;33(2):570-9.

[3] C. S. Patlak, R. G. Blasberg, J. D. Fenstermacher (March 1983). "Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data". J Cerebr Blood F Met 3 (1): 1–7

Figures

Hemodynamic parameters from DCE experiment, Ktrans, K1, CBF and Ve when fitting either Tofts Model or the 2CX model to DCE data using either of two fitting algorithms, LM and Bayes. Note how the spurious voxels are eliminated in the Ve map when Bayes algorithm is used. Note also, how 2CX is able to separate CBF and K1.

Rater scores of image quality in two hemodynamic maps, leakage K1, and extra vascular volume, Ve: 1: not assessable 2:not interpretable due to severe image degradation 3:considerable image degradation limiting interpretation 4:good quality 5:excellent quality. Blue: Levenberg Marquardt. Yellow: Bayes algorithm

Simulated parameters. A: Ktrans from Tofts model. B-D: K1, CBF and Ve from 2CX model. The boxplot demonstrates the distribution of estimated parameters. The colored line: green, cyan or magenta indicate the true underlying value. Blue: LM. Yellow: Bayes. Tofts model is unable to separate CBF and K1. LM and Bayes provide similar accuracy and precision in 2CX parameters, except for Ve in healthy tissue, where Bayes is without outliers.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1477