4D Spiral Flow in MR compatible Spinal Canal Phantom with and without Occlusion
Matthew Lee Dobson1, Bryan Gootee1, Michael Kendrick2, Robert Bert3, MJ Negahdar1, and Amir Amini1

1Electrical and Computer Engineering, University of Louisville, Louisville, KY, United States, 2VA Medical Center, Louisville, KY, United States, 3Department of Radiology, University of Louisville, Louisville, KY, United States

Synopsis

A 3’ clear polycarbonate tube together with a dowel rod extending the entire length of the tube, centered in the middle of the tube, were used to model the spinal canal and the spinal cord. Normal saline solution was used to mimic the Cerebrospinal (CSF) fluid. The dowel rod was centered with winged support structures that were 3D printed from a CAD model. A spinal canal occlude was also 3D printed. 4D flow MR imaging was performed and results indicate that the flow phantom has utility for validation and testing of MR methods for measurement of CSF flow.

Introduction

We present description and design of a CSF flow phantom which should prove useful in development and validation of flow imaging protocols for measurement of CSF flow. Numerous diseases and pathologies hamper flow of the CSF including hydrocephalus, Chiari malformation, and spinal cord injury.

Phantom Design

A 3 feet, .625” inside diameter, .75” outside diameter clear polycarbonate tube, was used as an idealized model of the spinal canal (figure 1). A clear tube was chosen to allow for visual inspection of the Phantom for bubbles and overseeing of the flow of fluid. The cord was modeled by a wooden .3125” diameter dowel rod which ran the entire extent of the tube. These dimensions closely match reported diameter ranges of the human spinal canal (~0.67’’) and spinal cord (~0.31’’). To center the dowel rod inside the canal, winged support structures were created in SolidWorks (figure 2a) and 3D printed to the exact dimension of the polycarbonate tube with a hole having the diameter of the dowel rod precisely centered. These support structures were placed on the ends with a third located 1ft from one end of the phantom. The winged support structures were then glued to the dowel rod. The support structures were placed at these locations so that the flow of the fluid would not be obstructed, and so that the flow would be fully developed distal to the structures where imaging was performed. Two dowel rods were utilized to mimic both a healthy spine and one with a stenosis. The occlusion modeling a 75% area stenosis in the spine was constructed using 3D printing (figure 2b) and glued to the second dowel rod. Experiments were carried out using a closed-loop flow system (figure 2c). An MR-compatible, computer-controlled pump (LB Pump; LB Technology LLC, Louisville, KY) was used with the capability to program user-defined flow waveforms. All flow connectors made use of common PVC couplings and connectors so that a pumping system could be connected to the phantom. It was also designed with screw off connectors for easy changing of the two spinal cords.

MR imaging

Imaging was performed on a Philips Achieva 1.5T scanner (Philips Healthcare, Best, NL) using an 8 element SENSE knee coil. To measure the velocity two 4D flow MRI methods were utilized: conventional 4D flow and 4D Spiral flow [1]. For the conventional 4D flow acquisitions, scan parameters were as follows: FOV = 80*80*60 mm3, TR = 11 ms, TE = 7 ms, tip angle = 6o, Venc = 20-60 cm/s for all 3 directions, in plane resolutions = 1*1 mm2, slice thickness = 3 mm, number of slices=20. For 4D Spiral flow, all parameters were identical other than TE = 3.1 ms and TR = 9.9 ms. In order to provide k-space coverage, 36 spiral arms with 4ms readout each were adopted. Image acquisition time for steady flow were (4D conventional: 1:24, 4D spiral: 0:46) and for pulsatile flow were (4D conventional: 27:46, 4D spiral: 15:38).

Results

Table 1 shows peak velocity and flow rate for steady flows at 4, 8, and 12 ml/s. Results illustrate that both acquisition methods measure velocity profiles with similar accuracy, however there is a light underestimation in Spiral acquisition. Flow rate variation inside the phantom through z direction was measured as an indicator of noise in the acquisition. Results show that 4D Spiral flow is more robust encountering phase contrast noise.

Conclusion

In this work we have reported initial results of using two 4D flow techniques on two CSF flow phantoms designed. At these low velocity and low flow rates, results indicate good agreement between the two 4D flow methods, however there is a slight underestimation of velocities with 4D Spiral Flow. It should be mentioned that relative to 4D conventional flow, 4D spiral flow reduced the scan time by around 40%.

Acknowledgements

No acknowledgement found.

References

1.MJ Negahdar, Mo Kadbi, M. Kendrick, R. Longaker, M. Stoddard, AA Amini, “4D Spiral Imaging of Flows in Stenotic Phantoms and Subjects with Aortic Stenosis, Magnetic Resonance in Medicine, in press, 2015, DOI: 10.1002/mrm.25636.

Figures

Figure 1. Clear Polycarbonate tube with a wooden dowel rod in the center as an idealized model of spinal canal.

Figure 2. a) Winged support structures for centering the dowel rods – 3 were used inside the tube. B) A 75% area flow occluder. C) The flow circuit, including a programmable pump capable of creating steady and pulsatile flows.

Figure 3. a) Mesh plot of velocity magnitudes in the flow at middle of phantom with no occlusion using 4D Spiral flow at Q= 12ml/s. b) Velocity pattern through phantom with the 75% area occlusion (imaged with 4D Spiral flow).

Figure 4. Measured pulsatile flow in the phantom.

Table 1. peak velocity and flow rate for steady flows at 4, 8, and 12 ml/s.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1448