Haisam Islam1, Christine Law2, Sean Mackey3, and Gary Glover4
1Bioengineering, Stanford University, Stanford, CA, United States, 2Stanford University, Stanford, CA, United States, 3Anesthesiology, Perioperative, and Pain Medicine, Stanford University School of Medicine, Stanford, CA, United States, 4Radiology, Stanford University, Stanford, CA, United States
Synopsis
Simultaneous functional
imaging of the brain and spinal cord would provide valuable understanding of
neural information processing. However, this is challenging due to the poor
field homogeneity of the spinal cord as well as the typically high spatial
resolution desired for it. The higher-order shims available on most scanners
are static, and thus cannot switch rapidly between brain and spinal cord
acquisitions Here, we use a dynamic slice-based shim for brain slices and a volume-based
shim + reduced FOV acquisition for a neck volume to perform simultaneous functional
imaging of both structures of interest.Target Audience
Researchers
and clinicians interested in simultaneous functional imaging of the spinal-cord
and brain
Purpose
Simultaneous functional
brain and spinal-cord MRI would be important for understanding how neural information is processed but is challenging due to poor $$$B_0$$$ homogeneity in the neck. This is exacerbated
by the small size of the spinal-cord, which prompts a higher spatial resolution
[1]. $$$B_0$$$ homogeneity can be improved by shimming [2],
but the higher-order shims on most scanners are static and limited to second-
or third-order, making it difficult to shim the brain and spinal-cord
simultaneously. One approach has been to vary the shims dynamically, i.e. slice-specific
linear shims and frequency offset between the two volumes [3], and perform
reduced FOV imaging using a 2D slice-selective RF pulse to shorten the readout
length [4]. There are problems with this approach, however. 1) Due to the (in-plane)
symmetry of the neck, varying the linear shims has limited effect. 2) a 2D
spinal-cord slice-selective RF pulse is too long and highly susceptible to
off-resonance. 3) A volume-based shim trades-off homogeneity in one region for
another, and may degrade image quality in the brain. Here, we propose
simultaneous brain and spinal-cord functional imaging using: 1) slice-based
shim + 2D imaging for the brain 2) volume-based shim + 2D slab-selective RF
pulse + 3D imaging for the spinal-cord with EPI readout direction orthogonal to
the RF pulse dimensions. The slice-based shim reduces trade-off in homogeneity
[5]. The purpose of 3D imaging in the spinal-cord is to reduce the length of
the RF pulse, and hence susceptibility to off-resonance. The orthogonal EPI
readout direction allows aliasing to be prevented in the non-selective
dimension of the RF pulse.
Methods
The static shims on our
scanner are the second-order spherical harmonics ($$$xy$$$, $$$zy$$$, $$$zx$$$, $$$x^2-y^2$$$, and $$$z^2-(x^2+y^2)/2$$$), and the dynamic
shims are the linear gradients ($$$x$$$, $$$y$$$, $$$z$$$) and a frequency
offset $$$\Delta f$$$. We minimize the total
off-resonance (in the sum-of-squares sense) over each voxel in a mask region
over the brain and spinal-cord using these shims, where one set of static shims
is used for the entire scan, and a separate set of dynamic shims for each slice
and for the neck volume. The $$$zx$$$ and $$$zy$$$ coils were not used for the static shim and
the shim was not used for the slices since they
were (completely or mostly) redundant and resulted in ill-conditioning of the
problem. Due to imperfections in the second-order shims, i.e. they do not
generate fields that correspond exactly to the spherical harmonics, we modeled
each one in terms of all the shims (static and dynamic), and calculated the
shim amplitudes required to produce the desired field most accurately (in the
sum-of-squares sense).
To test the excitation, we used a 2DFT readout with the
following sequence parameters common to head and neck: matrix-size $$$64 \times 64$$$, readout bandwidth $$$BW_r = \pm 125$$$ kHz, echo-time $$$TE = 10$$$ ms, repetition-time $$$TR = 100$$$ ms. For the head, we used: slice thickness $$$\Delta z = 4$$$ mm, $$$FOV_{head} = 22$$$ cm, # slices $N_s = 6$. For the neck, we
used: slab-thickness $$$D_z = 4$$$ cm, slab-width $$$D_y = 4$$$ cm, time-bandwidths $$$TB_z = TB_y = 4$$$, $$$FOV_x = 12$$$ cm, $$$FOV_y = 6$$$ cm, $$$FOV_z = 5$$$ cm, # z phase-encodes $$$N_z = 6$$$.
Discussion
The
dynamic shim improves the field in
the brain but has limited effect in the neck (see Figure 1). Using the dynamic
shim and a 2D EPI RF pulse, we achieved reasonable selectivity in the neck. We
acquired a slightly larger FOV in z (factor of 1.25) to reduce aliasing
effects. Ideally, the excited region would have constant width (and thickness),
but this may not be the case due to off-resonance. However,
since only the spinal-cord is of interest, imperfect excitation outside it in
the frequency-encode direction is not problematic. To avoid aliasing in the
frequency-encode direction, we increased the readout bandwidth and reconstructed
a larger FOV, since the traditional method of applying an anti-aliasing filter
does not work with ramp-sampling. Due to alternating between acquiring a 3D
volume and other acquisitions, the time between each z phase-encode in the neck
is not constant, i.e. the spins are more relaxed during the earlier phase
encodes. This may be addressed by varying the flip-angles to yield the maximum
total signal or interleaving head slices and neck phase-encodes to allow
greater $$$T_1$$$ relaxation.
We plan to apply this method to simultaneous brain + spinal cord fMRI using a
sensory-motor task with GRAPPA-accelerated EPI.
Acknowledgements
General
Electric Healthcare. NIH Grant: P41 EB0015891, R01 NS053961, K24 DA029262. ambhir-RSL Innovation Challenge GrantReferences
1. Stroman
PW. Clin Med & Research 3.3:146-156 (2005). 2. Kim D. Magn Reson Med
48:715-722 (2002) 3. Finsterbusch J. NeuroImage 79:153-161 (2013)
4. Pauly J. J Magn Reson 81:43-56 (1989) 5. Glover GH. Magn Reson Med
42:290-299 (1999).