Synopsis
The
usefulness of non-Gaussian diffusion parameters for the clinical diagnostic
ability in the head and neck was evaluated. 135 (62 malignant/63 benign/10
inflammation) patients were prospectively recruited, and non-Gaussian DWI and
IVIM parameters as well as synthetic ADC, which comprises both Gaussian and
non-Gaussian effect, were estimated from the DWI datasets with multiple b values.
Significant difference in each parameter was observed between malignant and
benign lesions. There was a significant difference between inflammation or
lymphatic vascular malformation and tumors in K or fIVIM values, providing the
potential to improve the DWI diagnostic accuracy by complementing their
diagnostic abilities.Introduction
Several
studies have reported the usefulness of DWI in the characterization of head and
neck tumors (1). DWI has been widely applied to the diagnosis and monitoring of
head and neck lesions, however, their DWI (mainly ADC) and IVIM parameters
varies between different intuitions and scanners (2), and their diagnostic
accuracy is still limited. Therefore, there is a need to establish the gold
standard in DWI and IVIM parameters to allow the robust comparison between the
scanners and improve the diagnostic accuracy beyond conventional ADC. Recently,
some authors have noted the importance of non-Gaussian DWI in the head and neck
tumor as the additional diagnostic tool to complement ADC (3). Based on these encouraging
results, our purpose was to evaluate the usefulness of non-Gaussian diffusion
parameters for the clinical diagnostic ability in head and neck.
Materials and Methods
This IRB
approved prospective study included 135 (62 malignant/63 benign/10 inflammation)
patients suspected of head and neck tumors. Head and neck MRI was performed
using a 3-T system (Skyra, Trio Tim; Siemens AG) or a 1.5-T system (Avanto; Siemens AG) equipped with a
dedicated head and neck coil. The following images were obtained after localizers
were acquired:
1. As the image quality of DWI for head and neck lesion suffers
from severe image distortion due to susceptibility artifact, a read-out
segmented EPI (RS-EPI) sequence combined with GRAPPA parallel acquisition and
2D-navigator-based reacquisition was used (4,5) with the following parameters: 9 b values of 0, 75, 150, 300, 600, 1000, 1400, 1800, 2200 sec/mm2; TR/TE 2,000/65 ms, FOV 220×220 mm2, matrix 148×148, slice thickness 5.0 mm, 5 readout
segments, parallel imaging factor 2, bandwidth 938 Hz/Px, echo spacing 0.36 ms and scan time 5 min 8 sec.
2. T1-weighted
image: matrix size 256x256, FOV 180x180 mm2, section thickness 4.0 mm; 20
sections without gap, TR/TE 700/12 ms.
3.
ROIs were
placed onto the lesion and images processing was performed using
software implemented in Matlab (Mathwork, Natick, MA) comprising the following
steps:
1/The diffusion signal acquired
with b>300 s/mm² was fitted using the kurtosis diffusion model to estimate
ADCo and K:
S/So = exp[-bADCo + K(bADCo)²/6] [2]
2/Then, the
fitted diffusion signal component was subtracted from the corrected raw signal
acquired with b<300s/mm² and the remaining signal was fitted using the IVIM
model (6) to get estimates of the flowing blood fraction, fIVIM, and
the pseudodiffusion, D*.
3/Synthetic
ADC encompassing both Gaussian and non-Gaussian diffusion effects (2), LbsADC-Hb,
was defined using only 2 b values as:
sADCLb-Hb, = ln [S(Lb)/S(Hb)]/Hb-Lb [3]
where Lb is a low “key” b value and Hb is high “key” b
value are selected to provide the highest sensitivity to non-Gaussian
diffusion (2). We tried 3 combinations for Lb and Hb: 0-1400, 300-1400 and
0-600 s/mm² (equivalent to a standard ADC).
Results
The DWI and IVIM parameters in malignant and benign lesions, inflammation and LVM (Lymphatic
Vascular Malformations) are shown in Table 1 and Table 2. All parameters showed
the significant difference for differentiating malignant from benign lesions,
and their AUC performance is shown in Figure 1. Regarding
sADC and ADCo
parameters, a significant difference was observed between
malignant and benign lesions (
p<0.01), benign lesions and
inflammation (
p<0.01), and
malignat lesions and
LVM. sADC0-1400 and sADC0-600 values in inflammation
was significantly lower
than in LVM (
p<0.05 and
p<0.01). Interestingly,
K value in inflammation was significantly higher than that of malignant or
benign lesions (
p<0.05 and
p<0.01). The combinations of low K and low
ADCo values were found in eight malignant lesions, but not in inflammation.
fIVIM in
LVM was significantly higher than malignant or benign lesions (w/o LVM)
(
p<0.01 and
p<0.01),
Overall,
the AUC performance of sADC0-1400 was highest
as 0.84 with 74.2% sensitivity and 80.3% specificity.
Discussion and Conclusion
AUC
of diffusion parameters for differentiating malignant and benign lesions was
reasonable, considering the data derived from different scanners. Distortion artifacts
were remarkably improved with RS-EPI in most cases, which enabled more
approximate estimation. The combinations of both low K and ADCo values in some
malignant cases suggest the different tumor characteristics other than
inflammation. Higher fIVIM in LVM might explain higher perfusion
fraction. Although more
work is needed to conclude these results, non-Gaussian DWI and IVIM parameters
have the potential to improve the diagnostic accuracy by complementing their
diagnostic abilities.
Acknowledgements
This work was supported
by Hakubi Project of Kyoto University and JSPS KAKENHI Grant.
The authors would like to thank Masayuki
Nakagawa, Taisuke Nagao, Katsutoshi Murata and Yuta Urushibata for the support in acquiring the
data.
References
(1) Chawla S et al. Diffusion-weighted imaging in head and neck cancers. Future Oncol. 2009;5;959-975.
(2) Iima M et al. Clinical Intravoxel Incoherent Motion and Diffusion MR Imaging: Past, Present and Future. Radiology. (in press, Dec 2015)
(3) Yuan J et al. Non-Gaussian Analysis of Diffusion Weighted Imaging in Head and Neck at 3T: A Pilot Study in Patients with Nasopharyngeal Carcinoma. PLoS One. 2014:23;9:e87024.
(4) Iima M et al. Reduced-distortion diffusion MRI of the craniovertebral junction. AJNR. 2012;33:1321-1325.
(5) Koyasu S et al. The clinical utility of reduced-distortion readout-segmented echo-planar imaging in the head and neck region: initial experience. European Radiology. 2014;24:3088-96.
(6) Iima M et al. Quantitative Non-Gaussian Diffusion and Intravoxel Incoherent Motion Magnetic Resonance Imaging: Differentiation of Malignant and Benign Breast Lesions. Investigative Radiology. 2015;50:205-11.