Zackary I. Cleveland1,2, Jinbang Guo1,3, Teckla Akinyi1,2, Hongjiang Wei4, S. Sivaram Kaushik5, Jason C. Woods1,3, Chunlei Liu4, Vivian S. Lee6, and Luke Xie6
11) Center for Pulmonary Imaging Research, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, United States, 22) Department of Biomedical, Chemical, and Environmental Engineering, University of Cincinnati, Cincinnati, OH, United States, 33) Department of Physics, Washington University, St. Louis, MO, United States, 4Brain Imaging and Analysis Center, Duke University Medical Center, Durham, NC, United States, 5Medical College of Wisconsin, Milwaukee, WI, United States, 66) Utah Center for Advanced Imaging Research, University of Utah, Salt Lake City, UT, United States
Synopsis
Magnetic susceptibility differences at gas-tissue
interfaces within the lungs have long been considered a significant obstacle to
performing high-resolution pulmonary MRI because of the resulting rapid T2*
relaxation. However, susceptibility differences in the lungs originate from
regional differences in blood oxygenation and alveolar O2 content. Thus,
if these differences are mapped, they have the potential to provide fundamental
information about regional lung function. Here we demonstrate that quantitative
susceptibility mapping (QSM) of the lungs is possible in vivo using multi-echo radial MRI. Additionally, we demonstrate that the
contrast observed in the lungs via QSM is sensitive to O2 partial
pressure.Introduction
The presence of gas-tissue interfaces at the
alveolar surface has historically been viewed a substantial obstacle to pulmonary
MR imaging. This is due to the >9 ppm susceptibility difference between
diamagnetic tissue and paramagnetic O2 in the alveolar airspaces that
generates a short T
2* of 1 ms or less, depending on field
strength (1,2). Despite
these challenges, the magnetic susceptibility of lung tissue should be
considered in the context of pulmonary function—namely gas exchange and blood
oxygenation. That is, the susceptibility difference between alveolar air and
adjacent pulmonary tissues is determined by the regional partial pressure of O
2,
which is of central physiological importance. Similarly, deoxyhemoglobin
(paramagnetic, with 4 unpaired Fe
3+ electrons) becomes diamagnetic when
oxygenated, a process that occurs efficiently only in regions where ventilation
and pulmonary perfusion are properly matched. Thus, mapping the magnetic susceptibility
of the lung tissue has the potential to provide fundamental insights into lung
function in health and disease. A promising approach for overcoming the
intrinsically short T
2* of lung tissue, while
simultaneously measuring its magnetic properties, is to perform quantitative
susceptibility mapping (QSM) with short,
multi-echo radial MRI (3). Here we demonstrate that: 1) when obtained with
sufficiently short echo times, QSM is able to transform the typically negative T
2*
contrast observed in lungs into positive susceptibility signal and 2) the
intensity of the positive susceptibility signal in the lungs is sensitive to O
2
partial pressure.
Methods
Animals: Free breathing C57BL/6 mice (n=6, ~25 g) were anesthetized
with isoflurane mixed with medical grade air or pure O2 gas. Respiratory
motion was tracked with a pressure bellows placed on the abdomen, and body
temperature was maintained at ~36°C with flowing warm air.
MRI: Images were acquired at end expiration using a 7T
Bruker Biospec scanner and a custom-built quadrature birdcage coil. Imaging was
performed with an RF-spoiled, 3D multi-echo radial sequence that employed interleaved,
golden-angle trajectories. Imaging parameters include: radial views=51472; TR=9
ms; TEs=80, 200, 300, 400 and 500 μs; BW=278 kHz, FOV=40×40×60 mm3,
matrix=128×128×128, resolution=313×313×469 μm3. For image
reconstruction, k-space data were re-gridded using an iterative algorithm (4,5).
Quantitative susceptibility
mapping: Phase was calculated from re-gridded
k-space data. The tissue phase was calculated using Laplacian-based phase
unwrapping and background phase removal algorithm called V-SHARP (6). The tissue phase was then inverted using the two-level
QSM reconstruction algorithm, STAR-QSM (7). Finally, the quantitative susceptibility maps obtained
from each of the echoes were summed to enhance susceptibility SNR.
Image analysis: The lung volume within the thoracic cavity was manually
segmented from the body using Amira. Subsequent analyses were performed in MATLAB
and ImageJ. Within the summed echo image, a region of interest (ROI) was
measured in the skeletal muscle near the lungs to provide a reference
susceptibility value. The susceptible value within this ROI was then subtracted
from susceptibility values within the lung mask to generate background-subtracted
images.
Results
Relative to surrounding tissues, positive
susceptibility signal was obtained from the lungs of all animals (e.g., see
Fig. 1). When mice breathed air (~20% O2), susceptibility values across
the lung volume varied by ~11 ppm and ranged from approximately -4 to 6 ppm. When
animals breathed 100% O2 rather than air, the mean susceptibility increased
from 1.6 to 2.0 ppm. Additionally, regional heterogeneity in susceptibility
values, as assessed by the coefficient of variation (CV, standard deviation
divided by mean), decreased by ~10% when the animal breathed pure O2.
Discussion
The positive susceptibility signal observed from
lung tissue relative to extra-pulmonary tissues is consistent with the expected
susceptibility shift caused by the presence of paramagnetic molecular O
2
gas. This interpretation of the data is further supported by the ~0.4 ppm
increase in mean susceptibility observed from the lungs when mice breathed pure
O
2 vs air. Note that no effort was made to exclude the pulmonary
vasculature during image analysis. Thus, the simultaneous increase in
susceptibility and decrease in regional heterogeneity (i.e., narrowing of the distribution
of susceptibility values) observed when mice breathed pure O
2 likely
reflects decreased deoxyhemoglobin content in the blood returning to the lungs
via the pulmonary arteries.
Conclusions
Quantitative susceptibility mapping of the lungs
is feasible using MRI sequences with sufficiently short echo times. The
positive contrast observed in lung QSM is sensitive to changes in pulmonary O
2
partial pressure and likely also to blood oxygenation. Together, these results demonstrate
that multi-echo, radial QSM holds the potential to become a powerful probe of
regional pulmonary function.
Acknowledgements
This study was funded by the
National Institutes of Health (NHLBI
R00HL111217-03) and a Cincinnati
Children's Research Foundation Trustee Award.References
1. Chen XJ, Moller HE, Chawla
MS, et al., Spatially resolved measurements of hyperpolarized gas properties in
the lung in vivo. Part II: T2*. Magnet Reson Med
1999;42(4):729-737.
2. Ohno Y, Koyama H, Yoshikawa
T, et al., T2* Measurements of 3-T MRI with Ultrashort
TEs: Capabilities of Pulmonary Function Assessment and Clinical Stage Classification
in Smokers. Am J Roentgenol 2011;197(2):W279-W285.
3. Xie L, Layton AT, Wang N,
et al., Dynamic contrast-enhanced quantitative susceptibility mapping with
ultrashort echo time MRI for evaluating renal function. Am J Physiol Renal
Physiol 2015: doi: 10.1152/ajprenal.00351.2015.
4. Zwart NR, Johnson KO, Pipe
JG. Efficient sample density estimation by combining gridding and an optimized
kernel. Magnet Reson Med 2012;67(3):701-710.
5. Johnson KO, Pipe JG.
Convolution Kernel Design and Efficient Algorithm for Sampling Density
Correction. Magnet Reson Med 2009;61(2):439-447.
6. Li W, Wu B, Liu C.
Quantitative susceptibility mapping of human brain reflects spatial variation
in tissue composition. NeuroImage 2011;55(4):1645-1656.
7. Wei H, Dibb R, Zhou Y, Sun
Y, Xu J, Wang N, Liu C. Streaking artifact reduction for quantitative
susceptibility mapping of sources with large dynamic range. NMR Biomed
2015;28(10):1294-1303.