Synopsis
RASER (rapid acquisition with sequential
excitation and refocusing) is an ultrafast imaging technique based on
spatiotemporal encoding (SPEN). The excitation with a chirp-pulse with a low bandwidth-time product (R-value) introduces blurring in the SPEN
dimension. Superresolution (SR) which removes the blurring fails as a result of
the spatially varying B1-phase
produced by radio-frequency coils at ultrahigh fields. A novel iterative
phase-correction of the SR-algorithm is presented. It is shown that the spatial
resolution and the SNR of blurred RASER images acquired at 7 T are significantly
improved employing phase-corrected SR.Motivation
In RASER (rapid
acquisition with sequential excitation and refocusing) (1,2), a 2D-image in a single shot is
acquired similar to EPI (Fig. 1). The phase-encoding in the echo-planar imaging
(EPI) readout train is replaced spatiotemporal encoding (SPEN) in RASER. A low
bandwidth-time product (
R-value) of
the frequency-swept excitation pulse causes blurring along the SPEN-dimension. Superresolution
(SR) eliminates the blurring providing high spatial resolution (3,4). SR-reconstruction fails in the presence of
spatial phase variations which are inevitably produced by the
B1-fields of all ultra-high
field radio-frequency (RF)-coils. A novel iterative self-calibrated phasing
algorithm, which determines the spatial phase distribution directly from
experimental RASER data, is developed. Benefits of SR-reconstruction, such as,
increased signal-to-noise ratio (SNR) and improved spatial resolution, are
quantified in RASER images acquired at 7 T.
Methods
SPEN in RASER is performed using a frequency-swept
chirp-pulse for excitation in combination with appropriate balancing of the
gradients resulting in a quadratic phase profile imprinted on the transverse
magnetization during the echo readout. Spatial localization of the echo signal
is based on dephasing of the transverse magnetization in regions of large phase
variations outside the vicinity of the vertex of the quadratic phase profile
producing the so-called signal attenuation function. The acquired series of
echoes originates from subsequent locations along the spatial coordinate.
SR-reconstruction can also be described as the matrix
multiplication of the SPEN vector Sx
corresponding to a line in the low-resolution 2D-image at position x in frequency-encoded dimension with
the so-called pulse encoding matrix (PEM) A
(3-5)
$$I_{x} = u S_{x}$$
$$u = {\left(A^{T} A\right)}^{-1}A^{T}$$
generating the high-resolution image vector Ix. The
superscript T represents the conjugate transpose. Spatial phase variations P,
which are not represented in the PEM A,
are obtained iteratively ($$$j \rightarrow j+1$$$) from the experimental data $$$S_{x} = P_x^S \mid S_{x} \mid$$$ and the reconstructed
image $$$I_{x} = P_x^I \mid I_{x} \mid$$$. The elements of PEM Aμν are corrected according
to
$$A_{\mu\nu}^{mod} = P_{x\nu}^{T} A_{\mu\nu} P_{x\nu}$$
$$P_{x,j+1} = P_{x,j}\sqrt{P_x^S P_x^I}$$
The SNR is determined by measuring the standard
deviation in a noise scan acquired with blanked RF-amplifiers and signal
intensity in RASER images of a homogenous saline phantom. Simulations of SNR
are based on the method in reference (6). The theoretical SNR before SR-reconstruction is
derived from the SNR of a spin echo (SE)-EPI image which is normalized by the
square root of the number of phase-encoding steps (5).
Results
Fig. 2 shows RASER and for comparison gradient echo (GRE)
images of a high-resolution LEGO phantom in gray. The labels ‘before’ and
‘after’ correspond to iteration and after convergence
of the phase-correction algorithm applied to the RASER images of each channel
of a 32-channel RF head coil. The spatial phase variation distributions are shown in color. The
reduction of striping artifacts in the RASER images achieved with the
phase-correction algorithm is evident.
Fig.
3 depicts the SNR-values for various R-values
obtained from the experimental RASER data (blue: before SR-reconstruction, green:
after SR-reconstruction) and SE-EPI image (yellow) as well as the theoretical SNR-values
before SR-reconstruction (red). The SNR of the RASER images before
SR-reconstruction (lores) increases with lower R-values confirming the theoretically predicted gains (red). The
SNR after SR-reconstruction (hires) also increases with lower R-values, but is lower than in
low-resolution images. This finding suggest that there are two competing
mechanisms for noise propagation in SR-reconstruction present: first, noise
enhancement as a result of matrix inversion reduces SNR. Secondly, averaging of
the noise weighted by the signal attenuation function across the field-of-view
(FOV) reduces the noise level.
Fig.
4 shows images of the calcarine sulcus of a subject. The RASER images are
reconstructed using the SR-algorithm with self-calibrated phasing. Similar
image quality and spatial resolution are achieved with RASER data acquired with
the three different R-values. For comparison, segmented gradient echo (GE)-EPI,
segmented SE-EPI and T1-weighted
(T1)-GRE were acquired from
the same slice. The yellow box marks the extent and location of the RASER images.
The details of sulci are depicted in the RASER images equally well to the T1-GRE (inverted contrast)
and the SE-EPI images. The GE-EPI provides T2*-weighted
contrast while RASER and SE-EPI produce T2-weighted
images accounting for the different gray-white matter contrast.
Conclusion
In
summary, self-calibrated phasing algorithm which enables SR-reconstruction of
SPEN data exhibiting strong phase variations was developed. The proposed
phase-correction algorithm is shown to substantially improve the quality of
RASER-images acquired at 7 T. It is concluded that this capability is essential
to be able to perform SR-reconstruction of data acquired at ultrahigh magnetic
field providing high spatial resolution and increased SNR.
Acknowledgements
Financial support by
the NIH-grants P41 EB015894 (NIBIB), P30 NS057091 (BTRC), S10 RR026783, R01
EB000331 and R24 MH105998-01 and the WM KECK Foundation is acknowledged.References
1. Chamberlain R, Park JY, Corum C,
Yacoub E, Ugurbil K, Jack CR, Garwood M. RASER: A new ultrafast magnetic
resonance imaging method. Magn Res Med 2007;58(4):794-799.
2. Goerke U, Garwood M, Ugurbil K.
Functional magnetic resonance imaging using RASER. NeuroImage
2011;54(1):350-360.
3. Ben-Eliezer N, Irani M, Frydman L.
Super-Resolved Spatially Encoded Single-Scan 2D MRI. Magn Res Med
2010;63(6):1594-1600.
4. Ben-Eliezer N, Shrot Y, Frydman L.
High-definition, single-scan 2D MRI in inhomogeneous fields using spatial
encoding methods. Magn Reson Imaging 2010;28(1):77-86.
5. Ben-Eliezer N, Shrot Y, Frydman L,
Sodickson DK. Parametric Analysis of the Spatial Resolution and Signal-to-Noise
Ratio in Super-Resolved Spatiotemporally Encoded (SPEN) MRI. Magn Res Med
2014;72(2):418-429.
6. Kellman P, McVeigh ER. Image
reconstruction in SNR units: A general method for SNR measurement. Magn Res Med
2005;54(6):1439-1447.