Jeffrey H Maki1 and Gregory J Wilson1
1Radiology, University of Washington, Seattle, WA, United States
Synopsis
Gadolinium
contrast for CE-MRA is typically injected at a fixed, relatively fast (1.5 –
2.0 mL/s) rate. This results in a peaked
bolus profile such that vascular signal intensity (SI) decays during latter k-space
acquisition, leading to blurring and ringing artifacts. A “tailored” test bolus-based predictive
algorithm was developed to determine a patient-individualized multi-phase injection
to achieve any arbitrary arterial SI “plateau” duration. This technique was tested on 10 patients and
compared to 10 patients receiving a fixed 1.6 mL/s contrast injection. The tailored bolus plateau duration was 24
vs. 9 s (p < 0.01) with only a 20% SI loss.Introduction
High resolution contrast-enhanced MRA (CE-MRA) is performed
during the 1st pass of gadolinium contrast. Typically, the full contrast
dose is injected at 1.5 – 2.0 mL/s, followed by saline flush at the same rate.
Centric image acquisition begins near peak enhancement using fluoroscopic
triggering or a test bolus. Any temporal variation in vascular enhancement during
acquisition (e.g., “peaked” bolus with
subsequent fall off) causes blurring and ringing artifacts in the images. The
purpose of this study was to individually “tailor” the contrast bolus enhancement
profile to be constant over the scan time, with the ultimate goal of improving
CE-MRA image quality.
Methods
With local IRB approval, 20 subjects were
randomized to receive either standard or “tailored” injection (Medrad Spectris
Solaris, Bayer, Whippany, NJ) of single-dose (0.1 mmol/kg) gadoteridol (Bracco,
Princeton, NJ), and resulting signal intensity (SI) profiles were measured for
100 sec to capture pre-Gd baseline, 1st pass, and recirculation. Prior
to each full injection, a test bolus was administered. Standard injections consisted
of non-diluted, single-dose contrast (less test bolus volume) injected at our
institutional standard 1.6 mL/s. Tailored injections utilized a novel optimization
algorithm to determine the best multi-phase (≤ 3
phases) injection of 38 mL diluted (single-dose diluted with saline to 40mL) contrast.
The test boluses consisted of 1 mL pure contrast @ 1.6 mL/s (standard), or 2 mL
diluted contrast @ 2 mL/s (tailored). The optimization utilized linear prediction of
contrast concentration profiles (based on the test bolus response) for arbitrary
injection schemes, and targeted a 20 s SI plateau duration. This was implemented in MATLAB (Mathworks,
Natick, MA). All contrast injections
were followed by 25-30 mL saline flush at the final contrast injection rate.
Signal intensity profiles for all injections were
measured (3T Ingenia, Philips, Best, the Netherlands) using an oblique sagittal,
multi-dynamic, thick slice 3D T1-FFE acquisition covering the entire aorta to eliminate
inflow effects. Time-intensity curves were gathered from a single ROI placed in
the supra-celiac abdominal aorta. After
normalization to baseline, maximum SI and full width at 80% max (FWM80)
SI were measured. Test bolus MR
parameters were: TR/TE/α = 2ms/0.8ms/7o,
temporal resolution ~ 1 s; and full bolus: TR/TE/α = 3.5ms/1.4ms/30o, temporal resolution
~ 1.7 s. Spatial resolution was uniformly
1.8 x.1.8 x 7 mm3.
Results
Figure 1 illustrates the
bolus tailoring concept for a single subject, demonstrating timing bolus
results, calculated 3-phase injection scheme, and resultant full bolus. Ranges for each phase (a-c) of the
patient-specific triphasic tailored bolus volumes and rates were: a) 8.3 +
3 mL @ 3.1 + 0.6 mL/s; b) 16.4 + 7.7 mL @ 1.4 + 0.2 mL/s; and
c) 10.0 + 5.0 mL @ 1.3 + 0.2 mL/s.
The single-phase
injections were markedly peaked, demonstrating a relatively short maximum SI
duration, with mean FWM80 = 8.8 + 3.8 s (Figure 2, left). The bolus-tailoring prediction algorithm worked
well to achieve the desired uniform SI plateau
with significantly increased duration; mean
FWM80 = 23.7 + 2.5 s (p < 0.01) (Figure 2, right). Only a relatively small mean signal loss (19.5%,
p = 0.05) was observed to achieve this much longer bolus duration (Figures 3
and 4).
Discussion
Using a predictive
algorithm with a small contrast test bolus allows for patient-specific enhancement
profile “shaping” to achieve a longer and more uniform plateau than does a typical
single-phase injection. This prolonged SI
plateau comes at the expense of a modest SI decrease (i.e., nearly 3x plateau duration while maintaining >80% of peak SI). This is largely explained by T1 and T2*
mediated sub-linearity of SI vs.
contrast concentration such that significant decrease in contrast injection
rates lead to only minimal SI decrease (1,2).
Of note, we found it
necessary to dilute the contrast for bolus tailoring, as calculated injection
rates for undiluted contrast, particularly the 2nd and 3rd
phase, were quite low (often ~ 0.5-0.8 mL/s).
Because the test bolus is performed at a fixed injection rate, system
linearity can only be assumed for that rate. In preliminary tests without
diluting, we found that slower bolus phases arrived later than predicted, distorting
the expected plateau phase (Figure 5). This was largely remedied in the study by
diluting the contrast as described, keeping injection rates higher where the
system behaved more linearly.
Conclusion
Bolus tailoring for
CE-MRA can be readily achieved and has significant potential to improve image
quality. This suggests the conventional
method of injecting Gd contrast at a relatively fast fixed rate is flawed. Future work will evaluate the impact of
tailored boluses on CE-MRA image quality.
Acknowledgements
This work was in part
funded by a grant from the University of Washington Research Royalty Fund. References
(1) Wilson GJ, Woods M, Springer CS Jr, Bastawrous S,
Bhargava P, Maki JH. Magn Reson Med
2014;72:1746–54.
(2) Wilson GJ, Springer
CS Jr, Woods M, Bastawrous S, Bhargava P, Maki JH. PISMRM, Milan, Italy, 2014, p3862.