Jessica AM Bastiaansen1, Davide Piccini1,2, Ruud B van Heeswijk1,3, and Matthias Stuber1,3
1Department of Radiology, University hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 2Advanced Clinical Imaging Technology, Siemens Healthcare, Lausanne, Switzerland, 3Center for Biomedical Imaging, Lausanne, Switzerland
Synopsis
Large
volume fat suppression is increasingly challenging at high magnetic field
strengths due to B0 and B1 inhomogeneities. In this study, we developed a novel
lipid-insensitive binomial off-resonant (LIBRE) radiofrequency excitation pulse
to achieve near-complete fat suppression in large 3D volumes and applied it to
whole-heart coronary imaging at 3T. In 6 healthy volunteers, we performed free-breathing
self-navigated whole-heart 3D radial coronary MRA, and quantitatively compared
the results to more commonly used methods for lipid nulling. We show that LIBRE
significantly improves the signal nulling of lipid resonances resulting in both
improved blood pool delineation for self-navigation and increased vessel conspicuity
in the final images.Purpose
The
signal suppression of unwanted lipid resonances is increasingly challenging at
high magnetic field strengths due to increased B0 and B1 inhomogeneities. In a
number of specific applications, e.g. coronary MRA, fat suppression is of vital
importance, since vessels are embedded in lipid tissue layers and fat
suppression needs to be homogeneous within a relatively large volume. Not only may
residual off-resonance lipid signals hinder the correct anatomical
visualization of the small-caliber coronary vessels, but the presence of such
signal from static structures (e.g. chest) can also be a major issue when
advanced motion compensation techniques are used. In particular respiratory
self-navigation [1, 2] relies on the precise detection of the left-ventricular
blood pool in a projection image of the whole thorax [3] to perform motion
correction and achieve a 100% scan efficiency in free-breathing coronary
MRA. When fat signal is not completely
suppressed, the blood pool can no longer be isolated and correct tracking
becomes increasingly challenging. The aim of this study was therefore to develop
and test a novel lipid insensitive binomial off-resonant excitation (LIBRE)
pulse for respiratory self-navigated whole-heart coronary MRA at 3T, and to quantitatively
assess the resulting image quality in comparison to more commonly used
lipid-nulling schemes.
Materials and Methods
Coronary MRA was
performed in (n=6) healthy adult volunteers on a 3T clinical scanner (PRISMA,
Siemens Healthcare, Erlangen, Germany). Data were acquired using a prototype respiratory-1D-self-navigated
ECG-triggered free-breathing 3D radial gradient-recalled-echo (GRE) imaging
sequence [3], preceded by an adiabatic T2 preparation module (T2Prep) to
improve blood myocardium contrast [4]. All datasets were 1D motion-corrected,
using the superior-inferior (SI) projection image of the thorax, and
reconstructed at the scanner. Three different scans were performed in each
volunteer: 1) a conventional lipid nulling using spectral presaturation of the
lipid resonances (fat-sat, or FS), 2) a conventional 1-1 binomial water
excitation scheme (WE), 3) the novel lipid insensitive excitation using LIBRE
pulses. The pulse design was based on binomial rectangular pulses which are
both frequency and phase modulated. Imaging parameters were: field-of-view (220
mm)
3, matrix size 192
3, TE
T2-Prep = 40 ms, RF
excitation angle 18°, TE/TR = 2.5 ms/5.6 ms, with 24 to 32 radial readouts per
heartbeat for a total of ~15k k-space lines. To compare the efficacy of the three different lipid nulling
methods, the signal-to-noise ratio (SNR) of blood, myocardial and fat tissue,
the contrast-to-noise ratio (CNR) between blood and myocardium, and between
blood and fat tissue, and the vessel sharpness of the right coronary artery
(RCA) were calculated using SoapBubble [5]. Statistics were computed via
two-tailed student’s t-test for paired data corrected for multiple comparisons.
Finally, the tracking signals of the algorithm for respiratory self-navigation
were compared by visual inspection.
Results and Discussion
Data were
successfully acquired and reconstructed in all volunteers. Representative MRA
images show a clear visual difference between
the three methods (Fig. 1, A to C). Both
RCA and LAD could be clearly visualized using the LIBRE method (Fig. 1, A1 to A3).
In all volunteers, the LIBRE method showed improved lipid nulling in the whole acquired
volume when compared to conventional spectral fat suppression and water
excitation techniques (Fig. 1). The value of all quantitative endpoints was
significantly improved for the LIBRE method, with respect to the two other
acquisition schemes. SNR and CNR measurements showed significant improvements
comparing the LIBRE water excitation method with conventional water excitation
and fat pre saturation (Fig. 2A and 2B). The vessel sharpness of the RCA was
improved by 30% on average (Fig. 2C). These results are most likely due to the
absence of ubiquitous streaking artifacts caused by unsuppressed chest fat. The
effects of the different fat suppression techniques can also be appreciated in sagittal
images (Fig. 3A-C), where the LIBRE excitation pulses achieve a near complete
suppression of the chest fat, thus generating a clean tracking signal for the
self-navigation algorithm (Fig. 3D). Conversely, the suboptimal fat suppression
of the other two techniques resulted either in a dampened tracking due to the
superimposition of static tissue (Fig. 3E), or to unreliable motion detection
due to the lack of sharpness in the delineation of the blood pool (Fig. 3F). The
suboptimal motion detection and correction in FS and WE also contribute to the
lower final image quality.
Conclusion
These
preliminary findings demonstrate that novel LIBRE excitation pulses achieve
improved fat saturation for coronary imaging at 3T, higher CNR of
blood/myocardium and increased vessel sharpness, leading to improved
visualization of the coronary arteries. Respiratory self-navigation may
particularly benefit from the improved efficacy of this novel pulse.
Acknowledgements
Center
for Biomedical Imaging, Lausanne, Switzerland (CIBM), NanoteraReferences
[1] Stehning, C., et al., Free-breathing
whole-heart coronary MRA with 3D radial SSFP and self-navigated image
reconstruction. Magn Reson Med, 2005. 54(2):
p. 476-80.
[2] Piccini, D., et al., Respiratory
self-navigated postcontrast whole-heart coronary MR angiography: initial
experience in patients. Radiology, 2014. 270(2): p. 378-86.
[3] Piccini, D., et al., Respiratory
self-navigation for whole-heart bright-blood coronary MRI: methods for robust
isolation and automatic segmentation of the blood pool. Magn Reson Med,
2012. 68(2): p. 571-9.
[4] Nezafat, R., et al., B1-insensitive
T2 preparation for improved coronary magnetic resonance angiography at 3 T.
Magn Reson Med, 2006. 55(4): p.
858-64.
[5] Etienne, A., et al., "Soap-Bubble"
visualization and quantitative analysis of 3D coronary magnetic resonance
angiograms. Magn Reson Med, 2002. 48(4):
p. 658-66.