Short-T2 specific excitation by a 'back-and-forth' composite RF pulse
Ethan M Johnson1, Adam B Kerr1, Kim Butts Pauly2, and John M Pauly1

1Electrical Engineering, Stanford University, Stanford, CA, United States, 2Radiology, Stanford University, Stanford, CA, United States

Synopsis

Images of cortical bone have previously been created by selection of RF pulse parameters giving short-$$$T_2$$$-specificity in excitation for a 3D UTE sequence. The previous demonstration required multiple excitations. Here a composite pulse is described that creates similar contrast for depicting cortical bone with bright signal.

Introduction

A technique for creating 'MR-sim-CT' images with positive signal in cortical bone by combining a pair of images acquired with a three-dimensional (3D) ultra-short echo-time (UTE) sequence has been demonstrated [1]. The acquisition method alternates application of differently-$$$T_2$$$-sensitised hard excitation pulses and encoding at staggered echo times (TE). Here, a method is presented for composing together the excitation pulses and acquiring a single set of gradient echoes that can be combined to generate similar contrast. Application of the composite pulse for imaging of bone reduces the net scan duration by half.

Theory

The previously-demonstrated method ([1]) is used as reference. It alternates between excitation by a 'fast' RF pulse that excites a broad range of $$$T_2$$$ signals and a 'slow' RF pulse that excites only longer-$$$T_2$$$ signals (Fig.1). The basic principle of setting excitation rate to modulate the response in $$$T_2$$$ is also previously demonstrated for long-$$$T_2$$$-suppression preparation pulses [2,3]. The reference method additionally staggers the echo times to match excitation rates. A magnitude subtraction of echo images create strong $$$T_2^*$$$ sensitivity specific to cortical bone and limited sensitivity to off-resonance phase accumulation. The subtraction is also normalised by the sum to mitigate proton density weighting, since bone has low water content relative to other tissues.

If the pair of excitation pulses is composed together with opposing polarity to create one excitation pulse (Fig.1), some of the $$$T_2$$$-specificity achieved by subtraction can be generated directly in excitation. During the slow 'tip-back' portion of the composite pulse, only longer-$$$T_2$$$ magnetisation is excited; during the fast 'tip-forth' duration all is excited in the opposite direction (Fig.2). For long-$$$T_2$$$ magnetisation the net area cancels, but for short-$$$T_2$$$ it does not. The combination pulse will also, however, tend to excite magnetisation precessing at a frequency different from that of the applied pulse because phase accrued during the pulse interferes with the cancellation effect (Figs.2,3).

Unintended off-resonance excitation can be addressed to some degree by collecting a second echo, which is also required to create the bone-specific image of the reference method. The high sensitivity of the excitation pulse to off-resonance is then mitigated for long-$$$T_2$$$ in the difference image (Fig.3b). The second echo can be acquired during the same repetition interval (TR), so only one excitation is required, and the net acquisition time is reduced by a factor of two.

Methods

A 3D UTE gradient echo sequence having facility for modifying the excitation pulse was used to acquire lower-leg images with a T/R extremity coil performing excitation and signal reception for a 1.5T commercially-available clinical scanner. The images were encoded with isotropic 1.2mm resolution and 29.4cm field-of-view using 250kHz bandwidth and 7.6ms TR, requiring 5.55min. scan time per volume encoding (i.e. 11.1min. total for separated-excitation scans).

Separated-excitation acquisitions applied 1.5$$$\mu$$$T or 24$$$\mu$$$T pulses for 512$$$\mu$$$s or 32$$$\mu$$$s respectively (nominal 12$$$^{\circ}$$$ tip angle) in an interleaved fashion. Combined, back-and-forth pulse acquisitions excited by application of '-1.5$$$\mu$$$T' ($$$-y$$$-direction) for 512$$$\mu$$$s and 24$$$\mu$$$T ($$$+y$$$-direction) for 32$$$\mu$$$s. Gradient echoes at 0.034ms and 2.034ms were recorded in both separated-excitation and composite excitation acquisitions.

Bone images from each acquisition were generated by magnitude difference of the different-pulse/different-echo images and normalisation by their sum. Images were masked to exclude voxels presumed to be air using region-growing classification formed from the late-echo image of each acquisition.

Results

In phantom images acquired with the back-and-forth excitation, short-$$$T_2$$$ phantoms give bright signal (Fig.4). Excitation of the relatively long-$$$T_2$$$ resolution phantom and oil bottle is mostly avoided. However, near the edges of the phantoms, susceptibility-change-induced off-resonance confounds long-$$$T_2$$$-nulling.

Bone-specific images acquired with two differently-$$$T_2$$$-selective excitations show bright signal in cortical bones of the leg (Fig.5), a contrast previously demonstrated and verified as CT-similar [4]. The tibia and fibula are hyperintense relative to soft tissue; other structures between the muscles also show similar contrast. These may be tendons or epimysium and have been shown with hyperintense signal using other methods, such as inversion-recovery UTE [5].

Conclusions

Composing differently-$$$T_2$$$-selective excitation hard pulses into a back-and-forth excitation pulse and acquiring two different-TE gradient echoes allows generation of bone-specific images. The contrast is similar to that given by the reference method [1], but the acquisition time is halved. Because the reference spends only half its scan duration for each image used in the subtraction there is no image SNR penalty to use of the composite pulse, but the soft-tissue suppression is less robust to off-resonance or other confounds. There can be benefits though, such as reduced motion artefacts derived from the shorter duration.

Acknowledgements

NIH: P01 CA159992

References

[1] E. M. Johnson, U. Vyas, K. B. Pauly, J. M. Pauly. "$$$T_2$$$-selective excitation with UTE imaging for bone imaging." ISMRM 23rd Annual Meeting and Exhibition (2015): 2436.

[2] P. E. Z. Larson, P. T. Gurney, K. Nayak, G. E. Gold, J. M. Pauly, D. G. Nishimura. "Designing Long-$$$T_2$$$ Suppression Pulses for Ultrashort Echo Time Imaging." Magnetic Resonance in Medicine (2006) 56:94–103.

[3] P. E. Z. Larson, S. M. Conolly, J. M. Pauly, D. G. Nishimura. "Using Adiabatic Inversion Pulses for Long-$$$T_2$$$ Suppression in Ultrashort Echo Time (UTE) Imaging." Magnetic Resonance in Medicine (2007) 58:952–961.

[4] E. M. Johnson, U. Vyas, P. Ghanouni, K. B. Pauly, J. M. Pauly. "Comparison of Ultrashort TE MRI with CT for MR-guided Focused Ultrasound Planning." ISTU 15th International Symposium of Therapeutic Ultrasound(2015): 78.

[5] J. Du, M. Carl, M. Bydder, A. Takahashi, C. B. Chung, G. M. Bydder. "Qualitative and Quantitative Ultrashort Echo Time (UTE) Imaging of Cortical Bone." Journal of Magnetic Resonance (2010) 207:304–311.

Figures

Fig.1:

(top) The reference method uses separated excitations rated to excite all $$$T_2$$$ magnetisation (first pulse) or avoid exciting short-$$$T_2$$$ magnetisation (second pulse) and staggers echo times.

(bottom) An alternative 'back-and-forth' pulse comprising the two pulses with opposed polarity creates similar short-$$$T_2$$$ specificity in excitation.

Imaging gradients are not shown.


Fig.2:

Simulation of magnetisations evolving with different $$$T_2$$$-relaxations and precession frequencies during the back-and-forth pulse illustrates the excitation effect. Long-$$$T_2$$$ magnetisation is tipped away then back to $$$z$$$. Short-$$$T_2$$$ magnetisation is slightly then significantly tipped, so it undergoes net excitation. Off-resonant magnetisation precesses during the pulse and is also excited.


Fig.3:

(a) The composite pulse excitation profile is similar to the difference in excitation profiles used by the reference method for magnetisation on resonance but significantly differs off-resonance.

(b) Imaging sensitivities from echo subtraction are similar for echoes acquired after the separated pulses and for ones from the back-and-forth pulse.


Fig.4:

UTE images acquired after excitation by separate, differently-$$$T_2$$$-sensitised pulses can be subtracted to show the difference in $$$T_2$$$ sensitisation. The difference in magnitudes is similar for $$$T_2$$$ at any resonance, but the complex difference is sensitive to off-resonance. The back-and-forth composite pulse excitation is similar to the complex difference.


Fig.5:

Axial slices from the volume (left) acquired by either method show similar contrast, highlighting cortical bone and other short-$$$T_2$$$ structures in the lower leg with hyperintense signal. Volume renderings (right) of the leg can be windowed and levelled to primarily show bone and tendons or to include soft tissue.




Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1014