Joshua S. Greer1,2, Yue Zhang2, Christopher Maroules2, Orhan K. Oz2, Ivan Pedrosa2,3, and Ananth J. Madhuranthakam2,3
1Bioengineering, University of Texas at Dallas, Richardson, TX, United States, 2Radiology, UT Southwestern Medical Center, Dallas, TX, United States, 3Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
Flow Alternating Inversion Recovery (FAIR) has been studied
extensively for pulmonary perfusion imaging at 1.5T, but suffers from low SNR,
and is often corrupted by bright signal in the major vasculature and image
misregistration artifacts due to respiratory motion. The purpose of this study was to evaluate FAIR at 3T for increased SNR and compare against SPECT perfusion, to combine FAIR with inflow saturation to reduce signal in the major pulmonary
vessels, and to combine FAIR with background suppression strategies to minimize
artifacts due to image misregistration.Introduction
Arterial spin labeling (ASL) can measure quantitative
perfusion noninvasively using blood as an endogenous tracer. Flow Alternating
Inversion Recovery (FAIR) based ASL has been extensively studied for pulmonary
perfusion imaging at 1.5T in research environments [1-4]. However, the clinical
applications of FAIR have been limited largely due to the low signal to noise
ratio (SNR), combined with bright signal from major pulmonary vasculature and
potential image misregistration.
The purpose of this study was three fold: 1) To
evaluate FAIR at 3T for increased SNR due to higher field strength and
prolonged T1 of blood and compare results against SPECT perfusion, 2) To combine
FAIR with inflow saturation to suppress signal from major pulmonary vasculature,
and 3) To combine FAIR with background suppression to minimize image misregistration
artifacts.
Methods
This was a prospective, HIPAA-compliant feasibility
study. All subjects signed informed consent prior to imaging. The FAIR sequence was implemented on
a 3T Ingenia scanner (Philips Healthcare, Best, The Netherlands). Similar to an
earlier description [5], the sequence began with saturation of the imaging
region, followed by labeling with a pair of selective and non-selective
inversions using a hyperbolic secant pulse. A post-labeling delay of one
cardiac cycle (i.e. R-R interval) [1] was used to allow labeled blood to
perfuse the lungs. A Single-Shot Turbo Spin Echo (SShTSE) acquisition was used
to minimize the susceptibility artifacts due to B
0 inhomogeneities in the
lungs. The beginning of the sequence (i.e. saturation pulse) was ECG-triggered such
that the data acquisition happens during the diastolic phase of the following
cardiac cycle. Coronal and sagittal images were acquired with an in-plane
resolution of 3x3 mm, 15mm slice thickness, TE of 44ms, and a TR of 3s using
three pairs of control/label images in an 18-second breath hold. Quantification
was performed using the perfusion equation for pulsed ASL [6] with the blood
proton density taken as the average signal from an ROI drawn over the aorta in
a separately acquired M
0 image. This sequence was evaluated in a cohort of 10
normal volunteers and compared against SPECT
perfusion, which was performed in the same subjects during a 15-minute
free breathing acquisition, following the injection of
technetium-99m-macroaggregated albumin (Tc99m-MAA). Subsequently, the inflow saturation for the FAIR
sequence was implemented using three equally-spaced saturation pulses applied
over the labeling regions on either side of the imaging plane (fig. 1) during
the final 500 ms prior to data acquisition. This inflow saturation suppresses
the blood signal that flows into major pulmonary vasculature towards the end of
the post-label delay. The FAIR sequence was also combined independently with a background suppression strategy using
four non-selective inversion pulses applied during the post-label delay with optimal
inversion times [7] (fig. 2).
Results
Figure 3 shows proton density, FAIR perfusion, quantified
FAIR perfusion, and SPECT perfusion images in the coronal and sagittal planes
in a representative volunteer. Perfusion was measured to be 508.9 $$$\pm$$$ 291.3 mL/100g/min (mean $$$\pm$$$ SD) across all subjects, consistent with previously
reported values [8]. The FAIR perfusion images show details of the lungs with
higher spatial resolution, but are also contaminated by the large pulmonary
vessels, which are not present in the SPECT images (fig. 3, arrows). The FAIR
perfusion images acquired with inflow saturation contain reduced signal in the
major vessels, including aorta and larger pulmonary vessels, allowing for
better evaluation of pulmonary parenchyma (fig. 4). Figure 5 shows the effect
of background suppression to reduce image misregistration artifacts. Pulmonary
perfusion appears similar between the two acquisition strategies, but
background suppression significantly reduced the artifacts caused by breathing,
seen in the diaphragm (red arrow) and fat (green arrow), and reduced background
signal variations at the edges of the field of
view (black arrow).
Discussion
Inflow saturation and background suppression improved
the quality of FAIR perfusion-weighted images by reducing the signal in the
major vasculature and minimizing image artifacts, respectively. In recent
years, several statistical methods have been proposed to identify signal
contribution from the major vasculature such that they can be eliminated from FAIR
perfusion-weighted images for improved robustness in measuring pulmonary
perfusion [3, 9]. The inflow saturation reported in this work eliminates the
need for such processing, making the FAIR perfusion images equivalent in appearance
to SPECT perfusion images. This inflow saturation, combined with the background
suppression, should further improve FAIR perfusion images to make it a clinically
feasible technique at 3T.
Acknowledgements
This work is partly supported by an RSNA research grant.References
[1] Bolar,
D. S., et al. "Quantification of regional pulmonary blood flow using ASL-FAIRER." Magnetic resonance in medicine 55.6 (2006): 1308-1317.
[2] Mai,
Vu M., and Stuart S. Berr. "MR perfusion imaging of pulmonary parenchyma
using pulsed arterial spin labeling techniques: FAIRER and FAIR."Journal of Magnetic Resonance Imaging 9.3 (1999): 483-487.
[3] Henderson,
A. Cortney, et al. "Characterizing pulmonary blood flow distribution
measured using arterial spin labeling." NMR in biomedicine 22.10 (2009): 1025-1035.
[4] Schraml,
Christina, et al. "Non-invasive pulmonary perfusion assessment in young
patients with cystic fibrosis using an arterial spin labeling MR technique at
1.5 T." Magnetic Resonance Materials in Physics,
Biology and Medicine25.2 (2012): 155-162.
[5] Mai,
V. M., et al. "Pulmonary perfusion using arterial spin labeling at 3 Tesla
MR scanner." Proceedings of the International Society for
Magnetic Resonance in Medicine, Eleventh Scientific Meeting, Toronto, Ontario,
Canada. 2003.
[6] Buxton,
Richard B., et al. "A general kinetic model for quantitative perfusion
imaging with arterial spin labeling." Magnetic resonance in medicine 40.3 (1998): 383-396.
[7] Maleki,
Nasim, Weiying Dai, and David C. Alsop. "Optimization of background
suppression for arterial spin labeling perfusion imaging." Magnetic
Resonance Materials in Physics, Biology and Medicine 25.2 (2012): 127-133.
[8] Wang,
Tungte, et al. "Quantitative perfusion mapping of the human lung using 1H
spin labeling." Journal of Magnetic Resonance Imaging 18.2 (2003): 260-265.
[9] Walker,
Shane C., et al. "A statistical clustering approach to discriminating
perfusion from conduit vessel signal contributions in a pulmonary ASL MR
image." NMR in Biomedicine 28.9 (2015): 1117-1124.