Mario Gilberto Baez Yanez1,2, Phillip Ehses1,3, and Klaus Scheffler1,3
1Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Graduate Training Centre of Neuroscience, Tuebingen, Germany, 3Department of Biomedical Magnetic Resonance, University of Tuebingen, Tuebingen, Germany
Synopsis
The
excellent sensitivity and stability of BOLD-imaging with balanced SSFP (bSSFP)
on humans at 9.4T has been demonstrated in a recent paper. Here, we analyze the
signal change of bSSFP for different vessel (spheres) sizes and susceptibility
differences for different repetition times and flip angles using Monte Carlo
simulations and experiments on micro spheres, and compare it to gradient echo
EPI. Simulated and measured signal changes (using values of susceptibility
changes and vessel sized comparable to a typical BOLD experiment at 9.4T) of
bSSFP are in the range of 10 to 15% with a peak sensitivity to the vessel
(sphere) size at about 3 mm, and a decreased sensitivity for larger
vessels (spheres). For GE-EPI, signal changes are similar to bSSFP, however, no
selectivity to small vessels is visiblePURPOSE
The purpose of this work is to
analyze BOLD-induced signal changes in bSSFP acquisitions at 9.4T, and compare
these results to conventional GE-EPI. The results are based on Monte Carlo
simulations and measurements from micro spheres mimicking a neurovascular
network.
INTRODUCTION
Up to date fMRI using GE-EPI
sequences have been the primary imaging tool to measure the neurovascular
hemodynamic changes associated with neuronal activity
1. The
sensitivity of GE-EPI and SE-EPI has been analyzed in detail using Monte Carlo
simulations and measurements on appropriate phantoms, and in animals and humans
2,3,4. In a recent paper the high quality
and stability of BOLD imaging using bSSFP at 9.4T has been demonstrated
5.
Figure 1 shows a representative example of these measurements for bSSFP, GE-EPI
and SE-EPI with corresponding signal change time courses. In this abstract we
analyze the sensitivity of bSSFP to oxygenation changes, and compare the
results to GE-EPI and SE-EPI. Our analysis is based on Monte Carlo simulations
using spheres with different diameter and concentration as a model for the
neurovascular network. In addition, we performed measurements on a phantom
with micro spheres applying bSSFP, GE-EPI and SE-EPI. Furthermore, we have
analyzed the influence of the repetition time and flip angle of bSSFP on
BOLD-induced signal changes.
METHODS
We created a 3D 2-compartment model filled with a volume fraction of
2%,3% and 4% of random position of spheres. A susceptibility
map correlated with the model was calculated with
the FPM model
6 using the main magnetic field, susceptibility
differences between compartments, oxygenation and hematocrit as parameters.
Diffusion was simulated using 10000 random walkers and the phase gain
was recorded in each time step of 20 ms. GRE, SE and bSSFP were simulated.
We build a spherical phantom
containing of 19 test tubes of 1 cm inner diameter. Samples contain Dy-DTPA with concentrations of 3 and 5 mM, precision
polystyrene microspheres and a reference sample. The phantom was placed in the isocenter of a custom-built head coil
7
(16 transmit / 31 receive channels). Several experiments with different
parameters were performed (bSSFP: TR = 5ms, FA = 5, 10, 15, 20, 25, 30 deg; TR
= 4, 6, 8 ms, FA = 15 deg; GRE EPI: TE= 10, 20, 30 ms). The sequence used for
human brain imaging at 9.4T shown in Fig. 1 are: bSSFP: TR/TE = 5/2.5ms, FA =
15 deg; GE-EPI: TE = 18 ms, SE-EPI: TE = 36 ms, 1 mm isotropic resolution for
all measurements.
RESULTS
For bSSFP, TR and flip angles have been varied between 2-8
ms and 10-70 degrees, respectively. Simulated and measured signal changes are on-resonance,
i.e. at the center of the passband of bSSFP. Figure 2 shows simulated signal changes for different flip
angles as a function of the sphere size for a) TR= 2 ms and b) TR = 8 ms. Gradient echo and spin echo simulations are
shown in Fig. 2 c) and d), respectively, for different echo times and as a
function of sphere size. Both bSSFP and gradient echo signal changes are of
similar range.
Figure 3 shows results obtained
from the measurements with micro spheres in Dy-DTPA solution for bSSFP and
gradient echo for different flip angles and echo times as a function of sphere
size. These measurements confirm the comparable sensitivity of bSSFP and GRE to
microscopic susceptibility at 9.4T
DISCUSSION/CONCLUSION
Our simulations and measurements
on spheres dissolved in Dy-DTPA solution confirm the comparable sensitivity of
bSSFP and GRE to BOLD-related effects for sphere (vessel) sizes around 1-3 mm. For sphere sized above 5 mm GRE shows a much stronger sensitivity to
BOLD than bSSFP. Therefore, the about three-fold higher signal change for
GGE-EPI compared to bSSFP in human data as shown in Fig. 1 can be explained by
the massive contribution of larger vessels. We thus conclude that in the range
of small vessels bSSFP is equally sensitive to GRE, with the advantage to be
more sensitive to this small vessel regime. Therefore, using bSSFP to detect
BOLD changes offers similar sensitivity as GRE but in addition is more
selective to the small vasculature, which is probably closer to the neuronal
event. The simulated and measured results shown in Fig. 2 and 3 also allow for
a further improvement of the sensitivity of bSSFP to the BOLD effect, i.e. to
go for high flip angles and longer TR. A significant advantage of bSSFP over
spin echoes (both are sensitive to the smaller vessels) is the much lower SAR
demand of bSSFP compared to spin echoes. Thus, whole brain coverage with bSSFP
is easily possible with bSSFP at 9.4T.
Acknowledgements
No acknowledgement found.References
1. Kim SG, Ogawa S. Biophysical
and physiological origins of blood oxygenation level-dependent fMRI signals. J
Cereb Bl Flow Metab (2012) 32, 1188-1206
2. Weisskoff et al, Microscopic
Susceptibility Variation and Transverse Relaxation: Theory and Experiment, MRM
(1994) 31:601-610
3. Frohlich et al, Theory of
susceptibility-induced transverse relaxation in the capillary network in the
diffusion narrowing regime, MRM (2005) 53:564-573
4. Miller K Jezzard P. Modeling SSFP Functional MRI contrast in the brain. MRM (2008) 60:661-673
5. Scheffler, K. and Ehses, P. (2015),
High-resolution mapping of neuronal activation with balanced SSFP at 9.4 tesla.
Magn Reson Med. doi: 10.1002/mrm.25890
6. Pathak A et al. A novel technique for
modeling susceptibility based contrast mechanisms for arbitrary microvascular
geometries: the finite perturber method. NeuroIm (2008) 1130-1143
7. Shajan G et al. A 16-channel dual-row
transmit array in combination with a 31-element receive array for human brain
imaging at 9.4T. MRM 72 (2014)