Wyger Brink1 and Andrew Webb1
1Radiology, Leiden University Medical Center, Leiden, Netherlands
Synopsis
Imaging
methods at high fields can suffer from receive non-uniformities from the body
coil, particularly when the body coil is used as a reference for intensity
correction. In this work we show that the DREAM B1 mapping sequence can be used for receive uniformity
correction in RF-shimmed whole-body imaging at 3T.Purpose
Imaging
methods at high fields can suffer from receive non-uniformities from the body
coil, particularly when the body coil is used as a reference for intensity
correction. The
B1 mapping
method DREAM is widely used for rapid mapping of the transmit
B1+ field.
1
Recently, it was shown that also the receive sensitivity (
B1-) can potentially be retrieved from DREAM
data.
2 Using this approach, we evaluate the use of DREAM for
correcting the receive non-uniformities of the body coil in whole body MRI at
3T.
Methods
A whole-body
imaging protocol was performed at different landmarks in a 3T dual-transmit MR
system (Ingenia, Philips Healthcare, Best, the Netherlands). The two channels of the
body coil were driven in RF-shimmed mode, using the vendor supplied RF-shimming
routine. The DREAM method was used for multislice B1 mapping;
in-plane resolution = 7 × 7 mm2, slice thickness = 7 mm, (nominal) STEAM/imaging
tip angle = 10°/50°, TR/TEste/TEfid=3.8/1.05/2.3 ms.
Receive sensitivity (B1-)
maps were obtained via the DREAM data as derived in Eqs. [1-3]. The
B1- profiles
were then smoothened by means of a regularized polynomial fitting procedure to
remove density weighting.
$$I_{\rm FID}=M_0 B_1^- \sin\beta \cos^2\alpha \left[1\right]$$
$$I_{\rm STE}=M_0 B_1^- \sin\beta \sin^2\alpha \left[2\right]$$
$$M_0 B_1^- = \frac{I_{\rm FID}+2I_{\rm STE}}{\sin\beta} \left[3\right]$$
Two-point mDIXON
images were acquired using the body coil for signal reception via a dual-echo 3D
gradient echo sequence; voxel size = 1.5×1.5×10 mm3, TR/TE1/TE2
= 3.6/1.2/2.0 ms, tip angle = 5°. T2-weighted Fast Spin Echo (FSE)
images were acquired using a torso receive array; voxel size = 1.3×1.3 mm2;
slice thickness = 5 mm; slice gap = 5 mm; TR/TE = 1000/80 ms;
excitation/refocusing tip angle = 90°/120°; 12 refocusing pulses;
sensitivity encoding (SENSE) acceleration = 2.
Results
Fig. 1
shows reconstructed B1+
and B1- profiles
from the DREAM sequence based on the single-slice calibration data combined to mimic
quadrature driving conditions. The results show that there is a mirror-symmetry
relation between the transmit and receive profiles, which is in agreement with basic
electromagnetic theory.3
In Fig. 2 RF-shimmed
multislice DREAM data is used to correct the in-phase images of the
reconstructed DIXON data. Improved image uniformity is especially visible in
the torso and legs. It is also worth noting that the image uniformity is improved
at the cardiac station, despite the flow-sensitivity of the DREAM method.
Fig. 3 shows FSE images in the liver and lower
extremity stations, before and after correction. Although less visible due to
the strong T2 weighting, image uniformity is improved in both examples. Window
levels have been adjusted to the level of the liver and muscle to better
visualize the improved uniformity.
Discussion
This study
shows that DREAM data can be used to perform uniformity correction in body
applications at 3T by extracting the B1- maps. The receive uniformity of the
body coil affects the overall image uniformity in situations when the body coil
is used as a reference during intensity correction. This also holds true in
parallel imaging reconstruction methods using receive arrays in which body coil
data are used to normalize the receive array sensitivity profiles, as confirmed
in these imaging experiments.4
The method proposed here can correct for the receive non-uniformities
by means of a single 15 s acquisition. Channel-wise volumetric B1- calibration
data may also be used for this purpose.5 This allows a whole-body
protocol to be corrected for the B1-
without the acquisition of any additional scans.
Conclusion
The DREAM
B1 mapping method can be used
for uniformity correction in whole-body applications at 3T. This allows for improving image quality at a minimal burden to the protocol.
Acknowledgements
The authors thank Dr. Peter Börnert for constructive discussions.References
1. Nehrke K, Börnert P. DREAM—A Novel
Approach for Robust, Ultrafast, Multislice B1 Mapping. Magn Reson Med
2012;68:1517–1526.
2. Brenner D,
Stirnberg R, Pracht ED, et al. Rapid MRI
System Calibration using 3DREAM. In: Proceedings of the 23rd Annual Meeting of ISMRM,
Toronto, Ontario, Canada 2015; p. 0491.
3. Wang
J, Yang QX, Zhang X, et al. Polarization
of the RF field in a human head at high field: a study with a quadrature
surface coil at 7.0 T. Magn Reson Med 2002;48:362–369.
4. Pruessmann
KP, Weiger M, Scheidegger MB, et al. SENSE:
sensitivity encoding for fast MRI. Magn Reson Med 1999;42:952–962.
5. Sprinkart AM, Nehrke K, Träber F, et al. Ultrafast volumetric B1 (+) mapping
for improved radiofrequency shimming in 3 tesla body MRI. J Magn Reson Imag
2014;40:857–863.