Kay Nehrke1 and Peter Börnert1,2
1Philips Research, Hamburg, Germany, 2Radiology, LUMC, Leiden, Netherlands
Synopsis
To improve the workflow for B1+ calibration on a dual transmit MRI system, the DREAM B1+ mapping sequence has been streamlined for acoustic noise reduction and free-breathing acquisition using a standard external respiratory motion sensor. About 10 dB reduction in sound pressure level were achieved by optimizing the echo order with respect to gradient strength reduction. Feasibility was shown in volunteer experiments on abdominal B1+ mapping.Introduction
In high field MRI, fast and robust in vivo B
1+ mapping is an essential prerequisite for many quantitative or parallel-transmit-based applications (1-3). However, most B
1+ mapping techniques are too slow for seamless integration into the clinical workflow. Recently, the DREAM (Dual Refocusing Echo Acquisition Mode) B
1+ mapping approach has been introduced (4), allowing RF shimming in a couple of seconds at low SAR without compromising the RF shimming performance (5). The present study aims at further refining the DREAM-based B
1+ calibration towards a silent, hidden preparation phase. For this purpose, the sequence is streamlined with respect to acoustic noise reduction. Moreover, a respiratory motion sensor is used to synchronize the short single-shot acquisitions of the individual tx-channel maps with successive expiratory phases. The feasibility of the approach is shown for in vivo B
1+ calibration of a clinical dual-channel MRI system at 3T.
Theory
The DREAM sequence employs a STEAM preparation module for B1+ encoding and a low-angle pulse train for the quasi-simultaneous acquisition of free-induction decay (FID) and stimulated echo (STE) signals as gradient-recalled echoes under a single readout gradient lobe, separated in time by appropriate gradient encoding (Fig.1). Thereby, the polarity of the STEAM dephaser gradient Gm2 determines the temporal order of the FID and STE signals. Given a fixed echo time of the FID (TEFID=2.3 ms for water-fat in-phase at 3T), the STE-first order generally results in a shorter TR of the sequence. However, strong dephaser and spoiler gradient areas Gm1 and Gm3, respectively, are required for proper encoding and spoiling (Fig.1a). In combination with the tight timing, this results in strong peak gradient strengths. In contrast, the more relaxed timing and the smaller dephaser and spoiler gradient areas of the FID-first order (Fig.1b) may be utilized to decrease the gradient demands of the sequence, thus reducing the acoustic noise.
Methods
Volunteer experiments were performed on a clinical 3T MRI system (Ingenia, Philips Healthcare, Best, The Netherlands). Abdominal DREAM B1+ calibration scans were acquired in 2D single shot acquisition for both echo orders (transversal orientation, slice width = 20 mm, FOV = 530x460 mm2, scan matrix = 64x56, STEAM/imaging flip angle = 50°/10°, shot duration ≈ 250 ms). The respiratory bellows of the MRI scanner was placed on the abdomen of the volunteer and used to synchronize the short single shot DREAM acquisition with end-expiration. Thus, the total scan interval for the two tx channels was two respiratory cycles during free-breathing (≈ 10 s). The acoustic noise in terms of the sound pressure was measured using a linear PCM recorder (TASCAM DR-05) . The relative sound pressure level Lp (in decibel) for the two echo orders was determined according to:
$$L_{p}=20\cdot\log_{10}{\frac{p_{FID\_first}}{p_{STE\_first}}}$$
where pFID_first and pSTE_first denote the root-mean-square sound pressures for the the respective echo orders. In addition, the measured in vivo B1+ maps were visually inspected and compared.
Results
The FID-first order reduced the required peak gradient strength from 20 mT/m to 4 mT/m, which resulted in a significant reduction of the sound pressure level by approx. 10 dB (Fig.2). The increased TR (from 3.8 ms to 4.6 ms) led to a moderate increase of the shot duration from 213 ms to 252 ms. Moreover, the SAR level was slightly reduced due to the stretched RF pulses (STE-first: local torso SAR < 16%, whole body SAR < 0.5 W/kg, FID-first: local torso SAR < 11%, whole body SAR < 0.4 W/kg). For both timing schemes, the in vivo B1+ maps show the typical shading pattern characteristic for the employed dual channel system (Fig.3), and no significant differences could be observed.
Discussion
The
proposed sequence enables fast, free-breathing, low-SAR and almost silent B1+ calibration on a dual transmit
MRI system. Hence, it potentially provides all required ingredients for turning
multi-channel B1+ calibration into a robust preparation
phase, without adding significant overhead to the examination. Moreover, the
preparation phase can be hidden from both, patient and MR technician. Thus,
patient discomfort and image quality problems related to breath-holding,
acoustic noise, and warming can be reduced, which represents a strong
improvement in workflow over the existing B1+ mapping technique.
The individual DREAM B1+ maps are acquired in only a fraction
of a second, and hence, freeze respiratory quite efficiently. Thus, the motion
sensor is merely used to acquire all B1+ maps in roughly the
same respiratory state, and therefore, other motion sensor concepts could be
used likewise.
Acknowledgements
No acknowledgement found.References
1. Hoult DI and Phil D. JMRI 2000;12:46-67.
2. Katscher U et al. MRM 2003;49:144-50.
3. Zhu Y. MRM 2004;51:775-84.
4. Nehrke K. and Börnert P., MRM 2012;68:1517-26.
5. Nehrke K, Sprinkart AM, Börnert P.MAGMA. 2015:185-94.