Iulius Dragonu1, Craig Buckley1, Peter Weale1, Matthew D Robson2, and Aaron T Hess2
1Siemens Healthcare Ltd, Frimley, Camberley, United Kingdom, 2University of Oxford Centre for Clinical Magnetic Resonance Research (OCMR), John Radcliffe Hospital, Headington, United Kingdom
Synopsis
Radiofrequency shimming with multiple channel
excitation is a well established method to increase the transverse magnetic
field homogeneity and reduce SAR at high magnetic field strength(≥7T). To harness the benefits of a
parallel transmit system, the magnitude and relative phase of each transmit
channel must be determined during a calibration scan. We propose a new strategy to accelerate the
acquisition of such calibration images by simultaneously exciting several transmit
channels and reconstructing the calibration images using the technique similar
to simultaneous multi-slice acquisitions.
Radiofrequency
shimming with multiple channel excitation is a well established method to
increase the transverse magnetic field homogeneity at high
magnetic field strength(≥7T)(1). To harness the benefits of a parallel transmit system, the
magnitude and relative phase of each transmit channel must be determined during
a calibration scan, this is achieved by acquiring images with each transmit channel
individually.
In cardiac
and abdominal applications, sufficiently high resolution is required in order
to avoid partial volume effects, particularly when the proton density images
are used as a surrogate for absolute B1+(2). For this purpose, we propose a new strategy to accelerate the acquisition
of such calibration images. It is hypothesised that simultaneously exciting
several transmit channels and reconstructing the calibration images using the
technique similar to simultaneous multi-slice acquisitions(3) with controlled
aliasing in order to increase the efficiency of the reconstruction and reduce
the geometry factors of the SENSE reconstruction(4).
Measurements
were performed on a whole-body Magnetom 7T MR-system (Siemens Healthcare,
Erlangen, Germany). Data acquisition was performed with a custom built eight-channel transmit/receive TEM coil (5). Images were acquired using a dual-echo
GRE sequence (TR=7.7ms,TE1,2=1.02,2.04ms, α=14°).
Acquisition
with multiple excitation channels and the controlled aliasing was simulated
offline using the raw kspace data with
acceleration factors of two and four (simultaneous excitation with two and four
channels respectively). Controlled aliasing with a shift of
FOV/2 was used with an acceleration factor of two by multiplying each even line
with a factor of pi and shifts of FOV/4,FOV/2 and 3*FOV/4 with an acceleration
factor of four.
Images corresponding to each transmit channel were calculated using the
SENSE reconstruction. Coil sensitivities(B1-)
were estimated using 16 phase encoding lines acquired in the centre of kspace with single channel excitation.
The images
acquired with two different echo times were combined using the following
formulation
$$Icombined =sqrt(MTE1·MTE2)·exp(j·(𝛟TE2- 𝛟TE1)) (Eq.1)$$
The
magnitude of coil sensitivities were calculated by dividing each of the
magnitude images by the sum of squares image. Furthermore, as B1-
coil sensitivities are independent of the transmit channel, they can be calculated
using a weighted mean of B1-
for each transmit channel.
In absence
of noise, the phase difference 𝛟TE2- 𝛟TE1(Eq.1) is the same for all receive/excitation channels and is proportional to the main magnetic field B0 inhomogeneities. The phase
difference mentioned previously was calculated using the complex sum over all transmit
and receive channels. Regions with higher signal magnitude will contribute more
substantially to the final phase difference than the regions with low signal
magnitude. A complex image corresponding to TE=0
was calculated using the following relationship.
$$I(TE=0)m+,n-=[M(TE1,2)m+,n-·exp(j𝛟(TE1,2)m+,n-)]·exp[-j·(𝛟TE2- 𝛟TE1)·TE1,2/(TE2-TE1)] (Eq.2)$$
The phase image obtained using Eq.2 only has contributions
due to the transmit channel m (Cm+) and reception
channel n (Cn-). For the receive sensitivity maps the phase information of the
transmit sensitivities is removed using one receive channel as a reference by
subtracting its phase from the other receive channels (as such the phase of
each receive channel is Cn-
- Cref-)
For accurate RF shimming, accurate magnitude images and
relative phase between channels is required, consequently, the relative phase
of the transmit channels will not be affected if the phase information of the
receive reference channel is present in all reconstructed images using SENSE.
Fig.1 summarizes the reconstructed transmit images
obtained for a full acquisition, acceleration factors of two, and of four respectively. It can be
observed that the phase difference between the full acquisition and the
accelerated acquisition is very accurate especially in regions with high B1+ values. In
regions with lower B1+
phase difference between the full acquisition and the accelerated is mainly
dominated by noise and no systematic error was observed.
Four representative images of transmit channels acquired
with two fold acceleration are shown in fig.2.
In order to estimate the accuracy of the proposed method the simulated accelerated acquisition was compared to the fully encoded acquisition. The ratio between the magnitude images for the accelerated case and
the full sampling was calculated for all acquired slices. The histogram of the ratio of magnitude images suggests that the
method maintained the accuracy of the maps for acceleration factors up to four while reducing the scan time.
A new
method to accelerate the acquisition of relative B1+ sensitivities using accelerated simultaneous
excitation with multiple transmit channels has been proposed. Accurate relative
transmit sensitivities were obtained with two and four fold acceleration factors
using controlled aliasing. In cardiac imaging
this strategy will enable sensitivities from two or four transmit channels to
be to be measured in a single heart beat.
Acknowledgements
No acknowledgement found.References
(1) Vaughan JT,
DelaBarre L, Snyder CJ, Adriany G, Collins CM, Van de Moortele P-F, Moeller S,
Ritter J, Strupp J, Andersen P, Tian J, Smith MB, Ugurbil K. RF image
optimization at 7T and 9.4T. In: Proceedings of the 13th Annual Meeting of the
ISMRM, Miami Beach, FL, USA, 2005. p 953.
(2) Van de Moortele P-F, Uburbil K. Very Fast
Multi Channel B1 Calibration at High Field in the Small Flip Angle Regime. In
Proceedings of the 17th Annual Meeting of ISMRM, Honolulu, Hawaii,
USA, 2009. p. 367.
(3) Breuer FA, Blaimer M, Heidemann RM, Mueller MF,
Griswold MA, Jakob PM. Controlled aliasing in parallel
imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging.
Magn Reson Med 2005;53:684-691.
(4)
Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: Sensitivity
encoding for fast MRI. Magn Reson Med 1999;42:952-962.
(5) Snyder CJ, DelaBarre L,
Metzger GJ, van de Moortele P-F, Akgun C, Ugurbil K, Vaughan JT. Initial
results of cardiac imaging at 7 Tesla. Magn Reson Med 2009;61:517–524.