Oliver Bieri1,2, Grzegorz Bauman1,2, and Carl Ganter3
1Radiology, University Hospital Basel, Basel, Switzerland, 2Biomedical Engineering, University of Basel, Basel, Switzerland, 3Diagnostic Radiology, Technical University Munich, Munich, Germany
Synopsis
A new method
for accurate and fast broadband frequency mapping with balanced steady state
free precession is introduced. The method mitigates the need for advanced phase
unwrapping algorithms from a matrix pencil analysis of sequentially shifted
echo times. Typically, the new method offers a spectral resolution in the
range of Hertz with a sensitivity range in the order of several thousands of
Hertz.Purpose
Fast and
accurate frequency mapping in the presence of strong magnetic field
inhomogeneities, e.g. as experienced at ultra-high fields, near prominent
susceptibility variations, such as tissue-air interfaces, or in the proximity
of implants, typically generates severe phase ambiguities using contemporary multi-echo
B0-mapping schemes. Here, we make use of sequentially shifted echo times in
combination with a balanced SSFP readout scheme to provide an unconstrained echo-spacing
for accurate and fast broadband B0 mapping with a matrix pencil approach that
does not require dedicated phase unwrapping methods.
Methods
Similar to the
work of Scheffler [1], we make use of balanced SSFP for fast frequency mapping,
but use sequentially shifted echo times instead of a multi-echo readout scheme
(see Fig. 1). The echo-spacing, $$$\Delta$$$TE, enters as an unconstrained parameter, and can be adapted to fit the
needs; e.g. a $$$\Delta$$$TE of 1 ms
mitigates phase ambiguities up to 1kHz, whereas a $$$\Delta$$$TE of 0.1ms offers a sensitivity range up to
10kHz. For B0 mapping, n acquisitions were performed with echo times,
TE(i)=TE(1)+ix$$$\Delta$$$TE, where i =
0,1,…,n-1, and the local frequency is then retrieved from the data by a rank-1 matrix
pencil approach [2].
Broadband
B0 mapping was evaluated in a long cylindrically shaped aqueous phantom (T2/T1
~ 1) and on a hip prosthesis immersed in a tissue-mimicking doped gel (T2/T1 ~
0.1). To this end, the following imaging
parameters were used: flip angle: 10° (hard pulse of 100us duration), $$$\Delta$$$TE = 100us; n = 10 scans (sampling
duration = 900us). The phantom was scanned in 3D with 2mm isotropic resolution,
whereas the hip prosthesis was scanned with an isotropic resolution of 4mm,
2mm, and 1.3mm. Other imaging parameters were: for the 4mm iso scan (approx. 30
sec scan time): 64x64x24, BW = 2298 Hz/Pixel, TR = 2.28 ms, TE = 0.69, 0.79, …
, 1.59 ms; for the 2mm iso scan (approx. 3 min scan time ): 128x128x40, BW =
2298 Hz/Pixel, TR = 2.88 ms, TE = 0.99, 1.09, … , 1.89 ms; for the 1.3mm iso
scan (approx. 10 min scan time): 192x192x80, BW = 1736 Hz/Pixel, TR = 3.39 ms,
TE = 1.24, 1.34, … , 2.14 ms. No filtering of the data, and no acceleration,
such as parallel imaging or partial Fourier, was used. All scans were performed
at 3T using the body coil for signal transmission and reception.
Results & Discussion
First, the
spectral resolution of the BFM-SSFP approach was investigated on a proper
phantom (see Fig. 1a). Generally, the B0-map appears smooth; no discretization
is visible. As a result, the spectral resolution appears to be limited only by
signal-to-noise (SNR) and appears to be in the range of Hertz. It is
interesting here to note that the MP has a much higher spectral resolution as
compared to a simple Fourier analysis, for which the spectral resolution is
given by 1/(nx$$$\Delta$$$TE) > 1kHz.
Generally,
banding becomes a major issue with balanced SSFP; especially in the combination
with the presence of severe perturbations in the homogeneity of the magnetic
field. To this end, the phantom was detuned with a linear gradient, and
prominent bands appear (see Fig. 1b). Nevertheless, expect for the loss in SNR,
banding represents no issue for B0-mapping and the frequency can be accurately resolved
throughout.
Finally,
broadband frequency mapping was tested in the proximity of a hip prosthesis
(see Fig. 3). Widespread banding artifacts even in combination with a TR as
short as 2 ms attest the presence of severe local field inhomogeneities.
Indeed, deviation from the on-resonance frequency roughly up to +4kHz and down
-3kHz are observed. Surprisingly, no drop-offs are visible, even in the very
close proximity of the implant.
Conclusion
We have
introduced a new method for fast broadband frequency mapping with balanced SSFP
that does not require phase unwrapping methods and offers a compelling spectral
resolution and sensitivity range. The method may be of special interest for
engineering and material studies, at ultra-high fields, or for temperature
monitoring near metallic implants.
Acknowledgements
Reinhard Elke for
providing us the hip implant
materialReferences
[1]
Scheffler K. Fast frequency mapping with balanced SSFP: theory and application
to proton-resonance frequency shift thermometry. Magn Reson Med. 2004
Jun;51(6):1205-11.
[2] Lin Y-Y, Hodgkinson P, Ernst M, Pines A
(1997) A Novel Detection–Estimation Scheme for Noisy NMR Signals: Applications
to Delayed Acquisition Data. Journal of Magnetic Resonance 128: 30-41.