Fast B0 first order inhomogeneity estimation using radial acquisition
Ali Aghaeifar1,2, Alexander Loktyushin1, Christian Mirkes1,3, Axel Thielscher1, and Klaus Scheffler1,3

1Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 2IMPRS for Cognitive and Systems Neuroscience, Tübingen, Germany, 3Department of Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany

Synopsis

B0 field inhomogeneity is a major source of distortion in MR images. Current approaches to dynamic shimming require extra acquisition time or external hardware. We propose a method that estimates first order shim errors by using projections of radial acquisition. The errors can be estimated from three projections multiple times in each measurement, which makes the method highly robust. The proposed method is evaluated in simulation and in vivo. Obtained results show a strong agreement between applied and measured first order shim errors.

Introduction

Modern MR scanners employ shim coils in order to make the B0 field homogeneous. However, susceptibility differences at the interfaces, intentional motion of the subject or involuntary physiological motion such as breathing can alter the shimmed fields and lead to image distortions and signal loss. To solve such problems, it is common to use dynamic shimming (real-time prospective shim correction), which updates the shim settings concurrently with the acquisition. This requires a periodic evaluation of the field fluctuations, which can be done by using the sensitivity information of coil elements 1, shim navigators 2, or external field sensors attached to the subject or the coil 3. The existing approaches, however, have limitations such as increased scan time duration, slow performance or the need for dedicated external field monitoring hardware. In this work, we employ a radial acquisition both for imaging and estimating the first order inhomogeneity of the B0 field, bypassing the need of dedicated navigators or hardware.

Method

Recently it was proposed to use cloverleaf navigators for combined real-time motion correction and shimming 4. Assuming a perfect shim and neglecting the phase of the receiver coil, the strongest signal intensity is found in the center of k-space which represents DC component of k-space. First order shim errors lead to a shift of the DC component away from the center. The three traversals of the cloverleaf navigator through the center of k-space provide the projections of this shift onto the X, Y, and Z axis. Based on this idea, we hypothesize that in radial acquisitions the spokes can be used not only for imaging but also for estimation of the first-order shim errors. Since each spoke is acquired in a different direction, the information provided by several projections can be used to determine the first order shim error in X, Y and Z axis. In Figure 1, we show how position of DC component changes in the XY plane given a good shim (Fig. 1A) and a small linear shim offset in X direction (Fig. 1B). Traversing k-space along a given axis in the presence of a first order field inhomogeneity oriented in the same direction can be mathematically described by Eq. [1].

$$k(t)=(\frac{3}{2}t_{ramp}+t_{deph}+t){\triangle G}_{}+(t_{ramp}+t_{deph})G_{deph}+(\frac{t_{ramp}}{2}+t)G_{readout} \quad\quad [1]$$

where t=0 is a starting time of the data acquisition, Gdeph and Greadout are the amplitudes of the dephasing and readout gradients, tramp the gradient ramp-up or ramp-down time (assumed to be identical for all gradients) , tdeph the flat-top time of the dephasing gradient, and ΔG the first order component of a field inhomogeneity modeled as a gradient offset. In 2D k-space, the signal will be maximum at tpeak when Kx2+Ky2 is minimum. Applying the derivative operator to Kx2+Ky2 with respect to time results in a circle (see Eq. [2]).

$$a_{3}(\triangle G_x^2 + \triangle G_y^2)+a_{2}\triangle G_{y}+a_{1}\triangle G_{x}+a_{0} = 0 \quad\quad [2]$$

$$a_{0}=(G_{xreadout}^2 +G_{yreadout}^2)(\frac{t_{ramp}}{2}+t_{peak})+(G_{xdeph}G_{xreadout}+G_{ydeph}G_{yreadout})(t_{ramp}+t_{deph})$$

$$a_{1} = G_{xreadout}(t_{deph}+2t_{peak}+2t_{ramp}) + G_{xdeph}(t_{ramp} +t_{deph} )$$

$$a_{2} = G_{yreadout}(t_{deph}+2t_{peak}+2t_{ramp})+ G_{ydeph}(t_{ramp} +t_{deph} )$$

$$a_{3} = \frac{3}{2}t_{ramp}+t_{deph}+t_{peak}$$

This circle is a set of solutions associated with all possible gradient offsets which result in a maximum intensity peak at tpeak. Finding a unique solution (the gradient offsets) requires several circles, and thus several projections. At least 3 non-identical circles are required to find the gradient offsets, as shown in Figure 2.

Results & Discussion

We have evaluated the proposed method in simulation and in vivo. A spoiled GRE sequence was modified to allow a projection-based acquisition of k-space. While scanning a subject, we altered the first order component of the shim in small steps of -20uT/m to +20uT/m in X and Y directions in each measurement. All experiments were performed on a MAGNETOM Prisma 3T scanner (SIEMENS Healthcare, Erlangen, Germany) using a two-channel head coil. Additionally, we computed gradient delays of the scanner and applied them to the measured samples. Figure 3 shows a comparison of applied shim offsets and the offsets estimated by our method. The precision depends on how well it is possible to pinpoint the position of the peak, which requires to maximally shorten the dwell-time of analogue-to-digital converter (ADC). As shown in Table 1, there is a good agreement between applied and the measured offsets.

Conclusion

We propose an approach for first order shim error estimation that does not require navigators, reference scans or external hardware, and is appropriate for radial imaging with short TRs. Shim gradients can be updated from three consecutive projection measurements allowing to perform the imaging with real-time shim correction. An extension of the method to 3D imaging is possible and is planned as a future work.

Acknowledgements

No acknowledgement found.

References

[1] D. N. Splitthoff and M. Zaitsev, “SENSE shimming (SSH): A fast approach for determining B0 field inhomogeneities using sensitivity coding,” Magn. Reson. Med. 2009;62(5):1319–1325.

[2] H. a. Ward, S. J. Riederer, and C. R. Jack, “Real-time autoshimming for echo planar timecourse imaging,” Magn. Reson. Med. 2002;48(5):771–780.

[3] M. Haeberlin, L. Kasper, K. P. Pruessmann, and C. Barmet, “Combined Real-Time Prospective Motion Correction and Concurrent Field Monitoring,” in ISMRM2011, 2011.

[4] a. J. W. Van Der Kouwe, T. Benner, and A. M. Dale, “Real-time rigid body motion correction and shimming using cloverleaf navigators,” Magn. Reson. Med. 2006;56(5):1019–1032.

Figures

Figure 1. Acquired signal in different radial projections. A: slice is rather well-shimmed. B: there is a -8uT/m offset in X direction. Arranged along the Y axis are 360 radial projections from 0 to 2pi and the X axis represents the samples.

Figure 2. The set of solutions for the first order shim offsets with three distinct projections, and the corresponding intersection point. Angles of the projections are 0, 2/3pi and 4/3pi and applied shim offset is +2uT/m in both X and Y direction.

Figure 3. Comparison between the applied Y (left) and X (right) offsets and the estimated first order shim measured in a human subject. Sequence parameters that we used were as following: 500mm FOV, TE/TR = 4.4/8.8ms, 350 Hz/Pixel Bandwidth and 20o flip angle.

Table 1. Covariance between the applied and calculated shim offsets measured in a volunteer. Ya and Xa are applied shim offsets, and Ym and Xm are the estimated shim offsets. Values are normalized to the variance of the applied offset.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0932