Shi Feng1, Shiloh Sison2, Jazmine Garcia3, Gabriel Mouchawar3, and Richard Williamson3
1Hardware development, St. Jude Medical, Sylmar, CA, United States, 2St. Jude Medical, Sunnyvale, CA, United States, 3St. Jude Medical, Sylmar, CA, United States
Synopsis
Metallic
leads of an implanted electronic device such as a pacemaker may behave as
antennae in the strong radio frequency electromagnetic field of MRI. The
induced current surrounding the electrodes may heat the local tissue. The
MRI-induced tissue heating around the electrodes of a pacemaker have only been
investigated for pacemakers employing a single lead. In this paper, we examine
the MRI-induced temperature rise (TR) of the tip electrode(s) associated with a
pacemaker system with two St. Jude Medical Tendril 2088 STS leads, and compare
it to the single result. Both transfer function and in vitro temperature rise
are investigated.Introduction
When
subjected to MRI and its radio frequency (RF) field, metallic leads of an
implanted electronic device such as a pacemaker may behave as antennae. The
induced current surrounding the electrodes may heat the local tissue [1]. The MRI-induced tissue heating around the
electrodes of a pacemaker have only been investigated for pacemakers employing
a single lead [2], [3]. In this paper, we examine the MRI-induced
temperature rise (TR) of the tip electrode(s) associated with a pacemaker
system with two St. Jude Medical Tendril 2088 STS leads.
Method
Due to the complex and tiny pacemaker leads
structure, it has an established transfer function (TF) modeling the
conversion of the incident electric field to an induced current which may be
employed in analyzing the electrode/tissue heating effect [4]. The tip TR can be estimated
by squaring the inner product of the incident electrical field along the lead
pathway and the TF. Consequently, the lead heating impact of adding a second
lead to a pacemaker system can be estimated by comparing the TFs of pacemaker
systems with 1 V.S. 2 leads. We measured the TFs using established means [2], [3] of 1 and 2-lead
pacemaker systems in a saline bath phantom with electrical conductivity of 1.2
S/m and a relative permittivity of 80 at 64 MHz, approximating the electrical
characteristics of blood. TR measurements were performed with leads located
along several phase reversal pathways in the ELIT1.5 phantom filled with HEC
gel with the same electrical parameters as the saline used in the TF
measurement. The two leads are separated at 3cm from distal end to avoid the interference
to the local SAR (Specific Absorption Rate) distribution from each other. The
phantom was loaded in the MITS 1.5 T vertical RF coil as show in Figure 1.
Result
The TF of a pacemaker system with two Tendril
2088 58cm STS leads was measured, scaled and compared with that of a system
with a single lead. The magnitude of the two lead TF was clearly lower than that
of the single lead, and the shape was also changed, as show in Figure 2 (a); meanwhile, the
TF phase of the dual lead system was almost identical to that of the single
lead system, as shown in Figure 2 (b). Based on the TF
value, a much lower TR would be expected for the dual lead system, which was verified
by the in vitro measurement result as shown in Figure 3. The triangle marks
and the circular marks represent the measured TR from a single lead and dual
leads, respectively. On average, a 47% drop in TR was achieved with the dual
lead system. However, this heating from the
dual lead pacemaker system cannot be directly predicted by scaling the single
lead result because the transfer function shape was also changed. A careful TF
validation process was accomplished by comparing the predicted and measured TR
on each lead pathway as shown in Figure 4.
Conclusions
We compared the MRI-induced lead heating
resulting from pacemaker systems employing one and two leads. The dual lead
system experienced significantly reduced tip heating compared to the single
lead system. We have shown that the RF heating TF methodology specified in
ISO/TS 10974 can be extended to dual lead TF and that the accuracy of dual lead
TFs are comparable to single lead TFs.
Acknowledgements
No acknowledgement found.References
[1] S. Pisa, G. Calcagnini, M. Cavagnaro, E.
Piuzzi, E. Mattei, and P. Bernardi, “A study of the interaction between
implanted pacemakers and the radio frequency field produced by magnetic
resonance imaging apparatus,” IEEE Trans. Electromag. Compat, vol. 50, no. 1,
pp. 35-42, February 2008.
[2] S. Feng, R. Qiang, W.
Kainz, and J. Chen, “A Technique to Evaluate MRI-induced Electric Fields at the
Ends of Practical Implanted Lead,” Microwave Theory and Techniques, IEEE
Transactions on, vol. 63, no.1, pp.305,313, Jan. 2015
[3] J. A. Nyenhuis, S.-M. Park, R.
Kamondetdacha, A. Amjad, F. G. Shellock, and A. R. Rezai, “MRI and Implanted
Medical Devices: Basic Interactions With an Emphasis on Heating,” IEEE
Transactions on Device and Materials Reliability, vol. 5, no. 3, pp. 467-480,
September 2005
[4]
S.-M. Park, R. Kamondetdacha, and J. A.
Nyenhuis, “Calculation of MRI-Induced Heating of an Implanted Medical Lead Wire
With an Electric Field Transfer Function,” Journal of Magnetic Resonance
Imaging, vol. 26, no. 5, pp. 1278-1285, November 2007