RF Induced Heating of Overlapped Stents
Peter Serano1,2, Maria Ida Iacono1, Leonardo M. Angelone1, and Sunder S. Rajan1

1U.S. Food and Drug Administration, Washington, DC, United States, 2Electrical and Computer Engineering, University of Maryland, College Park, MD, United States

Synopsis

In this study, the authors present an analysis of a potentially overlooked clinical scenario, namely overlapped stents separated with a layer of insulation. Electromagnetic and thermal simulations as well as measurements were performed to test such configurations. The results show that implanted medical devices that include gapped conductive structures, like overlapped stents, can affect the location and magnitude of peak heating near the implant.

Introduction

The RF induced heating of elongated medical implants like stents is typically exhibited by an increase in temperature in the tissue region around the extremities of the device. As such, ASTM 2182-11a [1] recommends measuring temperature increases at these locations. In this study, the authors present an analysis of a potentially overlooked clinical scenario, namely overlapped stents separated with a layer of insulation, which may both shift the location and magnitude of peak heating near the implant. Electromagnetic and thermal simulations as well as measurements were performed to test such configurations.

Methods

Four stent configurations were evaluated, each with an overall length of 100mm and radius of 3.175mm. The first was a single 100mm stent. The next three were configurations of two overlapping stents: 55mm stents with a 10mm overlap, 70mm stents with a 40mm overlap, and 85mm stents with a 70mm overlap. Electromagnetic and thermal simulations were performed using the ANSYS Workbench Platform with ANSYS HFSS and Mechanical modules (ANSYS Inc, Canonsburg, PA). Test implants were made from a sheet of copper mesh and PTFE heat-shrink tubing to create the overlap. Implants were placed in a gel phantom in accordance with ASTM 2182-11a and provided a circularly polarized 128MHz RF excitation via the MITS 3.0 system (Zurich Med Tech, Zurich, Switzerland). Temperature rise after 15 minutes of exposure to a WBSAR=0.75W/kg was recorded using fluoroptic temperature probes (Neoptix Inc, Quebec, Canada) at both the ends of the device as well as two additional locations at the edges of the overlap.

Results and Discussion

The results show that the overlapped stent configurations can affect both the location and magnitude of peak heating near the implant. In the case of the two 55mm stents with a 10mm overlap, an increase in the peak temperature compared with the single 100mm stent was observed in both simulation and measurements along with a shift in the location of peak heating from the end to the edge of the inner overlap. This effect is a result of an increased magnitude of the E-field between the two gapped conductive structures.

Conclusion

In this study, the authors have demonstrated that implanted medical devices that include gapped conductive structures, like overlapped stents, can affect the location and magnitude of peak heating near the implant.

Acknowledgements

This project was supported in part by an appointment to the Research Participation Program at the Center for Devices and Radiological Health / Office of Science and Engineering Laboratories, U.S. Food and Drug Administration, administered by the Oak Ridge Institute for Science and Education through an interagency agreement between the U.S. Department of Energy and FDA.

References

[1] American Society for Testing and Materials (ASTM) F2182 - 11a, “Standard test method for measurement of radio frequency induced heating on or near passive implants during magnetic resonance imaging.”

Figures

Figure 1: Thermal simulation results after 15 minutes of RF exposure to 0.75W/kg WBSAR

Figure 2: Measurement setup including MITS 3.0 test system and ASTM phantom

Figure 3: Temperature probe placement & numbering

Figure 4: Measured temperature rise during 15 minutes of RF exposure to 0.75W/kg WBSAR - single 100mm stent (top-left), two 55mm stents with a 10mm overlap (top-right), two 70mm stents with a 40mm overlap (bottom-left), and two 85mm stents with a 70mm overlap (bottom-right)



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0915