Volkan Acikel1, Patrick Magrath1,2, Scott E Parker1, Holden H Wu1, Peng Hu1, Paul J Finn1, and Daniel B Ennis1,2
1Department of Radiological Sciences, University of California Los Angeles, Los Angeles, CA, United States, 2Department of Bioengineering, University of California Los Angeles, Los Angeles, CA, United States
Synopsis
MRI exams for patients with pacemakers and implanted
cardioverter defibrillators (ICDs) are contraindicated at all clinical field
strengths. The aim of this study was to
measure directly RF induced lead tip heating during MRI exams of cadavers with
existing devices at both 1.5T and 3T.Introduction
MRI exams for patients with pacemakers ICDs are
contraindicated at all clinical field strengths. One principal concern is radiofrequency (RF)
induced lead-tip heating, which may transiently or irreversibly damage tissue
at the lead tip and inhibit pacing. The aim of this study was to measure
directly RF induced lead tip heating during MRI exams of cadavers with existing
devices at both 1.5T and 3T. The data show that pacemaker lead tip temperature
increases were below 4C for all cadavers at both 1.5T and 3T over a range of
iso-center positions despite the lack of cooling from tissue perfusion.
Methods
Experiments were conducted with six (N=6) cadavers (Table 1). Three to four fiberoptic temperature probes were
implanted adjacent to the right atrial (RA), right ventricular (RV) and/or
abandoned lead-tips under x-ray guidance (Figure 1). Whole body CT (1mm isotropic resolution) was used
to measure the tip-to-tip distances of leads and temperature probes. Each cadaver was exposed to 4W/kg whole body
SAR for 15-minutes at each of five different iso-center positions both at 1.5T
and 3T (Siemens Avanto and Prisma). The
first iso-center position was 6cm superior to the chin (LM
1) and the
other four iso-center positions were in 15cm increments inferior to the first
position (LM
2 to LM
5).
The maximum temperature increase (∆T
Max) was calculated as
the difference between the maximum temperature after 15 minutes of RF exposure
and a 20 second average before RF exposure. For single chamber pacemakers a temperature probe
was placed at the lead tip; for dual chamber pacemakers the temperature probes
were placed at the RA and RV lead tips; for the cadaver with abandoned leads
two temperature probes were placed close to the attached lead tip. In all cases the remaining probes were placed
in remote tissues for reference.
Results
All temperature probes were ≤10mm (5.74±2.59mm) from
the lead-tip (Table 2). However, in cadaver 2 the temperature sensor was too
distant from the RV lead tip to obtain temperature data. In the absence of perfusion, it is assumed
that the duration of RF exposure (15-minutes) was sufficient to ensure nearly maximum
heating at the probe location. Figure 1 shows the location of the
probes. Figure 2 shows ∆T
Max
for each probe in each cadaver at both 1.5T and 3T. Maximum lead tip heating
was observed at LM
2 and LM
3 for which the pacemaker/ICD+lead
was closest to the body transmit coil isocenter. ∆T
Max was >2C for 8 of 25 cases
at 1.5T and for 12 of 25 cases at 3T, but never exceeded 4C. Note that for 6 of these 8 cases at 1.5T and
10 of these 12 cases at 3T the background body temperature also increased >2C.
Discussion
The lead tip heating results were very similar
at 1.5T and 3T. The number of cases covered here is limited, but with an
increased number of cases these data will be helpful to evaluate the potential
severity of RF induced lead tip heating during routine clinical MRI scans. A
distinct advantage of these experiments is that they mimic as closely as
possible clinical MRI exams for living subjects except for the lack of
perfusion. The lack of perfusion,
however, creates a “worst-case-scenario” and despite the lack of perfusion ∆T
Max
did not exceed 4C for any measure.
Note that each implanted device was evaluated in the original position found
at the time of death. Consequently, the results do not represent the maximum
possible heating for that subject and device, which may only be found through
considerable manipulation of the subject or through simulation. High resolution
MRI was acquired for each cadaver, which is amenable to simulation and is the
subject of future work. Future work should also perform an analysis that
calculates the coupled RF power to the implants at both 1.5T and 3T.
Conclusion
Pacemaker lead tip temperature increases were below 4C for
all cadavers at both 1.5T and 3T over a range of iso-center positions despite
the lack of cooling from tissue perfusion. The results are encouraging, but further
work is needed before clinical exams for patients with implanted ICD/pacemakers
can be judiciously acquire at 3T.
Acknowledgements
This work was support by NIH/NHLBI R21-HL127433 to DBE and
the Department of Radiological Sciences at UCLA.References
No reference found.