MR-PET simultaneous acquisitions with attenuation correction using LSO background radiation.
Liliana Lourenco Caldeira1, Theodoros Kaltsas1, Jürgen Scheins1, Elena Rota Kops1, Lutz Tellmann1, Uwe Pietrzyk1, Christoph Lerche1, and N. Jon Shah1,2

1Institute of Neuroscience and Medicine, Forschungszentrum Jülich, Jülich, Germany, 2Department of Neurology, Faculty of Medicine, JARA, RWTH Aachen University, Aachen, Germany

Synopsis

In this work, the goal is to perform attenuation correction (AC) for MR-PET scanners using the background activity from LSO (Cerium-doped Lutetium Oxyorthosilicate) scintillator used in PET scanners. This approach has the advantage of obtaining a geometrically aligned AC map with the PET emission scans, which can be useful for coil AC maps. We demonstrate our approach for the Siemens 3T MR-BrainPET with a Tx/Rx 8-channel head coil and a 3-rod phantom.

Purpose

The goal is to perform attenuation correction (AC) for MR-PET scanners using the background activity from LSO (Cerium-doped Lutetium Oxyorthosilicate) scintillator used in PET scanners1, which are the most commonly used scintillator crystals in PET. To obtain quantifiable PET images AC is the most important correction. Furthermore, AC is not only required for the object being imaged (phantom or subject), but also for the MR coils normally placed inside the PET field-of-view (FoV).

Background

In MR-PET imaging, although the object being imaged can usually be visualized with MR images, the MR coils are not generally visible. Despite the low attenuation materials used for the fabrication of MR coils, it is still necessary to perform AC, especially if thorough PET image quantification is required. The AC factors depend highly on the material used and this is not precisely known for MR coils. Instead of using computed tomography (CT) or transmission scans, which require an additional separate acquisition in another scanner, intrinsic LSO radiation can be used to obtain an AC map in the same scanner if sufficiently long transmission scan times can be realised. This has the advantage of a geometric alignment of the generated AC map with the PET emission scan.

Methods

In this work we performed the experiments on a Siemens 3T MR-BrainPET2, which allows simultaneous acquisition of MR and PET. The BrainPET consists of 32 detector cassettes covering an axial FoV of 19.2 cm and a diameter of 31.4 cm. Each detector module is a 12×12 matrix of 2.5×2.5×20 mm3 LSO crystals coupled to a 3×3 array of avalanche photodiodes (APDs). Two types of measurements were carried out: with object (Tx/Rx 8-channel head coil or 3-rod phantom) and without object (blank scan). Figure 1 depicts the 3-rod phantom. Since natural activity of LSO due to presence of 176Lu is low (225 decays s-1cm-3), scan times must be long to minimize statistical errors. This was accomplished by concatenation of data from LSO scans during idle periods. Temperature and background activity were checked for stability. Temperature was simultaneously recorded for each detector cassette for gain stability control. Furthermore, the energy window for acquisition was widened so that all gamma rays coming from LSO would be detected (307 and 202 KeV). The detected counts were corrected for random events estimated using the delayed window events and applying a variance reduction method for noise reduction. Calculations were then performed to obtain quantities that can be reconstructed: the object scan is divided by the blank scan and the logarithm is applied (ACF=-log(I/I0), where I is the object scan and I0 is the blank scan). Maximum likelihood expectation maximisation (MLEM) with median root prior regularization was used to reconstruct images in both line-of-response (LOR) and sinogram space. As usually done for CT, AC values obtained from LSO background need to be extrapolated from 202 and 307 keV to 511 keV (energy of gamma rays used in PET imaging). Furthermore, to assess reconstruction using different AC maps, a 3-rod phantom activity measurement was performed with and without coil. Simultaneously, two MR images were acquired: an MPRAGE and a SPACE. These MR images can be used to obtain AC maps of the 3-rod phantom through segmentation methods.

Results

Mean temperatures of all cassettes remained stable with maximum difference to reference temperature 1.7°C ± 0.1 (Figure 2 shows example of four cassettes) and only negligible effect on APD gain is expected. Data acquired in transmission measurements (64 h each) allowed for the generation of AC maps from the reconstructed image. In Figure 3, the reconstructed maps of the head coil and 3-rod phantom can be seen. Since the coil goes beyond the PET reconstructed FoV, truncation artifacts were observed (Figure 3 (a)). We also found that sinogram-based calculations outperformed LOR-based calculations, which resulted in better AC maps. This is mainly due to statistics, since in sinogram space several LORs are grouped together, thus increasing the statistics per bin (Figure 4).

Conclusions/Future work

Obtaining AC maps for MR coils from LSO background seems feasible if the PET reconstruction FoV is adapted to the coil size or, alternatively, if AC factors for each sinogram entry are computed directly. More complex reconstruction algorithms can be used to obtain better reconstruction3. Furthermore, the use of MR surface coils, in which the exact location depends on the placement during acquisition, is also a problem that might be tackled using this approach.

Acknowledgements

No acknowledgement found.

References

[1] Rothfuss, Harold, et al. LSO background radiation as a transmission source using time of flight. Physics in medicine and biology 59.18 (2014): 5483.

[2] Herzog, H., et al. High resolution BrainPET combined with simultaneous MRI. Nuklearmedizin-Nuclear Medicine 50.2 (2011): 74.

[3] Erdogan, Hakan, and Jeffrey A. Fessler. Ordered subsets algorithms for transmission tomography. Physics in medicine and biology 44.11 (1999): 2835.

Figures

Figure 1: (a) The 3-rod phantom is a cylindrical phantom containing 3 rods, which are placed inside (b) the Tx/Rx 8-channel coil. The white rod is made of Teflon, while the red and black, as well as the surrounding cylinder, can be filled with different liquids and activity quantities.

Figure 2: Mean temperature of four individual cassettes plus its standard deviation over six months. Constant temperature means that temperature variations can be neglected.

Figure 3: Reconstructed AC maps of (a) the Tx/Rx 8-channel coil, (b) the 3-rod phantom and (c) the 3-rod phantom inside the Tx/Rx 8-channel coil, after 64 h of acquisition. Blank scan was also acquired for 64 h.

Figure 4: Percentage of zeros for three types of measurements: Tx/Rx 8-channel head coil (blue), 3-rod phantom (red) and blank scan (yellow) in sinogram (dashed lines) and LOR (full line) space. Each frame is 4 h.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0882