Michael Marx1, Pejman Ghanouni1, and Kim Butts Pauly1
1Radiology, Stanford University, Stanford, CA, United States
Synopsis
Multi-slice thermometry was
developed that overcomes several limitations of single-slice 2DFT thermometry
in MR-guided focused ultrasound brain treatment. Using multiple-echo spiral
imaging provides much greater imaging performance, which was applied to
improved focal spot localization and to improved ablation monitoring. High-resolution
higher-precision multi-slice focal spot localization can shorten treatment time
and improve patient safety. High-speed high-precision focal spot monitoring,
combined with full-brain monitoring and 3-dimensional focal spot
characterization during ablations can improve treatment guidance and feedback
while also improving patient safety. The new sequences were validated both in vivo and in a phantom within a
clinical transducer.Purpose
MR-guided
focused ultrasound (MRgFUS) brain treatments are currently guided by one
thermometry sequence: single-slice 2DFT MR thermometry. In this work, we divided
treatment monitoring into two tasks, with different thermometry design goals
for each, and developed sequences optimized for these goals: “Focal Spot
Localization” and “Monitoring”. These
sequences achieve greater imaging performance by utilizing multi-echo spiral thermometry1,
region-specific update rates, and MASTER2 slice interleaving.
Currently, focal spot targeting
confirmation requires several low-power sonications to obtain high-resolution
measurements in three dimensions. We developed a Focal Spot Localization
sequence that obtains high-resolution measurements in-plane, at improved
temperature precision compared to 2DFT, while also providing multiple-slices
for through-plane characterization. This would reduce the number of sonications
required, improving treatment time. Additionally, even lower-power sonications
could be detected, improving patient safety.
During ablative treatment
sonications, single-slice monitoring cannot detect through-plane shifts of the
ultrasound focus, or unexpected out-of-plane heating. We developed a multi-rate
thermometry Monitoring sequence that interleaves different sequences at
different update rates to simultaneously achieve fast and precise focal
monitoring, 3-dimensional focal spot measurement, and full brain monitoring.
Fast and precise multi-slice monitoring of the focus ensures accurate thermal
dose estimates for treatment feedback, while (slower) full-brain monitoring
ensures patient safety
Methods
All sequences were
implemented using RTHawk (HeartVista, Menlo Park, CA) on a GE 3T 750 Signa
scanner (GE Healthcare, Milwaukee, WI) equipped with the Insightec Exablate
Neuro (Insightec, Haifa, Israel). All imaging was performed with the body coil,
as is normally used with the Exablate system. Conventional 2DFT was implemented
as a gold standard for comparison. The new sequences all use a 36 cm FOV to
ensure that the transducer and water bath do not alias. Volunteer imaging was
done with informed consent under IRB approval. Multi-frequency reconstruction1 was performed on all data spiral data. Temperature uncertainty was measured as
the per-voxel temporal standard deviation of temperature measurements, and
averaged within manually segmented ROIs. Performance was also tested inside the
transducer, using a gel phantom. Sequence parameters and MR parameters are
compiled in Tables 1-2.
The Focal Spot Localization
sequence is a 5-slice 3-echo thermometry sequence, with doubled in-plane
resolution as compared to conventional 2DFT (1.1x1.1 mm vs 1.09x2.18 mm), and
acquisition time of 7 s. The Monitoring sequence interleaves 3 distinct
sequences at different rates to monitor 29 total slices. The “Focus” is
monitored using 3-slice 3-echo spiral imaging, for high-speed high-precision measurement
of focal heating. Two adjacent slices, the “Boundary”, are monitored at half
the temporal rate (also using 3-echo spiral) to fully characterize the focal
spot. The remaining 24 slices, “Background”, were acquired using 8 blocks of
3-slice MASTER2, with spiral readouts. Use of MASTER improves
temperature uncertainty, compared to traditional slice interleaving, by
increasing echo time. Limiting each MASTER block to 3 slices reduces inherent diffusion
and motion-encoding artifacts. The three component sequences were interleaved
such that Focus utilized 45% (15%
per-slice) of the timeline, Boundary used 15% (7.5% per-slice), and Background
the remaining 40% (1.7% per-slice).
Results
Figure 1 compares temperature
uncertainty
in vivo between 2DFT and
Focal Spot Localization while Figure 2 compares 2DFT with Monitoring. Large images
compare the center slice, while stacks of images at the right show additional
slices monitored by the new sequences. Figure 3 compares uncertainty within the
transducer. The new sequences obtained better uncertainty than 2DFT, with average
uncertainty compiled in Table 1. Relative “Efficiency” is also listed in Table
1, which accounts for differences in speed and voxel volume. Each multi-echo
spiral sequence is more than 150% as efficient as 2DFT. Background is 69% as
efficient (but collects 3 slices per TR, for an effective 120% efficiency.)
Conclusions
In this work, we have shown
that multi-slice multi-echo spiral thermometry is an effective imaging approach
for volumetric treatment monitoring. Improved imaging performance was successfully
used to achieve imaging objectives for different aspects of ablative
treatments. Focal spots may be localized faster and with reduced heating using
the higher-resolution higher-precision Focal Spot Localization sequence. The
mixed update rate Monitoring sequence successfully delivers high-speed
high-precision monitoring of the targeted focus, while improving safety by
simultaneously monitoring the full brain at a lower update rate. These
sequences have also been validated within the transducer, to help ensure they
will work in the clinical setting.
Acknowledgements
This work was supported by
funding from P01 CA159992, from InSightec, and from General Electric.References
1. Marx, Michael and Kim
Butts Pauly. "Improved MRI thermometry with multiple-echo spirals." Magnetic Resonance
in Medicine (2015). DOI: 10.1002/mrm.25914
2. Marx,
Michael, Juan Plata, and Kim Butts Pauly. “Toward volumetric MR Thermometry
With the MASTER Sequence.” IEEE Transactions on Medical Imaging. 34.1 (2015):
148-155.