Christopher M Rank1, Sebastian Sauppe1, Thorsten Heußer1, Andreas Wetscherek1, and Marc Kachelrieß1
1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany
Synopsis
We propose a new method for 5D respiratory
and cardiac motion compensation (MoCo), which employs highly undersampled MR
data and thus requires acquisition times as low as 2 minutes. Radial MR data of
the thorax of three free-breathing patients were acquired. Respiratory and
cardiac motion vector fields were estimated allowing for 5D MoCo
reconstructions, which employ 100% of the measured raw data for reconstruction
of each combination of respiratory and cardiac phase. These 5D MoCo reconstructions
clearly resolve different combinations of respiratory and cardiac phases while achieving high temporal and spatial resolution as well
as low noise and artifact levels.Purpose
Dynamic imaging of organs can
provide valuable information for radiotherapy applications or for studying
physiology. However, time-resolved 5D (3D + respiratory + cardiac) MR imaging
is time-consuming and recent approaches have reported acquisition times in the
range of 15 min
1,2. To allow for acquisition times as low as 2 min,
we propose a new method for 5D respiratory and cardiac motion-compensated image
reconstruction, which is based on radial MR data with very high undersampling.
Methods
Contrast-enhanced MR data
covering the thorax of three free-breathing patients
were acquired at 1.5 T (Magnetom Aera, Siemens Healthcare, Erlangen, Germany).
Data acquisition and evaluation was in accordance with the local ethics
committee and informed consent was obtained from each patient. We applied a
vendor-provided radial stack-of-stars sequence with golden angle radial spacing
and sagittal slice orientation: total acquisition time: 2.0 min, radial spokes
per slice: 720, field-of-view: 385×385×300 mm
3, voxel size:
1.5×1.5×5.0 mm
3, TR/TE = 3.77/1.69 ms, 60 slices (60% slice resolution,
33% slice oversampling, 6/8 partial Fourier), flip angle: 12°, fat supression
activated. Respiratory and cardiac motion signals used for
self-gating were estimated from the k-space center for each acquired spoke. A
bandpass filter was applied to distinguish between respiratory motion (filter
range: 0.1 – 0.5 Hz) and cardiac motion (filter range: 0.5 – 2.5 Hz) and the
signals were corrected for a baseline drift.
Motion
vector fields (MVFs) were estimated independently for respiratory and cardiac
motion using a newly-developed algorithm, which alternates between image
reconstruction and motion estimation. To increase robustness, deformable
registrations were carried out between adjacent motion phases and regularized
by cyclic constraints
3. In a first step, respiratory MVFs were estimated (Fig.
1) using
Nr = 20
overlapping respiratory motion bins with a width of Δ
r = 10% and
neglecting cardiac motion (
Nc
= 1, Δ
c = 100%). In a second
step cardiac MVFs were estimated (Fig. 2) assuming that respiratory motion in
the end-exhale plateau can be neglected. Thus 25% of the raw data most
consistent to end-exhale (
Nr
= 1, Δ
r = 25%,
r =
rref) divided
into
Nc = 10 overlapping
cardiac motion bins with a width of Δ
c =
20% were used for estimation of cardiac MVFs.
Having estimated
respiratory and cardiac MVFs, a 5D MoCo image reconstruction was performed
(Fig. 3). This was achieved by applying the estimated MVFs to a 5D double-gated
gridding reconstruction (
Nr
= 20, Δ
r = 10%,
Nc = 10, Δ
c =
20%, matrix size: 256×256×60×20×10, undersampling
factor: 27.9). That means for any arbitrary combination of respiratory
motion phase
r and cardiac motion
phase
c, all other combinations (
r’,
c’)
were warped onto (
r,
c) and averaged. Figure 3 shows the deformation
path (
r’,
c’) → (
rref,
c’) → (
rref,
c) → (
r,
c)
for one combination (
r’,
c’) as an example. Thus, 100% of the
measured raw data were used for the reconstruction of each cardio-respiratory combination
(
r,
c).
Results
Figure
4 shows representative reconstructions of 3D gridding, 5D double-gated gridding
and 5D MoCo. 3D gridding reconstructions yielded motion blur caused by
respiratory and cardiac motion. 5D double-gated gridding images exhibited high
noise levels and severe streak artifacts, which arose from the strong radial
undersampling as each combination (
r,
c) consisted of only 2% of the
measured raw data. The here-proposed 5D MoCo reconstruction achieved high image
sharpness because the images were fully compensated for organ motion. They
further showed low noise and low streak artifact levels because each combination
(
r,
c) was reconstructed from 100% of the measured raw data. As can be
seen in Fig. 5, different combinations of respiratory and cardiac motion phases
(
r,
c) were clearly resolved with identical image quality.
Conclusion and Discussion
In
this study, we proposed a new method for 5D respiratory and cardiac motion
compensation. The method enables time-resolved 5D MR imaging with acquisition
times as short as 2 min without compromises in temporal or spatial resolution,
size of field-of-view, noise and artifact level.
Acknowledgements
No acknowledgement found.References
1. Celicanin Z, Bieri O. 5DMRI of Moving Organs. Proc.
Intl. Soc. Mag. Reson. Med 2015;23:3683.
2. Feng L, Coppo S, Piccini D, et al. Five-dimensional
cardiac and respiratory motion-resolved whole-heart MRI. Proc. Intl. Soc. Mag.
Reson. Med 2015;23:0027.
3. Brehm
M, Paysan P, Oelhafen M, et al. Self-adapting cyclic registration for
motion-compensated cone-beam CT in image-guided radiation therapy. Med. Phys.
2012;39(12):7603-7618.