Five-Dimensional Respiratory and Cardiac Motion Compensation Based on Strongly Undersampled MR Data
Christopher M Rank1, Sebastian Sauppe1, Thorsten Heußer1, Andreas Wetscherek1, and Marc Kachelrieß1

1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany

Synopsis

We propose a new method for 5D respiratory and cardiac motion compensation (MoCo), which employs highly undersampled MR data and thus requires acquisition times as low as 2 minutes. Radial MR data of the thorax of three free-breathing patients were acquired. Respiratory and cardiac motion vector fields were estimated allowing for 5D MoCo reconstructions, which employ 100% of the measured raw data for reconstruction of each combination of respiratory and cardiac phase. These 5D MoCo reconstructions clearly resolve different combinations of respiratory and cardiac phases while achieving high temporal and spatial resolution as well as low noise and artifact levels.

Purpose

Dynamic imaging of organs can provide valuable information for radiotherapy applications or for studying physiology. However, time-resolved 5D (3D + respiratory + cardiac) MR imaging is time-consuming and recent approaches have reported acquisition times in the range of 15 min1,2. To allow for acquisition times as low as 2 min, we propose a new method for 5D respiratory and cardiac motion-compensated image reconstruction, which is based on radial MR data with very high undersampling.

Methods

Contrast-enhanced MR data covering the thorax of three free-breathing patients were acquired at 1.5 T (Magnetom Aera, Siemens Healthcare, Erlangen, Germany). Data acquisition and evaluation was in accordance with the local ethics committee and informed consent was obtained from each patient. We applied a vendor-provided radial stack-of-stars sequence with golden angle radial spacing and sagittal slice orientation: total acquisition time: 2.0 min, radial spokes per slice: 720, field-of-view: 385×385×300 mm3, voxel size: 1.5×1.5×5.0 mm3, TR/TE = 3.77/1.69 ms, 60 slices (60% slice resolution, 33% slice oversampling, 6/8 partial Fourier), flip angle: 12°, fat supression activated. Respiratory and cardiac motion signals used for self-gating were estimated from the k-space center for each acquired spoke. A bandpass filter was applied to distinguish between respiratory motion (filter range: 0.1 – 0.5 Hz) and cardiac motion (filter range: 0.5 – 2.5 Hz) and the signals were corrected for a baseline drift. Motion vector fields (MVFs) were estimated independently for respiratory and cardiac motion using a newly-developed algorithm, which alternates between image reconstruction and motion estimation. To increase robustness, deformable registrations were carried out between adjacent motion phases and regularized by cyclic constraints3. In a first step, respiratory MVFs were estimated (Fig. 1) using Nr = 20 overlapping respiratory motion bins with a width of Δr = 10% and neglecting cardiac motion (Nc = 1, Δc = 100%). In a second step cardiac MVFs were estimated (Fig. 2) assuming that respiratory motion in the end-exhale plateau can be neglected. Thus 25% of the raw data most consistent to end-exhale (Nr = 1, Δr = 25%, r = rref) divided into Nc = 10 overlapping cardiac motion bins with a width of Δc = 20% were used for estimation of cardiac MVFs. Having estimated respiratory and cardiac MVFs, a 5D MoCo image reconstruction was performed (Fig. 3). This was achieved by applying the estimated MVFs to a 5D double-gated gridding reconstruction (Nr = 20, Δr = 10%, Nc = 10, Δc = 20%, matrix size: 256×256×60×20×10, undersampling factor: 27.9). That means for any arbitrary combination of respiratory motion phase r and cardiac motion phase c, all other combinations (r’, c’) were warped onto (r, c) and averaged. Figure 3 shows the deformation path (r’, c’) → (rref, c’) → (rref, c) → (r, c) for one combination (r’, c’) as an example. Thus, 100% of the measured raw data were used for the reconstruction of each cardio-respiratory combination (r, c).

Results

Figure 4 shows representative reconstructions of 3D gridding, 5D double-gated gridding and 5D MoCo. 3D gridding reconstructions yielded motion blur caused by respiratory and cardiac motion. 5D double-gated gridding images exhibited high noise levels and severe streak artifacts, which arose from the strong radial undersampling as each combination (r, c) consisted of only 2% of the measured raw data. The here-proposed 5D MoCo reconstruction achieved high image sharpness because the images were fully compensated for organ motion. They further showed low noise and low streak artifact levels because each combination (r, c) was reconstructed from 100% of the measured raw data. As can be seen in Fig. 5, different combinations of respiratory and cardiac motion phases (r, c) were clearly resolved with identical image quality.

Conclusion and Discussion

In this study, we proposed a new method for 5D respiratory and cardiac motion compensation. The method enables time-resolved 5D MR imaging with acquisition times as short as 2 min without compromises in temporal or spatial resolution, size of field-of-view, noise and artifact level.

Acknowledgements

No acknowledgement found.

References

1. Celicanin Z, Bieri O. 5DMRI of Moving Organs. Proc. Intl. Soc. Mag. Reson. Med 2015;23:3683.

2. Feng L, Coppo S, Piccini D, et al. Five-dimensional cardiac and respiratory motion-resolved whole-heart MRI. Proc. Intl. Soc. Mag. Reson. Med 2015;23:0027.

3. Brehm M, Paysan P, Oelhafen M, et al. Self-adapting cyclic registration for motion-compensated cone-beam CT in image-guided radiation therapy. Med. Phys. 2012;39(12):7603-7618.

Figures

Gating for estimation of respiratory motion vector fields.

Gating for estimation of cardiac motion vector fields.

Example of one deformation path for MoCo reconstruction.

Comparison of different reconstruction methods.

MoCo reconstructions of different combinations of respiratory and cardiac motion phases (r, c).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0786