Synopsis
Sliding slice MRI is a technique which uses a magnetization
prepared sliding 2D slice to cast respiratory motion artifacts as geometric
distortions rather than diagnostically obscuring ghosting routinely associated
with 3D phase-encoding. In this work, we present the combination of
simultaneous-multi-slice with pseudo-random Cartesian based sliding slice
sampling. This combination allows increased frame rates, FOV tailoring, and
reduces sensitivity to off-resonance compared to past non-Cartesian radial and spiral
based approaches. Preliminary results are shown in moving phantoms and in-vivo
free breathing DCE, demonstrating very good image quality.Purpose
Respiratory motion remains a major challenge to the
use of MRI in the body. This is most notable in dynamic contrast enhanced (DCE)
imaging of the chest and abdomen, which requires continuous imaging to capture
bolus dynamics. While numerous motion-insensitive, free-breathing, accelerated 3D
imaging strategies have been proposed
1-2, they remain sensitive to
complex respiratory patterns. The alternative use of rapid 2D multi-slice
imaging would be appealing due to its ability to freeze motion; however, 2D
multi-slice imaging is highly sensitive to in-flow and steady state preparation.
Recently, sliding 2D slices were proposed as a means to alleviate these
artifacts and capture images in which motion artifacts manifest as correctable geometric
distortions
3. A drawback to this technique is the need for highly
accelerated 2D spiral imaging which is sensitive to off-resonance artifacts. In
this work, we investigate the synergistic combination of simultaneous
multi-slice, sliding slices, pseudo random Cartesian sampling with post-processing
registration to allow for high frame rates, lower demands on in plane acceleration,
and FOV tailoring.
Methods
A pulse sequence diagram is shown in Figure 1. As
in the initial spiral based implementation, imaging was performed by
interleaving a 3D saturation pulse with a 2D imaging module. Data is sampled as
shown in Fig 2 with data collected during a continuous sweep of the excitation
frequency of a multi-band RF pulse. During this sweep, readouts are collected
with ky phase encodings utilizing pseudorandom sampling, based on
the golden ratio4. As viewed as a function of time in Figure 2b,
this creates a set of images with overlapping replicates of the underlying
image. The signal of each sample is modeled as a function of the k-space
position (kx,ky) and the location of the slices (z1,z2):
$$d(k_x,k_y,z_1,z_2)=\int_{}^{}\int_{}^{}\int_{}^{} S(x,y,z)e^{i2\pi(xk_x+yk_y)}\left[G(z-z_1)I_{t+1}+G(z-z_2)I_{t}\right]$$
where G is the
slice excitation profile; S is the
sensitivity map; and It is
the image of time frame t. Making a discrete
approximation of Eq1, images can be recovered utilizing a penalized least
squares, with the penalty being a smoothness constraint to prevent super
resolution effects (i.e. the resolution in z should not exceed the slice
thickness). Note due to the interleaved nature of the data in time and space,
the reconstruction of any given slice requires the reconstruction of all slices
at all time points.
To evaluate feasibility, phantom and healthy volunteers
(after IRB approval and informed consent) were imaged on a clinical 3.0T system
(MR750, GE Healthcare, WI, USA). Images were first collected from the ADNI phantom
placed in an 8-channel head coil.
Sagittal images were acquired with 20mm, ~8Hz sinusoidal S/I motion of
the table with 2D, 3D, and simultaneous sliding slice. Sliding slice parameters included: resolution=1.5x1.9x4mm3,TE/TE=1.3/3.8ms, flipimaging=15°, flipsat=8°,
with 2 (ky) x 2 (slice) parallel imaging acceleration, 8s temporal
footprint, and 20 time frames. 2D and 3D
sequences were matched to the same temporal footprint and spatial resolution.
Apparent CNR was calculated using the ratio of the highest CNR sphere to that
of the background.
Dynamic in-vivo free breathing, DCE imaging
of the entire thoracic cavity was subsequently performed during bolus passage
of a 0.1mmol/kg dose of gadobenate dimeglumine with a 32-channel torso array coil (Neocoil, Pewaukee,
WI, USA). Imaging parameters were
identical to those used for phantom experiments.
Results
Figure 3a shows images acquired without slice
unwrapping, with the temporal dimension along the horizontal axis. Periodic
motion is clearly visible with two replicates, representing the two
slices. Replicates were successfully
unwrapped, creating images at twice the frame rate. Figure 3b shows 2D, 3D and sliding images.
Motion causes significant artifact and ghosting in the 3D images. 2D imaging
effectively freezes motion; however, images suffer from lack of contrast and
stair stepping artifacts. The sliding approach maintains high contrast and
motion insensitivity, which is reflected in high apparent CNR (23.2) compared
to 2D (1.15) and 3D (8.34). Figure 4,
shows the source sagittal images of in-vivo feasibility testing. Images
demonstrate complete freezing of motion, including Cardiac contraction. Reformatting the images into the coronal
plane, as in Figure 5b, shows the manifestation of motion in this case with the
motion state being different in the R/L dimension. This type of motion is
potentially correctable via non-rigid registration, preliminarily shown in Figure
5b.
Discussion and Conclusion
The combination of simultaneous multi-slice imaging
with sliding slices allows additional acceleration that translates to higher
temporal frame rates. The incorporation with Cartesian sampling facilitates FOV
tailoring and increases robustness to off-resonance artifacts. This allows for
sagittal acquisitions, casting most respiratory motion in-plane, which is amenable
to correction via registration.
Acknowledgements
We gratefully
acknowledge GE Healthcare and NIH-NS066982 for assistance and support.References
1. Feng, L., Axel, L., Chandarana, H., Block, K. T.,
Sodickson, D. K. and Otazo, R. (2015), XD-GRASP: Golden-angle radial MRI with
reconstruction of extra motion-state dimensions using compressed sensing. Magn
Reson Med. doi: 10.1002/mrm.25665
2. Cheng, J. Y., Zhang, T., Ruangwattanapaisarn, N.,
Alley, M. T., Uecker, M., Pauly, J. M., Lustig, M. and Vasanawala, S. S.
(2015), Free-breathing pediatric MRI with nonrigid motion correction and
acceleration. J. Magn. Reson. Imaging, 42: 407–420.
doi: 10.1002/jmri.24785
3. Johnson
KM, Holmes J, and Reeder S. Imaging in the presence of Motion with Sliding
Slice Distortions. (2015). Proceedings of the 23rd annual meeting of
ISMRM. Toronto, ON, Canada. #810
4. Prieto,
C., Doneva, M., Usman, M., Henningsson, M., Greil, G., Schaeffter, T. and
Botnar, R. M. (2015), Highly efficient respiratory motion compensated
free-breathing coronary mra using golden-step Cartesian acquisition. J. Magn.
Reson. Imaging, 41: 738–746. doi: 10.1002/jmri.24602