Synopsis
In contrast to studies at
3T, where the whole body coil is used for RF transmission, studies at 7T use
local transcieve coils, which have limited B1+ coverage
producing smaller temporal bolus widths that need to be estimated in order to
achieve proper renal blood flow (RBF) quantification. To estimate the temporal
bolus width and to quantify RBF at 7T, single breath-hold renal perfusion
studies were performed using the FAIR ss-FSE method with varied delay times.
Based on the results form multi-delay perfusion study, quantitative renal
perfusion imaging was further achieved by using a single-subtraction approach. PURPOSE
The
feasibility of performing renal perfusion imaging with arterial spin labeling
(ASL) methods has been demonstrated at 7T
1-3.
In particular, studies suggest that renal ASL perfusion imaging can be achieved within a
single breath-hold at 7T
3. Furthermore, the comparisons of
renal ASL imaging between 3T and 7T indicate that 7T can provide higher
SNR efficiency
4. However, quantitative renal perfusion
imaging has not been achieved at 7T. In contrast to studies at 3T, where the
whole body coil is used for RF transmission, studies at 7T use local transcieve
coils, which have limited B
1+ coverage producing smaller
temporal bolus widths that need to be estimated in order to achieve proper
renal blood flow (RBF) quantification. To estimate the temporal bolus width and
to quantify RBF at 7T, single breath-hold renal perfusion studies were performed
using the FAIR ss-FSE method with varied delay times. Based on the results from multi-delay perfusion study, quantitative renal perfusion imaging was achieved by
using a single-subtraction approach; a single perfusion scan was performed with
a fixed temporal bolus width that was defined by applying saturations
anteriorly outside the imaging volume to cutoff labeled blood supply at a
properly selected delay time.
METHODS
Studies were performed on a Siemens whole body MRI scanner
with an external 16-channel transceiver TEM stripline array driven by a series
of 16, 1 kW amplifiers (CPC, Pittsburgh, PA). Local B0 shimming
was achieved by using volumetric phase maps acquired within a single
breath-hold 5. A
dynamic B1+
shimming strategy 6 was
applied using 2 solutions at unique spatial locations,
including: 1) one for the arterial spin labeling
inversions covering the descending aorta; and 2) another for the combined pre-saturation and imaging slice location 3. All B1+
shim optimizations were based on a tradeoff solution between RF efficiency and
B1+ field homogeneity.
FAIR ss-FSE imaging
parameters for a single oblique coronal imaging slice were: TR/TE =
3500-4000/16 ms, parallel acceleration factor = 4, hyper echo flip angle = 90°,
resolution = 2 x 2 x 5 mm3, left-to-right phase encoding with 50-80%
oversampling, partial Fourier = 5/8, post-labeling delays = {0.3, 0.6, 0.9,
1.2, 1.5, 1.8, 2.1} s, selective/spatially-confined inversion slab size =
25/160 mm, and 4 pairs of label and control images. In
single-subtraction renal perfusion imaging, after inversion RF pulses, four
saturation RF pulses 50 ms apart were applied anteriorly outside the imaging
volume at a fixed delay time to cutoff the supply of labeled blood and
therefore prescribe a temporal bolus width, and followed by a post-bolus delay
time prior to imaging readout. The single-subtraction renal perfusion imaging
used the same MRI parameters as in multi-delay study, except that only one
perfusion scan was performed with a 600 ms temporal bolus width and a 600 ms post-bolus delay.
Post-processing,
including motion correction for small drifts during the breath-hold, was
performed within Matlab and SPM. Multi-delay perfusion signals of renal cortex
and medulla were fitted to a simplified three-phase, single-compartment model 7 using an
iterative nonlinear least-square model-fitting program to estimate the ATT,
temporal bolus width and RBF. RBF quantification using perfusion data from the
single-subtraction imaging study was also achieved using the single-compartment
model 7.
RESULTS AND DISCUSSION
The
imaging results of multi-delay perfusion imaging study from one representative
volunteer are presented in Figure 1, and the estimated ATT, temporal bolus
width, and RBF for renal cortex and medulla are shown in Figure 2. Figure 3
shows the results from the single-subtraction imaging study.
Study results indicate that
the temporal bolus widths achieved with the local transceive body coils at 7T
are larger than 600 ms across subjects for both renal cortex and medulla.
Although with renal perfusion studies using varied delay times, RBF can be
estimated from iterative model fitting, such an approach requires multiple ASL
acquisitions, and therefore time consuming. Knowledge of the temporal bolus width achieved at 7T will
allow the proper definition of a fixed bolus width, which is essential in
obtaining RBF measurements using the single-subtraction approach.
CONCLUSION
Both
temporal bolus width and RBF have been successfully estimated by performing
multi-delay perfusion studies using the single-breath FAIR ss-FSE method with
local
transcieve coils at 7T. The estimated temporal bolus width
provides the basis for the successful and efficient application of the
single-subtract approach for quantitative renal perfusion imaging.
Acknowledgements
P41 EB015894, NIH-R01 EB007327, 2R42EB013543-02, UL1TR000114 and the University
of Minnesota Foundation. The content is solely the responsibility of the
authors and does not necessarily represent the official views of the National
Institutes of Health.
References
1. Li X, Snyder C, Van de Moortele P-F, Ugurbil K, Metzger
GJ. Non-Contrast Enhanced Human Renal Perfusion Imaging Using Arterial Spin
Labeling at 7 T: Initial Experience. In: Proceedings of the 20th Annual
Meeting of ISMRM, Melbourne, Australia 2012:Abstract 1310.
2. Li X, Ugurbil K, Metzger GJ. Theoretical
Evaluation of Ultra High Field Benefits to Non-contrast Enhanced Renal
Perfusion Imaging Using FAIR-EPI. In: Proceedings of the 20th Annual
Meeting of ISMRM, Salt Lake City, USA 2013:Abstract 1540.
3. Li
X, Snyder C, Van de Moortele P-F, Ugurbil K, Metzger GJ. Feasibility of Single
Breath-hold Renal Perfusion Imaging at 7T. In: Proceedings of the 21st Annual Meeting of ISMRM, Salt Lake
City, USA 2013:Abstract 0030.
4. Li
X., Auerbach EJ, Van de Moortele Pierre-Francois, Ugurbil K, Metzger,
GJ. Theoretical and Experimental Comparisons of Single Breath-Hold Renal
Perfusion Imaging between 3T and 7T. In: Proceedings of the 23rd Annual Meeting of ISMRM, Toronto,
Ontario, Canada 2015:Abstract 0030.
5.
Shah S, Kellman P, Greiser A, Weale PJ, Zuehlsdorff S, Jerecic aR. Rapid
fieldmap estimation for cardiac shimming. In: Proceedings of the 17th Annual
Meeting of ISMRM, Honolulu, HI, 2009. Abstract 566.
6. Metzger
GJ, Auerbach EJ, Akgun C, Simonson J, Bi X, Ugurbil K, van de Moortele PF.
Dynamically applied B1+ shimming solutions for non-contrast enhanced renal
angiography at 7.0 tesla. Magn. Reson. Med. 2013; 69: 114–126.
7. Li X, Sarkar SN, Purdy DE, Spence JS,
Haley RW, Briggs RW. Anteroposterior perfusion heterogeneity in human
hippocampus measured by arterial spin labeling MRI. NMR Biomed. 2013; 26:
613–621.